Title: SU-E-T-454: Impact of Air Gap On PDDs of 6 MV Photon Beam for Various Field Sizes in Inhomogeneous Phantoms

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924816· OSTI ID:22548497
 [1];  [2];  [3];  [4];  [5]
  1. Cancer Centers of Southwest Oklahoma, Lawton, OK (United States)
  2. 21st Century Oncology, Estero, FL (United States)
  3. Procure Proton Therapy Center, Oklahoma City, OK (United States)
  4. University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma (United States)
  5. ProCure Proton Therapy Center, Oklahoma City, OK (United States)

Purpose: To investigate how the shape of air gap and its size will impact the percent depth doses (PDDs) of a 6MV photon beam for various field sizes. Methods: Two in-house phantoms were manufactured containing rectangular (phantom A) and circular (phantom B) air gaps. Both phantoms A and B were composed of same top layer (solid-water; 30×30cm{sup 2},5cm thickness) and bottom layer (solid-water; 30×30cm {sup 2},8cm thickness), but middle layer was varied to observe air gap effects and scatter contribution to the measurement point. In phantom A, a rectangular shaped air gap (30×30cm{sup 2},7cm thickness) was created by placing Styrofoam blocks between top and bottom layers of the phantom. In phantom B, middle layer was replaced by “inhomogenous block”, composed of acrylic plate (30×30cm{sup 2}, 4cm thickness) followed by PVC(30×30cm{sup 2},3cm thickness). Additionally, circular air gap was created by drilling a hole (diameter=2.8cm, length=7cm) at the center of “inhomogenous block”. In both phantoms, measurement readings were obtained at 13cm depth (i.e., 1cm after air gap) and depth of maximum dose(6MV energy; 100 MUs; field sizes ranged from 3×3cm{sup 2} to 10×10cm{sup 2}). The PDDs at 13cm depth were compared in both phantoms. Results: The measurements in both phantoms A and B showed an almost linear increase in PDDs with increasing field size, especially for smaller field sizes (from 3×3 to 7×7cm{sup 2}). For each field size, the PDD in phantom A was smaller compared to the one in phantom B. The difference in PDDs between two phantoms decreased with an increase in field size as the PDD difference decreased from 9.0% to 6.4%. Conclusion: The shape and size of air gap affect the PDD measurements in secondary build-up region as 6 MV primary beam traverses through the center of air gap. The scatter contribution due to increase in field size was more noticeable for field sizes ≤7×7cm{sup 2}.

OSTI ID:
22548497
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English