Purpose: To build a fast, accurate and easily-deployable research platform for Monte-Carlo dose calculations. We port the dose calculation engine PENELOPE to C++, and accelerate calculations using GPU acceleration. Simulations of a Co-60 beam model provided by ViewRay demonstrate the capabilities of the platform. Methods: We built software that incorporates a beam model interface, CT-phantom model, GPU-accelerated PENELOPE engine, and GUI front-end. We rewrote the PENELOPE kernel in C++ (from Fortran) and accelerated the code on a GPU. We seamlessly integrated a Co-60 beam model (obtained from ViewRay) into our platform. Simulations of various field sizes and SSDs using a homogeneous water phantom generated PDDs, dose profiles, and output factors that were compared to experiment data. Results: With GPU acceleration using a dated graphics card (Nvidia Tesla C2050), a highly accurate simulation – including 100*100*100 grid, 3×3×3 mm3 voxels, <1% uncertainty, and 4.2×4.2 cm2 field size – runs 24 times faster (20 minutes versus 8 hours) than when parallelizing on 8 threads across a new CPU (Intel i7-4770). Simulated PDDs, profiles and output ratios for the commercial system agree well with experiment data measured using radiographic film or ionization chamber. Based on our analysis, this beam model is precise enoughmore » for general applications. Conclusions: Using a beam model for a Co-60 system provided by ViewRay, we evaluate a dose calculation platform that we developed. Comparison to measurements demonstrates the promise of our software for use as a research platform for dose calculations, with applications including quality assurance and treatment plan verification.« less

Authors:
; ; ; ; ; ; ; ;  [1]
  1. Washington University School of Medicine, St. Louis, MO (United States)
Publication Date:
OSTI Identifier:
22545169
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BEAMS; COBALT 60; COMPUTER CODES; COMPUTERIZED TOMOGRAPHY; IONIZATION CHAMBERS; KERNELS; MONTE CARLO METHOD; PHANTOMS; QUALITY ASSURANCE; RADIATION DOSES; SIMULATION; VERIFICATION

Citation Formats

Wang, Y, Mazur, T, Green, O, Hu, Y, Wooten, H, Yang, D, Zhao, T, Mutic, S, and Li, H. SU-E-T-36: A GPU-Accelerated Monte-Carlo Dose Calculation Platform and Its Application Toward Validating a ViewRay Beam Model. United States: N. p., 2015. Web. doi:10.1118/1.4924397.
Wang, Y, Mazur, T, Green, O, Hu, Y, Wooten, H, Yang, D, Zhao, T, Mutic, S, & Li, H. SU-E-T-36: A GPU-Accelerated Monte-Carlo Dose Calculation Platform and Its Application Toward Validating a ViewRay Beam Model. United States. doi:10.1118/1.4924397.
Wang, Y, Mazur, T, Green, O, Hu, Y, Wooten, H, Yang, D, Zhao, T, Mutic, S, and Li, H. Mon . "SU-E-T-36: A GPU-Accelerated Monte-Carlo Dose Calculation Platform and Its Application Toward Validating a ViewRay Beam Model". United States. doi:10.1118/1.4924397.
@article{osti_22545169,
title = {SU-E-T-36: A GPU-Accelerated Monte-Carlo Dose Calculation Platform and Its Application Toward Validating a ViewRay Beam Model},
author = {Wang, Y and Mazur, T and Green, O and Hu, Y and Wooten, H and Yang, D and Zhao, T and Mutic, S and Li, H},
abstractNote = {Purpose: To build a fast, accurate and easily-deployable research platform for Monte-Carlo dose calculations. We port the dose calculation engine PENELOPE to C++, and accelerate calculations using GPU acceleration. Simulations of a Co-60 beam model provided by ViewRay demonstrate the capabilities of the platform. Methods: We built software that incorporates a beam model interface, CT-phantom model, GPU-accelerated PENELOPE engine, and GUI front-end. We rewrote the PENELOPE kernel in C++ (from Fortran) and accelerated the code on a GPU. We seamlessly integrated a Co-60 beam model (obtained from ViewRay) into our platform. Simulations of various field sizes and SSDs using a homogeneous water phantom generated PDDs, dose profiles, and output factors that were compared to experiment data. Results: With GPU acceleration using a dated graphics card (Nvidia Tesla C2050), a highly accurate simulation – including 100*100*100 grid, 3×3×3 mm3 voxels, <1% uncertainty, and 4.2×4.2 cm2 field size – runs 24 times faster (20 minutes versus 8 hours) than when parallelizing on 8 threads across a new CPU (Intel i7-4770). Simulated PDDs, profiles and output ratios for the commercial system agree well with experiment data measured using radiographic film or ionization chamber. Based on our analysis, this beam model is precise enough for general applications. Conclusions: Using a beam model for a Co-60 system provided by ViewRay, we evaluate a dose calculation platform that we developed. Comparison to measurements demonstrates the promise of our software for use as a research platform for dose calculations, with applications including quality assurance and treatment plan verification.},
doi = {10.1118/1.4924397},
