Title: SU-F-BRA-08: An Investigation of Well-Chamber Responses for An Electronic Brachytherapy Source

Abstract

Purpose: The aim of this study was to investigate the variation of well-type ionization chamber response between a Xoft Axxent™ electronic brachytherapy (EBT) source and a GE Oncoseed™ 6711 I-125 seed. Methods: A new EBT air-kerma standard has recently been introduced by the National Institute of Standards and Technology (NIST). Historically, the Axxent source strength has been based on a well chamber calibration from an I-125 brachytherapy source due to the lack of a primary standard. Xoft utilizes a calibration procedure that employs a GE 6711 seed calibration as a surrogate standard to represent the air-kerma strength of an Axxent source. This method is based on the premise that the energies of the two sources are similar and thus, a conversion factor would be a suitable interim solution until a NIST standard was established. For this investigation, a number of well chambers of the same model type and three different EBT sources were used to determine NIST-traceable calibration coefficients for both the GE 6711 seed and the Axxent source. The ratio of the two coefficients was analyzed for consistency and also to identify any possible correlations with chamber vintage or the sources themselves. Results: For all well chambers studied, themore » relative standard deviation of the ratio of calibration coefficients between the two standards is less than 1%. No specific trends were found with the well chamber vintage or between the three different EBT sources used. Conclusion: The variation of well chamber calibration coefficients between a Xoft Axxent™ EBT source versus a GE 6711 Oncoseed™ are consistent across well chamber vintage and between sources. The results of this investigation confirm the underlying assumptions and stability of the surrogate standard currently in use by Xoft, and establishes a migration path for future implementation of the new NIST air kerma standard. This research is supported in part by Xoft, a subsidiary of iCAD.« less

Authors:
;  [1]
  1. University of Wisconsin - Madison, Madison, WI (United States)
Publication Date:
OSTI Identifier:
22555214
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; BRACHYTHERAPY; CALIBRATION STANDARDS; IODINE 125; IONIZATION CHAMBERS; KERMA; RADIATION SOURCE IMPLANTS

Citation Formats

Culberson, W, and Micka, J. SU-F-BRA-08: An Investigation of Well-Chamber Responses for An Electronic Brachytherapy Source. United States: N. p., 2015. Web. doi:10.1118/1.4925219.
Culberson, W, & Micka, J. SU-F-BRA-08: An Investigation of Well-Chamber Responses for An Electronic Brachytherapy Source. United States. https://doi.org/10.1118/1.4925219
Culberson, W, and Micka, J. Mon . "SU-F-BRA-08: An Investigation of Well-Chamber Responses for An Electronic Brachytherapy Source". United States. https://doi.org/10.1118/1.4925219.
@article{osti_22555214,
title = {SU-F-BRA-08: An Investigation of Well-Chamber Responses for An Electronic Brachytherapy Source},
author = {Culberson, W and Micka, J},
abstractNote = {Purpose: The aim of this study was to investigate the variation of well-type ionization chamber response between a Xoft Axxent™ electronic brachytherapy (EBT) source and a GE Oncoseed™ 6711 I-125 seed. Methods: A new EBT air-kerma standard has recently been introduced by the National Institute of Standards and Technology (NIST). Historically, the Axxent source strength has been based on a well chamber calibration from an I-125 brachytherapy source due to the lack of a primary standard. Xoft utilizes a calibration procedure that employs a GE 6711 seed calibration as a surrogate standard to represent the air-kerma strength of an Axxent source. This method is based on the premise that the energies of the two sources are similar and thus, a conversion factor would be a suitable interim solution until a NIST standard was established. For this investigation, a number of well chambers of the same model type and three different EBT sources were used to determine NIST-traceable calibration coefficients for both the GE 6711 seed and the Axxent source. The ratio of the two coefficients was analyzed for consistency and also to identify any possible correlations with chamber vintage or the sources themselves. Results: For all well chambers studied, the relative standard deviation of the ratio of calibration coefficients between the two standards is less than 1%. No specific trends were found with the well chamber vintage or between the three different EBT sources used. Conclusion: The variation of well chamber calibration coefficients between a Xoft Axxent™ EBT source versus a GE 6711 Oncoseed™ are consistent across well chamber vintage and between sources. The results of this investigation confirm the underlying assumptions and stability of the surrogate standard currently in use by Xoft, and establishes a migration path for future implementation of the new NIST air kerma standard. This research is supported in part by Xoft, a subsidiary of iCAD.},
doi = {10.1118/1.4925219},
url = {https://www.osti.gov/biblio/22555214}, journal = {Medical Physics},
issn = {0094-2405},
number = 6,
volume = 42,
place = {United States},
year = {2015},
month = {6}
}