Title: SU-E-J-109: Evaluation of Deformable Accumulated Parotid Doses Using Different Registration Algorithms in Adaptive Head and Neck Radiotherapy

Abstract

Purpose: Three deformable image registration (DIR) algorithms are utilized to perform deformable dose accumulation for head and neck tomotherapy treatment, and the differences of the accumulated doses are evaluated. Methods: Daily MVCT data for 10 patients with pathologically proven nasopharyngeal cancers were analyzed. The data were acquired using tomotherapy (TomoTherapy, Accuray) at the PLA General Hospital. The prescription dose to the primary target was 70Gy in 33 fractions.Three DIR methods (B-spline, Diffeomorphic Demons and MIMvista) were used to propagate parotid structures from planning CTs to the daily CTs and accumulate fractionated dose on the planning CTs. The mean accumulated doses of parotids were quantitatively compared and the uncertainties of the propagated parotid contours were evaluated using Dice similarity index (DSI). Results: The planned mean dose of the ipsilateral parotids (32.42±3.13Gy) was slightly higher than those of the contralateral parotids (31.38±3.19Gy)in 10 patients. The difference between the accumulated mean doses of the ipsilateral parotids in the B-spline, Demons and MIMvista deformation algorithms (36.40±5.78Gy, 34.08±6.72Gy and 33.72±2.63Gy ) were statistically significant (B-spline vs Demons, P<0.0001, B-spline vs MIMvista, p =0.002). And The difference between those of the contralateral parotids in the B-spline, Demons and MIMvista deformation algorithms (34.08±4.82Gy, 32.42±4.80Gy and 33.92±4.65Gy )more » were also significant (B-spline vs Demons, p =0.009, B-spline vs MIMvista, p =0.074). For the DSI analysis, the scores of B-spline, Demons and MIMvista DIRs were 0.90, 0.89 and 0.76. Conclusion: Shrinkage of parotid volumes results in the dose increase to the parotid glands in adaptive head and neck radiotherapy. The accumulated doses of parotids show significant difference using the different DIR algorithms between kVCT and MVCT. Therefore, the volume-based criterion (i.e. DSI) as a quantitative evaluation of registration accuracy is essential besides the visual assessment by the treating physician. This work was supported in part by the grant from Chinese Natural Science Foundation (Grant No. 11105225)« less

Authors:
 [1];  [2];  [3]
  1. Key Laboratory of Particle & Radiation Imaging (Tsinghua University), Ministry of Education, Beijing, 100084 China (China)
  2. (China)
  3. Image processing center, Beihang University, Beijing, 100191 China (China)
Publication Date:
OSTI Identifier:
22494124
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; ACCURACY; ALGORITHMS; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; GLANDS; HEAD; HOSPITALS; NECK; NEOPLASMS; PATIENTS; PLANNING; RADIATION DOSES

