SU-E-T-137: Attenuation of Carbon Fiber IGRT Couch Top for SBRT By Using a MapCheck 2 and Ion Chamber
- MedStar Health RadAmerica, Mercy Radiation Oncology, Baltimore, MD (United States)
- Mercy Medical Center Radiation Oncology, Baltimore, MD (United States)
Purpose: It is common to use posterior oblique beams in intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) for regular and stereotactic body radiotherapy (SBRT). Beam attenuation by the treatment couch is not negligible when the couch is in the beam portal. In this study, we have correlated relative dose versus beam angle through a MapCheck 2TM diode array and an ionization chamber measurement. Methods: A Varian TrueBeam linear accelerator equipped with the image-guided radiation therapy (IGRT) carbon fiber couch was used for delivery of radiation with different photon energies of 6MV, 15MV and 6MV FFF. A MapCHECK 2TM diode array with a MapPhan was employed for the measurement at three field sizes (3 by 3, 5 by 5, and 10 by 10 cm). The independent measurement was performed with an ionization chamber placed at the center of an acrylic cylindrical phantom at the isocenter. The measured data with the MapCHECK 2TM diode array were averaged over the central 2×2 cm diodes for each measurement. The couch attenuation was deduced from the angular dependence with and without the couch involvement. Results: For the MapCHECK 2TM with MapPhan used only the obvious attenuation was observed at the beam angle between 5 and 10 degree from the angular dependence using a 6MV photon beam. The similar Result was obtained for the MapCHECK 2TM /MapPhan with the couch involvement. The couch attenuation was then deduced from the difference of these two sets of data measurements. Maximum couch attenuation was found up to 4.2% which was verified with the ion chamber measurement. Conclusion: A maximum attenuation of 4.2% was found for the carbon fiber couch top using the MapCHECK 2TM/MapPhan, which is consistent with the ion chamber measurement.
- OSTI ID:
- 22545263
- Journal Information:
- Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Evaluating dosimetric accuracy of flattening filter free compensator-based IMRT: Measurements with diode arrays
SU-E-T-315: Dosimetric Effects of Couch Top Shift On VMAT Delivery in Absence of Indexing