skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-432: Field Size Influence On the Electron and Photon Spectra Within Small MV Field Detectors

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924793· OSTI ID:22548477
;  [1]
  1. Department of Medical Physics, Karolinska University Hospital, Stockholm and Department of Physics, Medical Radiation Physics, Stockholm University (Sweden)

Purpose: To investigate the influence of photon field size on the electron and photon fluence spectra in the active volume of small field detectors. Methods: The PENELOPE MC system based usercode PenEasy was used to calculate the material influence on the spectra by scoring the differential fluence in inserts of silicon, carbon, phosphorus and aluminium having 3 mm diameter and height. The spectra were then calculated inside the active volume of eleven detectors (ion chambers and solid-state detectors) whose geometry was simulated with great detail. The inserts/detectors were placed at 10 cm depth in a 30 cm x 30 cm x 30 cm water phantom and irradiated with 2.5 MeV photons and Varian Clinac 6 MV beams of small, medium and large size. Results: For all configurations, photon spectra in the scoring volume were similar to that in a small water volume except for additional characteristic x-ray peaks resulting from the material itself and from the materials surrounding the detectors (i.e. high-Z shielding the silicon). Electron fluence calculated in the inserts were up to 60% larger than in water; the difference increased with material density and decreasing field size. MC-calculated doses were compared to analytically determined collision kerma and restricted cema (cut-off=15keV). For the inserts, with large and medium fields K-col agreed with MC-dose, but K-col overestimated the dose for small fields due to lack of lateral CPE. For the detectors, up to 15% differences between K-col and the MC-dose were found. For all configurations the C-delta and MC-dose agreed within ±2%. Conclusion: The most relevant findings were that shielding affects substantially the photon spectra and material conditions the electron spectra, their field size dependence varying with the geometry configuration. These affect the values of factors entering into relative dosimetry.

OSTI ID:
22548477
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English