Title: Dosimetric evaluation of 4 different treatment modalities for curative-intent stereotactic body radiation therapy for isolated thoracic spinal metastases

Abstract

To investigate the dosimetric characteristics of 4 SBRT-capable dose delivery systems, CyberKnife (CK), Helical TomoTherapy (HT), Volumetric Modulated Arc Therapy (VMAT) by Varian RapidArc (RA), and segmental step-and-shoot intensity-modulated radiation therapy (IMRT) by Elekta, on isolated thoracic spinal lesions. CK, HT, RA, and IMRT planning were performed simultaneously for 10 randomly selected patients with 6 body types and 6 body + pedicle types with isolated thoracic lesions. The prescription was set with curative intent and dose of either 33 Gy in 3 fractions (3F) or 40 Gy in 5F to cover at least 90% of the planning target volume (PTV), correspondingly. Different dosimetric indices, beam-on time, and monitor units (MUs) were evaluated to compare the advantages/disadvantages of each delivery modality. In ensuring the dose-volume constraints for cord and esophagus of the premise, CK, HT, and RA all achieved a sharp conformity index (CI) and a small penumbra volume compared to IMRT. RA achieved a CI comparable to those from CK, HT, and IMRT. CK had a heterogeneous dose distribution in the target as its radiosurgical nature with less dose uniformity inside the target. CK had the longest beam-on time and the largest MUs, followed by HT and RA. IMRT presentedmore » the shortest beam-on time and the least MUs delivery. For the body-type lesions, CK, HT, and RA satisfied the target coverage criterion in 6 cases, but the criterion was satisfied in only 3 (50%) cases with the IMRT technique. For the body + pedicle-type lesions, HT satisfied the criterion of the target coverage of ≥90% in 4 of the 6 cases, and reached a target coverage of 89.0% in another case. However, the criterion of the target coverage of ≥90% was reached in 2 cases by CK and RA, and only in 1 case by IMRT. For curative-intent SBRT of isolated thoracic spinal lesions, RA is the first choice for the body-type lesions owing to its delivery efficiency (time); the second choice is CK or HT; HT is the preferential choice for the body + pedicle-type lesions. This study suggests further clinical investigations with longer follow-up for these studied cases.« less

Authors:
 [1];  [2];  [1];  [2]; ; ; ; ;  [1];  [3];  [1];  [4]
  1. Department of Radiation Oncology, Chinese PLA General Hospital, 28 Fuxing Road, Beijing, 100853 (China)
  2. (China)
  3. Department of Radiation Oncology, Hainan Branch of Chinese PLA General Hospital, Haitang Bay, Sanya, 572000 (China)
  4. Department of Radiation Oncology, Monmouth Medical Center, 300 2nd Avenue, Long Branch, NJ 07740 (United States)
Publication Date:
OSTI Identifier:
22577869
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Dosimetry; Journal Volume: 41; Journal Issue: 2; Other Information: Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; 62 RADIOLOGY AND NUCLEAR MEDICINE; BEAM MONITORS; BEAMS; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; DOSIMETRY; EFFICIENCY; ESOPHAGUS; METASTASES; PATIENTS; RADIATION DOSES

Citation Formats

Yang, Jun, Department of Oncology, First Affiliated Hospital of Xinxiang Medical University, 88 Jiankang Road, Weihui, Henan, 453100, Ma, Lin, Department of Radiation Oncology, Hainan Branch of Chinese PLA General Hospital, Haitang Bay, Sanya, 572000, Wang, Xiao-Shen, Xu, Wei Xu, Cong, Xiao-Hu, Xu, Shou-Ping, Ju, Zhong-Jian, Du, Lei, Cai, Bo-Ning, and Yang, Jack. Dosimetric evaluation of 4 different treatment modalities for curative-intent stereotactic body radiation therapy for isolated thoracic spinal metastases. United States: N. p., 2016. Web. doi:10.1016/J.MEDDOS.2015.10.003.
Yang, Jun, Department of Oncology, First Affiliated Hospital of Xinxiang Medical University, 88 Jiankang Road, Weihui, Henan, 453100, Ma, Lin, Department of Radiation Oncology, Hainan Branch of Chinese PLA General Hospital, Haitang Bay, Sanya, 572000, Wang, Xiao-Shen, Xu, Wei Xu, Cong, Xiao-Hu, Xu, Shou-Ping, Ju, Zhong-Jian, Du, Lei, Cai, Bo-Ning, & Yang, Jack. Dosimetric evaluation of 4 different treatment modalities for curative-intent stereotactic body radiation therapy for isolated thoracic spinal metastases. United States. doi:10.1016/J.MEDDOS.2015.10.003.