journal = {Medical Physics},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}
  • Purpose: The clinical commissioning of IMRT subject to a magnetic field is challenging. The purpose of this work is to develop a GPU-accelerated Monte Carlo dose calculation platform based on PENELOPE and then use the platform to validate a vendor-provided MRIdian head model toward quality assurance of clinical IMRT treatment plans subject to a 0.35 T magnetic field. Methods: PENELOPE was first translated from FORTRAN to C++ and the result was confirmed to produce equivalent results to the original code. The C++ code was then adapted to CUDA in a workflow optimized for GPU architecture. The original code was expandedmore » to include voxelized transport with Woodcock tracking, faster electron/positron propagation in a magnetic field, and several features that make gPENELOPE highly user-friendly. Moreover, the vendor-provided MRIdian head model was incorporated into the code in an effort to apply gPENELOPE as both an accurate and rapid dose validation system. A set of experimental measurements were performed on the MRIdian system to examine the accuracy of both the head model and gPENELOPE. Ultimately, gPENELOPE was applied toward independent validation of patient doses calculated by MRIdian’s KMC. Results: An acceleration factor of 152 was achieved in comparison to the original single-thread FORTRAN implementation with the original accuracy being preserved. For 16 treatment plans including stomach (4), lung (2), liver (3), adrenal gland (2), pancreas (2), spleen(1), mediastinum (1), and breast (1), the MRIdian dose calculation engine agrees with gPENELOPE with a mean gamma passing rate of 99.1% ± 0.6% (2%/2 mm). Conclusions: A Monte Carlo simulation platform was developed based on a GPU- accelerated version of PENELOPE. This platform was used to validate that both the vendor-provided head model and fast Monte Carlo engine used by the MRIdian system are accurate in modeling radiation transport in a patient using 2%/2 mm gamma criteria. Future applications of this platform will include dose validation and accumulation, IMRT optimization, and dosimetry system modeling for next generation MR-IGRT systems.« less
  • Purpose: A new commercial GPU-based Monte Carlo dose calculation algorithm (GPUMCD) developed by the vendor Elekta™ to be used in the Monaco Treatment Planning System (TPS) is capable of modeling dose for both a standard linear accelerator and for an Elekta MRI-Linear accelerator (modeling magnetic field effects). We are evaluating this algorithm in two parts: commissioning the algorithm for an Elekta Agility linear accelerator (the focus of this work) and evaluating the algorithm’s ability to model magnetic field effects for an MRI-linear accelerator. Methods: A beam model was developed in the Monaco TPS (v.5.09.06) using the commissioned beam data formore » a 6MV Agility linac. A heterogeneous phantom representing tumor-in-lung, lung, bone-in-tissue, and prosthetic was designed/built. Dose calculations in Monaco were done using the current clinical algorithm (XVMC) and the new GPUMCD algorithm (1 mm3 voxel size, 0.5% statistical uncertainty) and in the Pinnacle TPS using the collapsed cone convolution (CCC) algorithm. These were compared with the measured doses using an ionization chamber (A1SL) and Gafchromic EBT3 films for 2×2 cm{sup 2}, 5×5 cm{sup 2}, and 10×10 cm{sup 2} field sizes. Results: The calculated central axis percentage depth doses (PDDs) in homogeneous solid water were within 2% compared to measurements for XVMC and GPUMCD. For tumor-in-lung and lung phantoms, doses calculated by all of the algorithms were within the experimental uncertainty of the measurements (±2% in the homogeneous phantom and ±3% for the tumor-in-lung or lung phantoms), except for 2×2 cm{sup 2} field size where only the CCC algorithm differs from film by 5% in the lung region. The analysis for bone-in-tissue and the prosthetic phantoms are ongoing. Conclusion: The new GPUMCD algorithm calculated dose comparable to both the XVMC algorithm and to measurements in both a homogeneous solid water medium and the heterogeneous phantom representing lung or tumor-in-lung for 2×2 cm{sup 2}-10×10 cm{sup 2} field sizes. Funding support was obtained from Elekta.« less
