Title: SU-E-T-10: A Clinical Implementation and the Dosimetric Evidence in High Dose Rate Vaginal Multichannel Applicator Brachytherapy

Abstract

Purpose: The multichannel cylindrical applicator has a distinctive modification of the traditional single channel cylindrical applicator. The novel multichannel applicator has additional peripheral channels that provide more flexibility both in treatment planning process and outcomes. To protect by reducing doses to adjacent organ at risk (OAR) while maintaining target coverage with inverse plan optimization are the goals for such novel Brachytherapy device. Through a series of comparison and analysis of reults in more than forty patients who received HDR Brachytherapy using multichannel vaginal applicator, this procedure has been implemented in our institution. Methods: Multichannel planning was CT image based. The CTV of 5mm vaginal cuff rind with prescribed length was well reconstructed as well as bladder and rectum. At least D95 of CTV coverage is 95% of prescribed dose. Multichannel inverse plan optimization algorithm not only shapes target dose cloud but set dose avoids to OAR’s exclusively. The doses of D2cc, D5cc and D5; volume of V2Gy in OAR’s were selected to compare with single channel results when sole central channel is only possibility. Results: Study demonstrates plan superiorly in OAR’s doe reduction in multi-channel plan. The D2cc of the rectum and bladder were showing a little lower for multichannelmore » vs. single channel. The V2Gy of the rectum was 93.72% vs. 83.79% (p=0.007) for single channel vs. multichannel respectively. Absolute reduced mean dose of D5 by multichannel was 17 cGy (s.d.=6.4) and 44 cGy (s.d.=15.2) in bladder and rectum respectively. Conclusion: The optimization solution in multichannel was to maintain D95 CTV coverage while reducing the dose to OAR’s. Dosimetric advantage in sparing critical organs by using a multichannel applicator in HDR Brachytherapy treatment of the vaginal cuff is so promising and has been implemented clinically.« less

Authors:
; ; ; ; ;  [1]
  1. Willis-Knighton Medical Center, Shreveport, LA (United States)
Publication Date:
OSTI Identifier:
22545145
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; ALGORITHMS; BLADDER; BRACHYTHERAPY; COMPUTERIZED TOMOGRAPHY; CRITICAL ORGANS; DOSE RATES; IMAGE PROCESSING; OPTIMIZATION; RADIATION DOSES; RECTUM