Yang, Jun, Department of Oncology, First Affiliated Hospital of Xinxiang Medical University, 88 Jiankang Road, Weihui, Henan, 453100, Ma, Lin, Department of Radiation Oncology, Hainan Branch of Chinese PLA General Hospital, Haitang Bay, Sanya, 572000, Wang, Xiao-Shen, Xu, Wei Xu, Cong, Xiao-Hu, Xu, Shou-Ping, Ju, Zhong-Jian, Du, Lei, Cai, Bo-Ning, and Yang, Jack. 2016. "Dosimetric evaluation of 4 different treatment modalities for curative-intent stereotactic body radiation therapy for isolated thoracic spinal metastases". United States. doi:10.1016/J.MEDDOS.2015.10.003.
@article{osti_22577869,
title = {Dosimetric evaluation of 4 different treatment modalities for curative-intent stereotactic body radiation therapy for isolated thoracic spinal metastases},
author = {Yang, Jun and Department of Oncology, First Affiliated Hospital of Xinxiang Medical University, 88 Jiankang Road, Weihui, Henan, 453100 and Ma, Lin and Department of Radiation Oncology, Hainan Branch of Chinese PLA General Hospital, Haitang Bay, Sanya, 572000 and Wang, Xiao-Shen and Xu, Wei Xu and Cong, Xiao-Hu and Xu, Shou-Ping and Ju, Zhong-Jian and Du, Lei and Cai, Bo-Ning and Yang, Jack},
abstractNote = {To investigate the dosimetric characteristics of 4 SBRT-capable dose delivery systems, CyberKnife (CK), Helical TomoTherapy (HT), Volumetric Modulated Arc Therapy (VMAT) by Varian RapidArc (RA), and segmental step-and-shoot intensity-modulated radiation therapy (IMRT) by Elekta, on isolated thoracic spinal lesions. CK, HT, RA, and IMRT planning were performed simultaneously for 10 randomly selected patients with 6 body types and 6 body + pedicle types with isolated thoracic lesions. The prescription was set with curative intent and dose of either 33 Gy in 3 fractions (3F) or 40 Gy in 5F to cover at least 90% of the planning target volume (PTV), correspondingly. Different dosimetric indices, beam-on time, and monitor units (MUs) were evaluated to compare the advantages/disadvantages of each delivery modality. In ensuring the dose-volume constraints for cord and esophagus of the premise, CK, HT, and RA all achieved a sharp conformity index (CI) and a small penumbra volume compared to IMRT. RA achieved a CI comparable to those from CK, HT, and IMRT. CK had a heterogeneous dose distribution in the target as its radiosurgical nature with less dose uniformity inside the target. CK had the longest beam-on time and the largest MUs, followed by HT and RA. IMRT presented the shortest beam-on time and the least MUs delivery. For the body-type lesions, CK, HT, and RA satisfied the target coverage criterion in 6 cases, but the criterion was satisfied in only 3 (50%) cases with the IMRT technique. For the body + pedicle-type lesions, HT satisfied the criterion of the target coverage of ≥90% in 4 of the 6 cases, and reached a target coverage of 89.0% in another case. However, the criterion of the target coverage of ≥90% was reached in 2 cases by CK and RA, and only in 1 case by IMRT. For curative-intent SBRT of isolated thoracic spinal lesions, RA is the first choice for the body-type lesions owing to its delivery efficiency (time); the second choice is CK or HT; HT is the preferential choice for the body + pedicle-type lesions. This study suggests further clinical investigations with longer follow-up for these studied cases.},
doi = {10.1016/J.MEDDOS.2015.10.003},
journal = {Medical Dosimetry},
number = 2,
volume = 41,
place = {United States},
year = 2016,
month = 7
}
  • Purpose: SBRT allows the delivery of high dose radiation treatments to localized tumors while minimizing dose to surrounding tissues. Due to the large doses delivered, accurate contouring of organs at risk is essential. In this study, differences between the true spinal cord as seen using MRI and CT myelogram (CTM) have been assessed in patients with spinal metastases treated using SBRT. Methods: Ten patients were identified that have both a CTM and a MRI. Using rigid registration tools, the MRI was fused to the CTM. The thecal sac and true cord were contoured using each imaging modality. Images were exportedmore » and analyzed for similarity by computing the Dice similarity coefficient and the modified Hausdorff distance (greatest distance from a point in one set to the closest point in the other set). Results: The Dice coefficient was calculated for the thecal sac (0.81 ±0.06) and true cord (0.63 ±0.13). These two measures are correlated; however, some points show a low true cord overlap despite a high overlap for the thecal sac. The Hausdorff distance for structure comparisons was also calculated. For thecal sac structures, the average value, 1.6mm (±1.1), indicates good overlap. For true cord comparison, the average value, 0.3mm (±0.16), indicates very good overlap. The minimum Hausdorff distance between the true cord and thecal sac was on average 1.6mm (±0.9) Conclusion: The true cord position as seen in MRI and CTM is fairly constant, although care should be taken as large differences can be seen in individual patients. Avoidning the true cord in spine SBRT is critical, so the ability to visualize the true cord before performing SBRT to the vertebrae is essential. Here, CT myelogram appears an excellent, robust option, that can be obtained the day of treatment planning and is unaffected by uncertainties in image fusion.« less
