Title: SU-F-P-15: Report On AAPM TG 178 Gamma Knife Dosimetry and Quality Assurance

Abstract

Purpose: AAPM Task Group 178 Gamma Stereotactic Radiosurgery Dosimetry and Quality Assurance was formed in August, 2008. The Task Group has 12 medical physicists, two physicians and two consultants. Methods: A round robin dosimetry intercomparison of proposed ionization chambers, electrometer and dosimetry phantoms was conducted over a 15 month period in 2011 and 2012 (Med Phys 42, 11, Nov, 2015). The data obtained at 9 institutions (with ten different Elekta Gamma Knife units) was analyzed by the lead author using several protocols. Results: The most consistent results were obtained using the Elekta ABS 16cm diameter phantom, with the TG-51 protocol modified as recommended by Alfonso et al (Med Phys 35, 11, Nov 2008). A key white paper (Med Phys, in press) sponsored by Elekta Corporation, was used to obtain correction factors for the ionization chambers and phantoms used in this intercomparison. Consistent results were obtained for both Elekta Gamma Knife Model 4C and Gamma Knife Perfexion units as measured with each of two miniature ionization chambers Conclusion: The full TG 178 report gives clinical history and background of gamma stereotactic radiosurgery, clinical examples and history, quality assurance recommendations and outline of possible dosimetry protocols. The report will be reviewed bymore » the AAPM Working Group on Recommendations for Radiotherapy External Beam Quality Assurance and then by the AAPM Science Council before publication in Medical Physics. Consultant to Elekta, Inc.« less

Authors:
 [1]
  1. San Diego Medical Physics, Solana Beach, CA (United States)
Publication Date:
OSTI Identifier:
22624457
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; CONSULTANTS; CORRECTIONS; DOSIMETRY; ELECTROMETERS; IONIZATION CHAMBERS; PHANTOMS; QUALITY ASSURANCE; RADIOTHERAPY; RECOMMENDATIONS; REVIEWS; SURGERY

Citation Formats

Goetsch, S. SU-F-P-15: Report On AAPM TG 178 Gamma Knife Dosimetry and Quality Assurance. United States: N. p., 2016. Web. doi:10.1118/1.4955722.
Goetsch, S. SU-F-P-15: Report On AAPM TG 178 Gamma Knife Dosimetry and Quality Assurance. United States. doi:10.1118/1.4955722.
Goetsch, S. Wed . "SU-F-P-15: Report On AAPM TG 178 Gamma Knife Dosimetry and Quality Assurance". United States. doi:10.1118/1.4955722.
@article{osti_22624457,
title = {SU-F-P-15: Report On AAPM TG 178 Gamma Knife Dosimetry and Quality Assurance},
author = {Goetsch, S},
abstractNote = {Purpose: AAPM Task Group 178 Gamma Stereotactic Radiosurgery Dosimetry and Quality Assurance was formed in August, 2008. The Task Group has 12 medical physicists, two physicians and two consultants. Methods: A round robin dosimetry intercomparison of proposed ionization chambers, electrometer and dosimetry phantoms was conducted over a 15 month period in 2011 and 2012 (Med Phys 42, 11, Nov, 2015). The data obtained at 9 institutions (with ten different Elekta Gamma Knife units) was analyzed by the lead author using several protocols. Results: The most consistent results were obtained using the Elekta ABS 16cm diameter phantom, with the TG-51 protocol modified as recommended by Alfonso et al (Med Phys 35, 11, Nov 2008). A key white paper (Med Phys, in press) sponsored by Elekta Corporation, was used to obtain correction factors for the ionization chambers and phantoms used in this intercomparison. Consistent results were obtained for both Elekta Gamma Knife Model 4C and Gamma Knife Perfexion units as measured with each of two miniature ionization chambers Conclusion: The full TG 178 report gives clinical history and background of gamma stereotactic radiosurgery, clinical examples and history, quality assurance recommendations and outline of possible dosimetry protocols. The report will be reviewed by the AAPM Working Group on Recommendations for Radiotherapy External Beam Quality Assurance and then by the AAPM Science Council before publication in Medical Physics. Consultant to Elekta, Inc.},
doi = {10.1118/1.4955722},
journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 43,
place = {United States},
year = {2016},
month = {6}
}