Citation Formats

Syh, J, Syh, J, Patel, B, Zhang, J, Wu, H, and Rosen, L. SU-E-T-10: A Clinical Implementation and the Dosimetric Evidence in High Dose Rate Vaginal Multichannel Applicator Brachytherapy. United States: N. p., 2015. Web. doi:10.1118/1.4924371.
Syh, J, Syh, J, Patel, B, Zhang, J, Wu, H, & Rosen, L. SU-E-T-10: A Clinical Implementation and the Dosimetric Evidence in High Dose Rate Vaginal Multichannel Applicator Brachytherapy. United States. doi:10.1118/1.4924371.
Syh, J, Syh, J, Patel, B, Zhang, J, Wu, H, and Rosen, L. Mon . "SU-E-T-10: A Clinical Implementation and the Dosimetric Evidence in High Dose Rate Vaginal Multichannel Applicator Brachytherapy". United States. doi:10.1118/1.4924371.
@article{osti_22545145,
title = {SU-E-T-10: A Clinical Implementation and the Dosimetric Evidence in High Dose Rate Vaginal Multichannel Applicator Brachytherapy},
author = {Syh, J and Syh, J and Patel, B and Zhang, J and Wu, H and Rosen, L},
abstractNote = {Purpose: The multichannel cylindrical applicator has a distinctive modification of the traditional single channel cylindrical applicator. The novel multichannel applicator has additional peripheral channels that provide more flexibility both in treatment planning process and outcomes. To protect by reducing doses to adjacent organ at risk (OAR) while maintaining target coverage with inverse plan optimization are the goals for such novel Brachytherapy device. Through a series of comparison and analysis of reults in more than forty patients who received HDR Brachytherapy using multichannel vaginal applicator, this procedure has been implemented in our institution. Methods: Multichannel planning was CT image based. The CTV of 5mm vaginal cuff rind with prescribed length was well reconstructed as well as bladder and rectum. At least D95 of CTV coverage is 95% of prescribed dose. Multichannel inverse plan optimization algorithm not only shapes target dose cloud but set dose avoids to OAR’s exclusively. The doses of D2cc, D5cc and D5; volume of V2Gy in OAR’s were selected to compare with single channel results when sole central channel is only possibility. Results: Study demonstrates plan superiorly in OAR’s doe reduction in multi-channel plan. The D2cc of the rectum and bladder were showing a little lower for multichannel vs. single channel. The V2Gy of the rectum was 93.72% vs. 83.79% (p=0.007) for single channel vs. multichannel respectively. Absolute reduced mean dose of D5 by multichannel was 17 cGy (s.d.=6.4) and 44 cGy (s.d.=15.2) in bladder and rectum respectively. Conclusion: The optimization solution in multichannel was to maintain D95 CTV coverage while reducing the dose to OAR’s. Dosimetric advantage in sparing critical organs by using a multichannel applicator in HDR Brachytherapy treatment of the vaginal cuff is so promising and has been implemented clinically.},
doi = {10.1118/1.4924371},
journal = {Medical Physics},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}
  • Purpose: The multichannel cylindrical vaginal applicator is a variation of traditional single channel cylindrical vaginal applicator. The multichannel applicator has additional peripheral channels that provide more flexibility in the planning process. The dosimetric advantage is to reduce dose to adjacent organ at risk (OAR) such as bladder and rectum while maintaining target coverage with the dose optimization from additional channels. Methods: Vaginal HDR brachytherapy plans are all CT based. CT images were acquired in 2 mm thickness to keep integrity of cylinder contouring. The CTV of 5mm Rind with prescribed treatment length was reconstructed from 5mm expansion of inserted cylinder.more » The goal was 95% of CTV covered by 95% of prescribed dose in both single channel planning (SCP)and multichannel planning (MCP) before proceeding any further optimization for dose reduction to critical structures with emphasis on D2cc and V2Gy . Results: This study demonstrated noticeable dose reduction to OAR was apparent in multichannel plans. The D2cc of the rectum and bladder were showing the reduced dose for multichannel versus single channel. The V2Gy of the rectum was 93.72% and 83.79% (p=0.007) for single channel and multichannel respectively (Figure 1 and Table 1). To assure adequate coverage to target while reducing the dose to the OAR without any compromise is the main goal in using multichannel vaginal applicator in HDR brachytherapy. Conclusion: Multichannel plans were optimized using anatomical based inverse optimization algorithm of inverse planning simulation annealing. The optimization solution of the algorithm was to improve the clinical target volume dose coverage while reducing the dose to critical organs such as bladder, rectum and bowels. The comparison between SCP and MCP demonstrated MCP is superior to SCP where the dwell positions were based on geometric array only. It concluded that MCP is preferable and is able to provide certain features superior to SCP.« less
  • Purpose: To compare the dose of an in-house 3D-printed gynecology applicator (TMHGA) for vaginal vault recurrence of corpus cancer patients after operation for high dose rate brachytherapy treatment with commercially available applicators. Methods: A newly designed applicator is made from 3D-printing methods using ABSM30i. The isodose of the applicator is compared with Elekta multi-channel (MC) applicator and titanium Rotterdam applicator with coupling central tube and vaginal cylinder (RC). Three plans are created using three applicators in a CT set of water phantom. The applicators are anchored using the applicator library and implant library in the Elekta Oncentra treatment planning systemmore » (ver.4.5). The rectum is mimicked by creating a 2cm diameter cylinder, with a distance 1mm posteriorly away from the high risk CTV (HR-CTV). Similarly, the bladder is replicated by a 6cm diameter cylinder with distance 1mm anteriorly from the HR-CTV. Three plans are all normalized 1.5cm superior, 0.5cm anterior and 0.5cm posterior of the applicator surface. By fixing D90 of HR-CTV to 6Gy, the D2cc of rectum and bladder of three plans are compared. Results: The D2cc of the bladder for using TMHGA is lower than MC and RC by 14.0% and 11.9% respectively. While the D2cc of the rectum for using TMHGA is lower than MC and RC by 18.9% and 12.4% respectively. The total treatment time of TMHGA plan is shorter than MC and RC by 11.2% and 12.9%. Conclusion: The applicator created via 3D printing delivers a lower dose to the bladder and the rectum while keeping the same coverage to HR-CTV as other commercially available applicators. Additionally, the new applicator resulted in a reduction of treatment time, which is always welcome.« less