  • Purpose: To assess motion of the spinal cord and cauda equina, which are critical neural tissues (CNT), which is important when evaluating the planning organ-at-risk margin required for stereotactic body radiation therapy. Methods and Materials: We analyzed CNT motion in 65 patients with spinal metastases (11 cervical, 39 thoracic, and 24 lumbar spinal segments) in the supine position using dynamic axial and sagittal magnetic resonance imaging (dMRI, 3T Verio, Siemens) over a 137-second interval. Motion was segregated according to physiologic cardiorespiratory oscillatory motion (characterized by the average root mean square deviation) and random bulk shifts associated with gross patient motionmore » (characterized by the range). Displacement was evaluated in the anteroposterior (AP), lateral (LR), and superior-inferior (SI) directions by use of a correlation coefficient template matching algorithm, with quantification of random motion measure error over 3 separate trials. Statistical significance was defined according to P<.05. Results: In the AP, LR, and SI directions, significant oscillatory motion was observed in 39.2%, 35.1%, and 10.8% of spinal segments, respectively, and significant bulk motions in all cases. The median oscillatory CNT motions in the AP, LR, and SI directions were 0.16 mm, 0.17 mm, and 0.44 mm, respectively, and the maximal statistically significant oscillatory motions were 0.39 mm, 0.41 mm, and 0.77 mm, respectively. The median bulk displacements in the AP, LR, and SI directions were 0.51 mm, 0.59 mm, and 0.66 mm, and the maximal statistically significant displacements were 2.21 mm, 2.87 mm, and 3.90 mm, respectively. In the AP, LR, and SI directions, bulk displacements were greater than 1.5 mm in 5.4%, 9.0%, and 14.9% of spinal segments, respectively. No significant differences in axial motion were observed according to cord level or cauda equina. Conclusions: Oscillatory CNT motion was observed to be relatively minor. Our results support the importance of controlling bulk patient motion and the practice of applying a planning organ-at-risk margin.« less
  • Stereotactic body radiation therapy (SBRT) treatments have high-dose gradients and even slight patient misalignment from the simulation to treatment could lead to target underdosing or organ at risk (OAR) overdosing. Daily real-time SBRT treatment planning could minimize the risk of geographic miss. As an initial step toward determining the clinical feasibility of developing real-time SBRT treatment planning, we determined the calculation time of helical TomoTherapy-based STAT radiation therapy (RT) treatment plans for simple liver, lung, and spine SBRT treatments to assess whether the planning process was fast enough for practical clinical implementation. Representative SBRT planning target volumes for hypothetical liver,more » peripheral lung, and thoracic spine lesions and adjacent OARs were contoured onto a planning computed tomography scan (CT) of an anthropomorphic phantom. Treatment plans were generated using both STAT RT 'full scatter' and conventional helical TomoTherapy 'beamlet' algorithms. Optimized plans were compared with respect to conformality index (CI), heterogeneity index (HI), and maximum dose to regional OARs to determine clinical equivalence and the number of required STAT RT optimization iterations and calculation times were determined. The liver and lung dosimetry for the STAT RT and standard planning algorithms were clinically and statistically equivalent. For the liver lesions, 'full scatter' and 'beamlet' algorithms showed a CI of 1.04 and 1.04 and HI of 1.03 and 1.03, respectively. For the lung lesions, 'full scatter' and 'beamlet' algorithms showed a CI of 1.05 and 1.03 and HI of 1.05and 1.05, respectively. For spine lesions, 'full scatter' and 'beamlet' algorithms showed a CI of 1.15 and 1.14 and HI of 1.22 and 1.14, respectively. There was no difference between treatment algorithms with respect to maximum doses to the OARs. The STAT RT iteration