Citation Formats

Xu, S, Chinese PLA General Hospital, Beijing, 100853 China, and Liu, B. SU-E-J-109: Evaluation of Deformable Accumulated Parotid Doses Using Different Registration Algorithms in Adaptive Head and Neck Radiotherapy. United States: N. p., 2015. Web. doi:10.1118/1.4924196.
Xu, S, Chinese PLA General Hospital, Beijing, 100853 China, & Liu, B. SU-E-J-109: Evaluation of Deformable Accumulated Parotid Doses Using Different Registration Algorithms in Adaptive Head and Neck Radiotherapy. United States. doi:10.1118/1.4924196.
Xu, S, Chinese PLA General Hospital, Beijing, 100853 China, and Liu, B. Mon . "SU-E-J-109: Evaluation of Deformable Accumulated Parotid Doses Using Different Registration Algorithms in Adaptive Head and Neck Radiotherapy". United States. doi:10.1118/1.4924196.
@article{osti_22494124,
title = {SU-E-J-109: Evaluation of Deformable Accumulated Parotid Doses Using Different Registration Algorithms in Adaptive Head and Neck Radiotherapy},
author = {Xu, S and Chinese PLA General Hospital, Beijing, 100853 China and Liu, B},
abstractNote = {Purpose: Three deformable image registration (DIR) algorithms are utilized to perform deformable dose accumulation for head and neck tomotherapy treatment, and the differences of the accumulated doses are evaluated. Methods: Daily MVCT data for 10 patients with pathologically proven nasopharyngeal cancers were analyzed. The data were acquired using tomotherapy (TomoTherapy, Accuray) at the PLA General Hospital. The prescription dose to the primary target was 70Gy in 33 fractions.Three DIR methods (B-spline, Diffeomorphic Demons and MIMvista) were used to propagate parotid structures from planning CTs to the daily CTs and accumulate fractionated dose on the planning CTs. The mean accumulated doses of parotids were quantitatively compared and the uncertainties of the propagated parotid contours were evaluated using Dice similarity index (DSI). Results: The planned mean dose of the ipsilateral parotids (32.42±3.13Gy) was slightly higher than those of the contralateral parotids (31.38±3.19Gy)in 10 patients. The difference between the accumulated mean doses of the ipsilateral parotids in the B-spline, Demons and MIMvista deformation algorithms (36.40±5.78Gy, 34.08±6.72Gy and 33.72±2.63Gy ) were statistically significant (B-spline vs Demons, P<0.0001, B-spline vs MIMvista, p =0.002). And The difference between those of the contralateral parotids in the B-spline, Demons and MIMvista deformation algorithms (34.08±4.82Gy, 32.42±4.80Gy and 33.92±4.65Gy ) were also significant (B-spline vs Demons, p =0.009, B-spline vs MIMvista, p =0.074). For the DSI analysis, the scores of B-spline, Demons and MIMvista DIRs were 0.90, 0.89 and 0.76. Conclusion: Shrinkage of parotid volumes results in the dose increase to the parotid glands in adaptive head and neck radiotherapy. The accumulated doses of parotids show significant difference using the different DIR algorithms between kVCT and MVCT. Therefore, the volume-based criterion (i.e. DSI) as a quantitative evaluation of registration accuracy is essential besides the visual assessment by the treating physician. This work was supported in part by the grant from Chinese Natural Science Foundation (Grant No. 11105225)},
doi = {10.1118/1.4924196},
journal = {Medical Physics},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}
  • Purpose: To evaluate the ROI contours and accumulated dose difference using different deformable image registration (DIR) algorithms for head and neck (H&N) adaptive radiotherapy. Methods: Eight H&N cancer patients were randomly selected from the affiliated hospital. During the treatment, patients were rescanned every week with ROIs well delineated by radiation oncologist on each weekly CT. New weekly treatment plans were also re-designed with consistent dose prescription on the rescanned CT and executed for one week on Siemens CT-on-rails accelerator. At the end, we got six weekly CT scans from CT1 to CT6 including six weekly treatment plans for each patient.more » The primary CT1 was set as the reference CT for DIR proceeding with the left five weekly CTs using ANACONDA and MORFEUS algorithms separately in RayStation and the external skin ROI was set to be the controlling ROI both. The entire calculated weekly dose were deformed and accumulated on corresponding reference CT1 according to the deformation vector field (DVFs) generated by the two different DIR algorithms respectively. Thus we got both the ANACONDA-based and MORFEUS-based accumulated total dose on CT1 for each patient. At the same time, we mapped the ROIs on CT1 to generate the corresponding ROIs on CT6 using ANACONDA and MORFEUS DIR algorithms. DICE coefficients between the DIR deformed and radiation oncologist delineated ROIs on CT6 were calculated. Results: For DIR accumulated dose, PTV D95 and Left-Eyeball Dmax show significant differences with 67.13 cGy and 109.29 cGy respectively (Table1). For DIR mapped ROIs, PTV, Spinal cord and Left-Optic nerve show difference with −0.025, −0.127 and −0.124 (Table2). Conclusion: Even two excellent DIR algorithms can give divergent results for ROI deformation and dose accumulation. As more and more TPS get DIR module integrated, there is an urgent need to realize the potential risk using DIR in clinical.« less
