Title: SU-C-209-05: Monte Carlo Model of a Prototype Backscatter X-Ray (BSX) Imager for Projective and Selective Object-Plane Imaging

Abstract

Purpose: To develop a Monte Carlo N-Particle (MCNP) model for the validation of a prototype backscatter x-ray (BSX) imager, and optimization of BSX technology for medical applications, including selective object-plane imaging. Methods: BSX is an emerging technology that represents an alternative to conventional computed tomography (CT) and projective digital radiography (DR). It employs detectors located on the same side as the incident x-ray source, making use of backscatter and avoiding ring geometry to enclose the imaging object. Current BSX imagers suffer from low spatial resolution. A MCNP model was designed to replicate a BSX prototype used for flaw detection in industrial materials. This prototype consisted of a 1.5mm diameter 60kVp pencil beam surrounded by a ring of four 5.0cm diameter NaI scintillation detectors. The imaging phantom consisted of a 2.9cm thick aluminum plate with five 0.6cm diameter holes drilled halfway. The experimental image was created using a raster scanning motion (in 1.5mm increments). Results: A qualitative comparison between the physical and simulated images showed very good agreement with 1.5mm spatial resolution in plane perpendicular to incident x-ray beam. The MCNP model developed the concept of radiography by selective plane detection (RSPD) for BSX, whereby specific object planes can be imagedmore » by varying kVp. 10keV increments in mean x-ray energy yielded 4mm thick slice resolution in the phantom. Image resolution in the MCNP model can be further increased by increasing the number of detectors, and decreasing raster step size. Conclusion: MCNP modelling was used to validate a prototype BSX imager and introduce the RSPD concept, allowing for selective object-plane imaging. There was very good visual agreement between the experimental and MCNP imaging. Beyond optimizing system parameters for the existing prototype, new geometries can be investigated for volumetric image acquisition in medical applications. This material is based upon work supported under an Integrated University Program Graduate Fellowship sponsored by the Department of Energy Office of Nuclear Energy.« less

Authors:
; ; ;  [1]
  1. University of Florida, Gainesville, FL (United States)
Publication Date:
OSTI Identifier:
22624348
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; COMPUTERIZED TOMOGRAPHY; ENERGY YIELD; IMAGES; MONTE CARLO METHOD; NAI DETECTORS; OPTIMIZATION; PHANTOMS; SCINTILLATIONS; SIMULATION; SODIUM IODIDES; SPATIAL RESOLUTION; VALIDATION; X-RAY SOURCES