  • Purpose: To enable an existing web application for GPU-based Monte Carlo (MC) 3D dosimetry quality assurance (QA) to compute “delivered dose” from linac logfile data. Methods: We added significant features to an IMRT/VMAT QA web application which is based on existing technologies (HTML5, Python, and Django). This tool interfaces with python, c-code libraries, and command line-based GPU applications to perform a MC-based IMRT/VMAT QA. The web app automates many complicated aspects of interfacing clinical DICOM and logfile data with cutting-edge GPU software to run a MC dose calculation. The resultant web app is powerful, easy to use, and is ablemore » to re-compute both plan dose (from DICOM data) and delivered dose (from logfile data). Both dynalog and trajectorylog file formats are supported. Users upload zipped DICOM RP, CT, and RD data and set the expected statistic uncertainty for the MC dose calculation. A 3D gamma index map, 3D dose distribution, gamma histogram, dosimetric statistics, and DVH curves are displayed to the user. Additional the user may upload the delivery logfile data from the linac to compute a 'delivered dose' calculation and corresponding gamma tests. A comprehensive PDF QA report summarizing the results can also be downloaded. Results: We successfully improved a web app for a GPU-based QA tool that consists of logfile parcing, fluence map generation, CT image processing, GPU based MC dose calculation, gamma index calculation, and DVH calculation. The result is an IMRT and VMAT QA tool that conducts an independent dose calculation for a given treatment plan and delivery log file. The system takes both DICOM data and logfile data to compute plan dose and delivered dose respectively. Conclusion: We sucessfully improved a GPU-based MC QA tool to allow for logfile dose calculation. The high efficiency and accessibility will greatly facilitate IMRT and VMAT QA.« less
  • Purpose: The current standard TG-43 dose calculation method for SAVI-based Accelerated Partial Breast Irradiation (APBI) assumes an ideal geometry of infinite homogeneous water. However, in SAVI treatments, the air cavity inside the device and the short source-to-skin distance raise concerns about the dose accuracy of the TG-43 method. This study is to evaluate TG-43 dose calculation accuracy in SAVI treatments using Monte Carlo (MC) simulations. Methods: We recalculated the dose distributions of 15 APBI patients treated with SAVI devices, including five cases with a size of 6–1, five with 8−1 and five with 10−1, using our in-house developed fast MCmore » dose package for HDR brachytherapy (gBMC). A phase-space file was used to model the Ir-192 HDR source. For each case, the patient CT was converted into a voxelized phantom and the dwell positions and times were extracted from treatment plans for MC dose calculations. Clinically relevant dosimetric parameters of the recalculated dose were compared to those computed via the TG-43 approach. Results: A systematic overestimation of doses was found for the 15 cases in TG-43 results, with D90, V150, and V200 for PTV-eval 2.8±1.8%, 2.0±2.2%, and 1.8±3.5% higher than MC results. TG-43 also overestimated the dose to skin with the maximum dose 4.4±8.4% higher on average. The relatively large standard deviation seen in the difference of maximum skin dose is partially ascribed to the statistical uncertainty of MC simulations when computing the maximum dose. It took gBMC ∼1 minute to compute dose for a SAVI plan. Conclusion: The high efficiency of our gBMC package facilitated the studies with a relatively large number of cases. An overestimation of TG-43 doses was found when using this MC package to recompute doses in SAVI cases. Clinical utilization of TG-43 dose calculation method in this scenario should be aware of this fact.« less
  • Purpose: Investigating the relative sensitivity of Monte Carlo (MC) and Pencil Beam (PB) dose calculation algorithms to low-Z (titanium) metallic artifacts is important for accurate and consistent dose reporting in post¬operative spinal RS. Methods: Sensitivity analysis of MC and PB dose calculation algorithms on the Monaco v.3.3 treatment planning system (Elekta CMS, Maryland Heights, MO, USA) was performed using CT images reconstructed without (plain) and with Orthopedic Metal Artifact Reduction (OMAR; Philips Healthcare system, Cleveland, OH, USA). 6MV and 10MV volumetric-modulated arc (VMAT) RS plans were obtained for MC and PB on the plain and OMAR images (MC-plain/OMAR and PB-plain/OMAR).more » Results: Maximum differences in dose to 0.2cc (D0.2cc) of spinal cord and cord +2mm for 6MV and 10MV VMAT plans were 0.1Gy between MC-OMAR and MC-plain, and between PB-OMAR and PB-plain. Planning target volume (PTV) dose coverage changed by 0.1±0.7% and 0.2±0.3% for 6MV and 10MV from MC-OMAR to MC-plain, and by 0.1±0.1% for both 6MV and 10 MV from PB-OMAR to PB-plain, respectively. In no case for both MC and PB the D0.2cc to spinal cord was found to exceed the planned tolerance changing from OMAR to plain CT in dose calculations. Conclusion: Dosimetric impacts of metallic artifacts caused by low-Z metallic spinal hardware (mainly titanium alloy) are not clinically important in VMAT-based spine RS, without significant dependence on dose calculation methods (MC and PB) and photon energy ≥ 6MV. There is no need to use one algorithm instead of the other to reduce uncertainty for dose reporting. The dose calculation method that should be used in spine RS shall be consistent with the usual clinical practice.« less