Title: SU-F-BRA-07: Dosimetric Verification of the Valencia Skin Applicator Using Gafchromic EBT3 Film

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4925218· OSTI ID:22555213
; ; ; ;  [1]
  1. Christiana Care Hospital, Newark, DE (United States)

Purpose: The Valencia applicators have recently been introduced for HDR treatment of small and shallow superficial skin lesions (< 20 mm diameter and 3-mm depth). Per AAPM TG 56, any HDR applicator internal dimensions must be verified prior to clinical use. However radiographic and tomographic imaging to validate the Valencia applicators is impractical due to the Tungsten alloy housing and flattening filter. In this study, we propose to use EBT3 film to indirectly confirm the physical integrity of the Valencia applicators. Methods: Treatment plans were created using the Oncentra MasterPlan TPS v4.5 for the H2 (20-mm dia.) and H3 (30-mm dia.) Valencia Applicators. A virtual CT phantom (2-mm slice thickness) was created with one source position in water. The published effective depth method was used for each applicator to delivery 500 cGy to a 3-mm depth using the TG-43 formalism. Film measurements (n=3) at 3-mm depth and vertical plane in solid water were performed for each applicator to verify the prescribed dose calculated by the TPS. Percent depth dose curves and off-axis profiles (phantom surface and 3-mm depth) were measured and compared to published data. Films were analyzed using an in-house written program and RIT113 v6 software. Film calibration was performed per TG-55 protocol using the Ir-192 source with NIST-traceable calibration. Results: The prescription absolute dose difference was 1% for the Valencia H2 applicator and 4% for the Valencia H3 applicator. The measured percent depth dose curves and off-axis dose profiles measured for the H2 and H2 Valencia applicators are in excellent agreement with the Granero et al. Monte Carlo results{sup 1}. Conclusion: Gafchromic EBT3 film can be used to indirectly verify the internal components of special HDR skin applicators constructed from high Z materials.{sup 1}Granero et al. “Design and evaluation of a HDR skin applicator with flattening filter”, Med. Phys. 35(2), 495–503, 2008.

OSTI ID:
22555213
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English