  • Purpose: As an alternative to cylindrical applicators, air inflated balloon applicators have been introduced into HDR vaginal cuff brachytherapy treatment to achieve sufficient dose to vagina mucosa as well as to spare rectum and bladder. In general, TG43 formulae based treatment planning systems do not take into account tissue inhomogeneity, and air in the balloon applicator can cause higher delivered dose to mucosa than treatment plan reported. We investigated dosimetric effect of air in balloon applicator using the Monte Carlo method. Methods: The thirteen-catheter Capri applicator with a Nucletron Ir-192 seed was modeled for various balloon diameters (2cm to 3.5cm)more » using the MCNP Monte Carlo code. Ir-192 seed was placed in both central and peripheral catheters to replicate real patient situations. Existence of charged particle equilibrium (CPE) with air balloon was evaluated by comparing kerma and dose at various distances (1mm to 70mm) from surface of air-filled applicator. Also mucosa dose by an air-filled applicator was compared with by a water-filled applicator to evaluate dosimetry accuracy of planning system without tissue inhomogeneity correction. Results: Beyond 1mm from air/tissue interface, the difference between kerma and dose was within 2%. CPE (or transient CPE) condition was deemed existent, and in this region no electron transport was necessary in Monte Carlo simulations. At 1mm or less, the deviation of dose from kerma became more apparent. Increase of dose to mucosa depended on diameter of air balloon. The increment of dose to mucosa was 2.5% and 4.3% on average for 2cm and 3.5cm applicators, respectively. Conclusion: After introduction of air balloon applicator, CPE fails only at the proximity of air/tissue interface. Although dose to mucosa is increased, there is no significant dosimetric difference (<5%) between air and water filled applicators. Tissue inhomogeneity correction is not necessary for air-filled applicators.« less
  • Purpose: This case study was designated to confirm the optimized plan was used to treat skin surface of left leg in three stages. 1. To evaluate dose distribution and plan quality by alternating of the source loading catheters pattern in flexible Freiberg Flap skin surface (FFSS) applicator. 2. To investigate any impact on Dose Volume Histogram (DVH) of large superficial surface target volume coverage. 3. To compare the dose distribution if it was treated with electron beam. Methods: The Freiburg Flap is a flexible mesh style surface mold for skin radiation or intraoperative surface treatments. The Freiburg Flap consists ofmore » multiple spheres that are attached to each other, holding and guiding up to 18 treatment catheters. The Freiburg Flap also ensures a constant distance of 5mm from the treatment catheter to the surface. Three treatment trials with individual planning optimization were employed: 18 channels, 9 channels of FF and 6 MeV electron beam. The comparisons were highlighted in target coverage, dose conformity and dose sparing of surrounding tissues. Results: The first 18 channels brachytherapy plan was generated with 18 catheters inside the skin-wrapped up flap (Figure 1A). A second 9 catheters plan was generated associated with the same calculation points which were assigned to match prescription for target coverage as 18 catheters plan (Figure 1B). The optimized inverse plan was employed to reduce the dose to adjacent structures such as tibia or fibula. The comparison of DVH’s was depicted on Figure 2. External beam of electron RT plan was depicted in Figure 3. Overcall comparisons among these three were illustrated in Conclusion: The 9-channel Freiburg flap flexible skin applicator offers a reasonably acceptable plan without compromising the coverage. Electron beam was discouraged to use to treat curved skin surface because of low target coverage and high dose in adjacent tissues.« less
  • Purpose: Acuros BV has become available to perform accurate dose calculations in high-dose-rate (HDR) brachytherapy with phantom heterogeneity considered by solving the Boltzmann transport equation. In this work, we performed validation studies regarding the dose calculation accuracy of Acuros BV in cases with a shielded cylinder applicator using Monte Carlo (MC) simulations. Methods: Fifteen cases were considered in our studies, covering five different diameters of the applicator and three different shielding degrees. For each case, a digital phantom was created in Varian BrachyVision with the cylinder applicator inserted in the middle of a large water phantom. A treatment plan withmore » eight dwell positions was generated for these fifteen cases. Dose calculations were performed with Acuros BV. We then generated a voxelized phantom of the same geometry, and the materials were modeled according to the vendor’s specifications. MC dose calculations were then performed using our in-house developed fast MC dose engine for HDR brachytherapy (gBMC) on a GPU platform, which is able to simulate both photon transport and electron transport in a voxelized geometry. A phase-space file for the Ir-192 HDR source was used as a source model for MC simulations. Results: Satisfactory agreements between the dose distributions calculated by Acuros BV and those calculated by gBMC were observed in all cases. Quantitatively, we computed point-wise dose difference within the region that receives a dose higher than 10% of the reference dose, defined to be the dose at 5mm outward away from the applicator surface. The mean dose difference was ∼0.45%–0.51% and the 95-percentile maximum difference was ∼1.24%–1.47%. Conclusion: Acuros BV is able to accurately perform dose calculations in HDR brachytherapy with a shielded cylinder applicator.« less