Title: Technical Note: Radiation properties of tissue- and water-equivalent materials formulated using the stoichiometric analysis method in charged particle therapy

Abstract

Purpose: Five tissue- and water-equivalent materials (TEMs) mimicking ICRU real tissues have been formulated using a previously established stoichiometric analysis method (SAM) to be applied in charged particle therapy. The purpose of this study was an experimental verification of the TEMs-SAM against charged particle beam measurements and for different computed tomography (CT) scanners. The potential of the TEMs-SAM to be employed in the dosimetry was also investigated. Methods: Experimental verification with three CT scanners was carried out to validate the calculated Hounsfield units (HUs) of the TEMs. Water-equivalent path lengths (WEPLs) of the TEMs for proton (106.8 MeV/u), helium (107.93 MeV/u), and carbon (200.3 MeV/u) ions were measured to be compared with the computed relative stopping powers. HU calibration curves were also generated. Results: Differences between the measured HUs of the TEMs and the calculated HUs of the ICRU real tissues for all CT scanners were smaller than 4 HU except for the skeletal tissues which deviated up to 21 HU. The measured WEPLs verified the calculated WEPLs of the TEMs (maximum deviation was 0.17 mm) and were in good agreement with the calculated WEPLs of the ICRU real tissues (maximum deviation was 0.23 mm). Moreover, the relative stopping powersmore » converted from the measured WEPLs differed less than 0.8% and 1.3% from the calculated values of the SAM and the ICRU, respectively. Regarding the relative nonelastic cross section per unit of volume for 200 MeV protons, the ICRU real tissues were generally well represented by the TEMs except for adipose which differed 3.8%. Further, the HU calibration curves yielded the mean and the standard deviation of the errors not larger than 0.5% and 1.9%, respectively. Conclusions: The results of this investigation implied the potential of the TEMs formulated using the SAM to be employed for both, beam dosimetry and HU calibration in charged particle therapy.« less

Authors:
;  [1];  [2];  [3];  [4];  [1]
  1. Radiation Oncology, University Hospital Erlangen, Erlangen 91054 (Germany)
  2. Faculty of Medicine, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen 91054, Germany and Department of Biophysics, GSI - Helmholtz Centre for Heavy Ion Research, Darmstadt 64291 (Germany)
  3. QRM - Quality Assurance in Radiology and Medicine GmbH, Möhrendorf 91096 (Germany)
  4. Department of Biophysics, GSI - Helmholtz Centre for Heavy Ion Research, Darmstadt 64291 (Germany)
Publication Date:
OSTI Identifier:
22579816
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 43; Journal Issue: 1; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; 61 RADIATION PROTECTION AND DOSIMETRY; ANIMAL TISSUES; CALIBRATION; COMPUTERIZED TOMOGRAPHY; DOSIMETRY; ICRU; PARTICLE BEAMS; RADIOTHERAPY; STOICHIOMETRY; STOPPING POWER; VERIFICATION

Citation Formats

Yohannes, Indra, Vasiliniuc, Stefan, Hild, Sebastian, Langner, Oliver, Graeff, Christian, Bert, Christoph, Faculty of Medicine, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen 91054, and Department of Biophysics, GSI - Helmholtz Centre for Heavy Ion Research, Darmstadt 64291. Technical Note: Radiation properties of tissue- and water-equivalent materials formulated using the stoichiometric analysis method in charged particle therapy. United States: N. p., 2016. Web. doi:10.1118/1.4938587.
Yohannes, Indra, Vasiliniuc, Stefan, Hild, Sebastian, Langner, Oliver, Graeff, Christian, Bert, Christoph, Faculty of Medicine, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen 91054, & Department of Biophysics, GSI - Helmholtz Centre for Heavy Ion Research, Darmstadt 64291. Technical Note: Radiation properties of tissue- and water-equivalent materials formulated using the stoichiometric analysis method in charged particle therapy. United States. https://doi.org/10.1118/1.4938587
Yohannes, Indra, Vasiliniuc, Stefan, Hild, Sebastian, Langner, Oliver, Graeff, Christian, Bert, Christoph, Faculty of Medicine, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen 91054, and Department of Biophysics, GSI - Helmholtz Centre for Heavy Ion Research, Darmstadt 64291. 2016. "Technical Note: Radiation properties of tissue- and water-equivalent materials formulated using the stoichiometric analysis method in charged particle therapy". United States. https://doi.org/10.1118/1.4938587.
@article{osti_22579816,
title = {Technical Note: Radiation properties of tissue- and water-equivalent materials formulated using the stoichiometric analysis method in charged particle therapy},
author = {Yohannes, Indra and Vasiliniuc, Stefan and Hild, Sebastian and Langner, Oliver and Graeff, Christian and Bert, Christoph and Faculty of Medicine, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen 91054 and Department of Biophysics, GSI - Helmholtz Centre for Heavy Ion Research, Darmstadt 64291},
abstractNote = {Purpose: Five tissue- and water-equivalent materials (TEMs) mimicking ICRU real tissues have been formulated using a previously established stoichiometric analysis method (SAM) to be applied in charged particle therapy. The purpose of this study was an experimental verification of the TEMs-SAM against charged particle beam measurements and for different computed tomography (CT) scanners. The potential of the TEMs-SAM to be employed in the dosimetry was also investigated. Methods: Experimental verification with three CT scanners was carried out to validate the calculated Hounsfield units (HUs) of the TEMs. Water-equivalent path lengths (WEPLs) of the TEMs for proton (106.8 MeV/u), helium (107.93 MeV/u), and carbon (200.3 MeV/u) ions were measured to be compared with the computed relative stopping powers. HU calibration curves were also generated. Results: Differences between the measured HUs of the TEMs and the calculated HUs of the ICRU real tissues for all CT scanners were smaller than 4 HU except for the skeletal tissues which deviated up to 21 HU. The measured WEPLs verified the calculated WEPLs of the TEMs (maximum deviation was 0.17 mm) and were in good agreement with the calculated WEPLs of the ICRU real tissues (maximum deviation was 0.23 mm). Moreover, the relative stopping powers converted from the measured WEPLs differed less than 0.8% and 1.3% from the calculated values of the SAM and the ICRU, respectively. Regarding the relative nonelastic cross section per unit of volume for 200 MeV protons, the ICRU real tissues were generally well represented by the TEMs except for adipose which differed 3.8%. Further, the HU calibration curves yielded the mean and the standard deviation of the errors not larger than 0.5% and 1.9%, respectively. Conclusions: The results of this investigation implied the potential of the TEMs formulated using the SAM to be employed for both, beam dosimetry and HU calibration in charged particle therapy.},
doi = {10.1118/1.4938587},
url = {https://www.osti.gov/biblio/22579816}, journal = {Medical Physics},
issn = {0094-2405},
number = 1,
volume = 43,
place = {United States},
year = {Fri Jan 15 00:00:00 EST 2016},
month = {Fri Jan 15 00:00:00 EST 2016}
}