Citation Formats

Rolison, L, Samant, S, Baciak, J, and Jordan, K. SU-C-209-05: Monte Carlo Model of a Prototype Backscatter X-Ray (BSX) Imager for Projective and Selective Object-Plane Imaging. United States: N. p., 2016. Web. doi:10.1118/1.4955594.
Rolison, L, Samant, S, Baciak, J, & Jordan, K. SU-C-209-05: Monte Carlo Model of a Prototype Backscatter X-Ray (BSX) Imager for Projective and Selective Object-Plane Imaging. United States. doi:10.1118/1.4955594.
Rolison, L, Samant, S, Baciak, J, and Jordan, K. 2016. "SU-C-209-05: Monte Carlo Model of a Prototype Backscatter X-Ray (BSX) Imager for Projective and Selective Object-Plane Imaging". United States. doi:10.1118/1.4955594.
@article{osti_22624348,
title = {SU-C-209-05: Monte Carlo Model of a Prototype Backscatter X-Ray (BSX) Imager for Projective and Selective Object-Plane Imaging},
author = {Rolison, L and Samant, S and Baciak, J and Jordan, K},
abstractNote = {Purpose: To develop a Monte Carlo N-Particle (MCNP) model for the validation of a prototype backscatter x-ray (BSX) imager, and optimization of BSX technology for medical applications, including selective object-plane imaging. Methods: BSX is an emerging technology that represents an alternative to conventional computed tomography (CT) and projective digital radiography (DR). It employs detectors located on the same side as the incident x-ray source, making use of backscatter and avoiding ring geometry to enclose the imaging object. Current BSX imagers suffer from low spatial resolution. A MCNP model was designed to replicate a BSX prototype used for flaw detection in industrial materials. This prototype consisted of a 1.5mm diameter 60kVp pencil beam surrounded by a ring of four 5.0cm diameter NaI scintillation detectors. The imaging phantom consisted of a 2.9cm thick aluminum plate with five 0.6cm diameter holes drilled halfway. The experimental image was created using a raster scanning motion (in 1.5mm increments). Results: A qualitative comparison between the physical and simulated images showed very good agreement with 1.5mm spatial resolution in plane perpendicular to incident x-ray beam. The MCNP model developed the concept of radiography by selective plane detection (RSPD) for BSX, whereby specific object planes can be imaged by varying kVp. 10keV increments in mean x-ray energy yielded 4mm thick slice resolution in the phantom. Image resolution in the MCNP model can be further increased by increasing the number of detectors, and decreasing raster step size. Conclusion: MCNP modelling was used to validate a prototype BSX imager and introduce the RSPD concept, allowing for selective object-plane imaging. There was very good visual agreement between the experimental and MCNP imaging. Beyond optimizing system parameters for the existing prototype, new geometries can be investigated for volumetric image acquisition in medical applications. This material is based upon work supported under an Integrated University Program Graduate Fellowship sponsored by the Department of Energy Office of Nuclear Energy.},
doi = {10.1118/1.4955594},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • On metallic specimens in which a secondary phase has been selectively removed by a chemical etchant, the use of backscatter electron (BSE) imaging yields images that are more readily segmented with image processing algorithms than other modes of imaging in the scanning electron microscope. The contrast mechanisms in this imaging mode, however, produce a bias in the observation of particle sizes and volume fractions due to the effects of the electron interaction volume in the specimen. This stereological bias is quantified using Monte Carlo (MC) simulation of backscatter images. It is observed that the overprojection of features with centroids residingmore » beneath the plane of polish is largely canceled out by the reduced segmentation size of features with centroids residing above the plane of polish. - Research Highlights: {yields} Backscatter imaging of selectively-etched surfaces can facilitate segmentation. {yields} Backscatter imaging of voids is simulated to estimate imaging/observation biases. {yields} The biases are quantified and incorporated into the stereological calculation. {yields} Systematic errors and imaging biases are observed to counteract one another. {yields} Results are illustrated using a bimodal gamma prime distribution in a Ni superalloy.« less
  • Purpose: Current, clinically applicable intracavitary brachytherapy applicators that utilize shielded ovoids contain a pair of tungsten-alloy shields which serve to reduce dose delivered to the rectum and bladder during source afterloading. After applicator insertion, these fixed shields are not necessarily positioned to provide optimal shielding of these critical structures due to variations in patient anatomies. The authors present a dosimetric evaluation of a novel prototype intracavitary brachytherapy ovoid [anatomically adaptive applicator (A{sup 3})], featuring a single shield whose position can be adjusted with two degrees of freedom: Rotation about and translation along the long axis of the ovoid. Methods: Themore » dosimetry of the device for a HDR {sup 192}Ir was characterized using radiochromic film measurements for various shield orientations. A MCNPX Monte Carlo model was developed of the prototype ovoid and integrated with a previously validated model of a v2 mHDR {sup 192}Ir source (Nucletron Co.). The model was validated for three distinct shield orientations using film measurements. Results: For the most complex case, 91% of the absolute simulated and measured dose points agreed within 2% or 2 mm and 96% agreed within 10% or 2 mm. Conclusions: Validation of the Monte Carlo model facilitates future investigations into any dosimetric advantages the use of the A{sup 3} may have over the current state of art with respect to optimization and customization of dose delivery as a function of patient anatomical geometries.« less
  • In ICF, the understanding of laser plasma scattering processes is essential for laser target coupling and for controlling the symmetry of indirect drive implosions. The existing Nova full aperture backscatter station has been useful in understanding laser plasma instabilities occurring in hohlraums by measuring the quantity, spectral distribution, and near-field spatial distributions of Brillouin and more recently Raman backscatter. Equally important is an understanding of the far-field spatial intensity distribution which could help in understanding filamentation, threshold and saturation processes. This article describes a broadband, color-corrected far-field imager and associated diagnostics capable of imaging the source of scattered light tomore » better than 25 {mu}m resolution. Brillouin and Raman backscatter can be imaged through the Nova beam-7 focusing lens or the imager can be used like a microscope to image side scatter from other beams. {copyright} {ital 1997 American Institute of Physics.}« less
  • We present a computation of {pi}{sup 0} nuclear modification factor with respect to the reaction plane in Au+Au collisions at {radical}(s{sub NN})=200 GeV, based on a Monte Carlo model of elastic energy loss of hard partons traversing the bulk hydrodynamical medium created in ultrarelativistic heavy-ion collisions. We find the incoherent nature of elastic energy loss incompatible with the measured data.
  • Purpose: To validate the physics parameters of a Monte Carlo model for patient plane leakage calculations on the 6MV Unique linac by comparing the simulations against IEC patient plane leakage measurements. The benchmarked model can further be used for shielding design optimization, to predict leakage in the proximity of intended treatment fields, reduce the system weight and cost, and improve components reliability. Methods: The treatment head geometry of the Unique linac was simulated in Geant4 (v9.4.p02 with “Opt3” standard electromagnetic physics list) based on CAD drawings of all collimation and shielding components projected from the target to the area withinmore » 2m from isocenter. A 4×4m2 scorer was inserted 1m from the target in the patient plane and multiple phase space files were recorded by performing a 40-node computing cluster simulation on the EC2 cloud. The photon energy fluence was calculated relative to the value at isocenter for a 10×10cm2 field using 10×10mm2 bins. Tungsten blocks were parked accordingly to represent MLC120. The secondary particle contamination to patient plane was eliminated by “killing” those particles prior to the primary collimator entrance using a “kill-plane”, which represented the upper head shielding components not being modeled. Both IEC patient-plane leakage and X/Y-jaws transmission were simulated. Results: The contribution of photons to energy fluence was 0.064% on average, in excellent agreement with the experimental data available at 0.5, 1.0, and 1.5m from isocenter, characterized by an average leakage of 0.045% and a maximum leakage of 0.085%. X- and Y-jaws transmissions of 0.43% and 0.44% were found in good agreement with measurements of 0.48% and 0.43%, respectively. Conclusion: A Geant4 model based on energy fluence calculations for the 6MV Unique linac was created and validated using IEC patient plane leakage measurements. The “kill-plane” has effectively eliminated electron contamination to patient plane in these simulations.« less