Control and automation of the Pegasus multi-point Thomson scattering system
Journal Article
·
· Review of Scientific Instruments
- Department of Engineering Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706 (United States)
A new control system for the Pegasus Thomson scattering diagnostic has recently been deployed to automate the laser operation, data collection process, and interface with the system-wide Pegasus control code. Automation has been extended to areas outside of data collection, such as manipulation of beamline cameras and remotely controlled turning mirror actuators to enable intra-shot beam alignment. Additionally, the system has been upgraded with a set of fast (∼1 ms) mechanical shutters to mitigate contamination from background light. Modification and automation of the Thomson system have improved both data quality and diagnostic reliability.
- OSTI ID:
- 22596509
- Journal Information:
- Review of Scientific Instruments, Vol. 87, Issue 11; Other Information: (c) 2016 Author(s); Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Control and automation of the Pegasus multi-point Thomson scattering system
A novel, cost-effective, multi-point Thomson scattering system on the Pegasus Toroidal Experiment (invited)
Public Data Set: Control and Automation of the Pegasus Multi-point Thomson Scattering System
Journal Article
·
Fri Aug 12 00:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:22596509
+3 more
A novel, cost-effective, multi-point Thomson scattering system on the Pegasus Toroidal Experiment (invited)
Journal Article
·
Fri Sep 16 00:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:22596509
+3 more
Public Data Set: Control and Automation of the Pegasus Multi-point Thomson Scattering System
Dataset
·
Fri Aug 12 00:00:00 EDT 2016
·
OSTI ID:22596509
+3 more