Title: SU-E-T-319: Dosimetric Evaluation of IMRT with Mix-Energy Beam for Deep Seated Targets

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924680· OSTI ID:22548374

Purpose: IMRT is preferred in the range of 6–10MV X-rays. Partially adding high energy (>10MV) treatment fields, may provide advantage of both higher and lower energies. To study IMRT dose distribution obtained from treatment plans with single (6MV) and mixed-energy (6MV and 15MV) for deep seated targets (separation more than 30cm). Methods: Five patients of carcinoma of cervix were studied using eclipse planning system. Two different dynamic IMRT plans were generated for Varian CL2300C/D linear accelerator; one is by using 6MV X-ray with seven equally spaced coplanar beams. In second plan, 2 lateral oblique fields (gantry angle 102°, 255°) beam energy was modified to 15MV by keeping all other parameters and dose volume constraints constant. Dose prescription for the planning target volume (PTV) was (5040cGy/28f). For plan comparison, dose volume histogram (DVH) was used and PTV coverage index (CI=Target volume covered by prescription dose/Target volume), heterogeneity index (D5/D95), mean dose to organ at risk (OAR) and normal tissue integral dose (NTID, liter-Gray) was also noted. Total monitor unit (MU) required to deliver a plan was also noted. Results: Mixed-energy plan showed a better conformity and CI values were 0.942±0.032 and 0.960±0.040 for single-energy and mixed-energy plan, respectively. In addition, HI value of mixed energy beam is comparable to that of single energy and the values were within 1.084±0.034 and 1.082±0.032 for single energy and mixed-energy plan, respectively. Variation in mean dose to bladder, rectum and bowel were within 1.05%, 0.87% and 0.90%. NTID was lesser for mixed-energy beam due to use of two high-energy fields. NTID were 1573.40±214.60 and 1510.20±249.80 litre-Gray for single energy and mixed-energy plan. MU needed to deliver a plan was similar in both plans and MUs were 238±45 and 237±47. Conclusion: Partial use of 15MV treatment fields in IMRT plan for deep seated targets showed dosimetric advantage over 6MV plan.

OSTI ID:
22548374
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English