time with current treatment planning systems is 45 sec, and the treatment planning required 3 iterations or 135 sec for STAT RT liver and lung SBRT plans and 7 iterations or 315 sec for STAT RT spine SBRT plans. Helical TomoTherapy-based STAT RT treatment planning with the 'full scatter' algorithm provides levels of dosimetric conformality, heterogeneity, and OAR avoidance for SBRT treatments that are clinically equivalent to those generated with the Helical TomoTherapy 'beamlet' algorithm. STAT RT calculation times for simple SBRT treatments are fast enough to warrant further investigation into their potential incorporation into an SBRT program with daily real-time planning. Development of methods for accurate target and OAR determination on megavoltage computed tomography scans incorporating high-resolution diagnostic image co-registration software and CT detector-based exit dose measurement for quality assurance are necessary to build a real-time SBRT planning and delivery program.« less
  • Early experience with stereotactic body radiation therapy (SBRT) of centrally located lung tumors indicated increased rate of high-grade toxicity in the lungs. These clinical results were based on treatment plans that were computed using pencil beam–like algorithms and without tissue inhomogeneity corrections. In this study, we evaluated the dosimetric errors in plans with and without inhomogeneity corrections and with planning target volumes (PTVs) that were within the zone of the proximal bronchial tree (BT). For 10 patients, the PTV, lungs, and sections of the BT either inside or within 2 cm of the PTV were delineated. Two treatment plans weremore » generated for each patient using the following dose-calculation methods: (1) pencil beam (PB) algorithm without inhomogeneity correction (IC) (PB − IC) and (2) PB with inhomogeneity correction (PB + IC). Both plans had identical beam geometry but different beam segment shapes and monitor units (MU) to achieve similar conformal dose coverage of PTV. To obtain the baseline dose distributions, each plan was recalculated using a Monte Carlo (MC) algorithm by keeping MUs the same in the respective plans. The median maximum dose to the proximal BT and PTV dose coverage in the PB + IC plans were overestimated by 8% and 11%, respectively. However, the median maximum dose to the proximal BT and PTV dose coverage in PB − IC plans were underestimated by 15% and 9%. Similar trends were observed in low-dose regions of the lung within the irradiated volume. Our study indicates that dosimetric bias introduced by unit tissue density plans cannot be characterized as underestimation or overestimation of dose without taking the tumor location into account. This issue should be considered when analyzing clinical toxicity data from early lung SBRT trials that utilized unit tissue density for dose calculations.« less
  • Purpose: To provide actuarial outcomes and dosimetric data for spinal/paraspinal metastases, with and without prior radiation, treated with stereotactic body radiotherapy (SBRT). Methods and Materials: A total of 39 consecutive patients (60 metastases) were treated with SBRT between April 2003 and August 2006 and retrospectively reviewed. In all, 23 of 60 tumors had no previous radiation (unirradiated) and 37/60 tumors had previous irradiation (reirradiated). Of 37 reirradiated tumors, 31 were treated for 'salvage' given image-based tumor progression. Local failure was defined as progression by imaging and/or clinically. Results: At last follow-up, 19 patients were deceased. Median patient survival time measuredmore » was 21 months (95% CI = 8-27 months), and the 2-year survival probability was 45%. The median total dose prescribed was 24 Gy in three fractions prescribed to the 67% and 60% isodose for the unirradiated and reirradiated cohorts, respectively. The median tumor follow-up for the unirradiated and reirradiated group was 9 months (range, 1-26) and 7 months (range, 1-48) respectively. Eight of 60 tumors have progressed, and the 1- and 2-year progression-free probability (PFP) was 85% and 69%, respectively. For the salvage group the 1 year PFP was 96%. There was no significant difference in overall survival or PFP between the salvage reirradiated vs. all other tumors treated (p = 0.08 and p = 0.31, respectively). In six of eight failures the minimum distance from the tumor to the thecal sac was {<=}1 mm. Of 60 tumors treated, 39 have {>=}6 months follow-up and no radiation-induced myelopathy or radiculopathy has occurred. Conclusion: Spine SBRT has shown preliminary efficacy and safety in patients with image-based progression of previously irradiated metastases.« less