  • Purposes: To systematically monitor anatomic variations and their dosimetric consequences during head-and-neck (H'N) radiation therapy using a GPU-based deformable image registration (DIR) framework. Methods: Eleven H'N IMRT patients comprised the subject population. The daily megavoltage CT and weekly kVCT scans were acquired for each patient. The pre-treatment CTs were automatically registered with their corresponding planning CT through an in-house GPU-based DIR framework. The deformation of each contoured structure was computed to account for non-rigid change in the patient setup. The Jacobian determinant for the PTVs and critical structures was used to quantify anatomical volume changes. Dose accumulation was performed tomore » determine the actual delivered dose and dose accumulation. A landmark tool was developed to determine the uncertainty in the dose distribution due to registration error. Results: Dramatic interfraction anatomic changes leading to dosimetric variations were observed. During the treatment courses of 6–7 weeks, the parotid gland volumes changed up to 34.7%, the center-of-mass displacement of the two parotids varied in the range of 0.9–8.8mm. Mean doses were within 5% and 3% of the planned mean doses for all PTVs and CTVs, respectively. The cumulative minimum/mean/EUD doses were lower than the planned doses by 18%, 2%, and 7%, respectively for the PTV1. The ratio of the averaged cumulative cord maximum doses to the plan was 1.06±0.15. The cumulative mean doses assessed by the weekly kVCTs were significantly higher than the planned dose for the left-parotid (p=0.03) and right-parotid gland (p=0.006). The computation time was nearly real-time (∼ 45 seconds) for registering each pre-treatment CT to the planning CT and dose accumulation with registration accuracy (for kVCT) at sub-voxel level (<1.5mm). Conclusions: Real-time assessment of anatomic and dosimetric variations is feasible using the GPU-based DIR framework. Clinical implementation of this technology may enable timely plan adaption and potentially lead to improved outcome.« less
  • Purpose: To estimate the accumulated dose to targets and organs at risk (OAR) for head and neck (H'N) radiotherapy using 3 deformable image registration (DIR) algorithms. Methods: Five H'N patients, who had daily CBCTs taken during the course of treatment, were retrospectively studied. All plans had 5 mm CTV-to-PTV expansions. To overcome the small field of view (FOV) limitations and HU uncertainties of CBCTs, CT images were deformably registered using a parameter-optimized B-spline DIR algorithm (Elastix, elastix.isi.uu.nl) and resampled onto each CBCT with a 4 cm uniform FOV expansion. The dose of the day was calculated on these resampled CTmore » images. Calculated daily dose matrices were warped and accumulated to the planning CT using 3 DIR algorithms; SmartAdapt (Eclipse/Varian), Velocity (Velocity Medical Solutions), and Elastix. Dosimetric indices for targets and OARs were determined from the DVHs and compared with corresponding planned quantities. Results: The cumulative dose deviation was less than 2%, on average, for PTVs from the corresponding plan dose, for all algorithms/patients. However, the parotids show as much as a 37% deviation from the intended dose, possibly due to significant patient weight loss during the first 3 weeks of treatment (15.3 lbs in this case). The mean(±SD) cumulative dose deviations of the 5 patients estimated using the 3 algorithms (SmartAdapt, Velocity, and Elastix) were (0.8±0.9%, 0.5±0.9%, 0.6±1.3%) for PTVs, (1.6±1.9%, 1.4±2.0%, 1.7±1.9%) for GTVs, (10.4±12.1%, 10.7±10.6%, 6.5±10.1%) for parotid glands, and (4.5±4.6%, 3.4±5.7%, 3.9±5.7%) for mucosa, respectively. The differences among the three DIR algorithms in the estimated cumulative mean doses (1SD (in Gy)) were: 0.1 for PTVs, 0.1 for GTVs, 1.9 for parotid glands, and 0.4 for mucosa. Conclusion: Results of this study are suggestive that more frequent plan adaptation for organs, such as the parotid glands, might be beneficial during the course of H'N RT. This study was supported in part by a research grant from Varian Medical Systems, Palo Alto, CA.« less
  • Purpose: To evaluate the anatomical changes and associated dosimetric consequences to the pharyngeal constrictor (PC) that occurs during head and neck radiotherapy (H&N RT). Methods: A cohort of 13 oro-pharyngeal cancer patients, who had daily CBCT’s for localization, was retrospectively studied. On every 5th CBCT, PC was manually delineated by a radiation oncologist. The anterior-posterior PC thickness was measured at the C3 level. Delivered dose to PC was estimated by calculating daily doses on CBCT’s, and accumulating to corresponding planning CT images. For accumulation, a parameter-optimized B- spline-based deformable image registration algorithm (Elastix) was used, in conjunction with an energy-massmore » mapping dose transfer algorithm. Mean and maximum dose (Dmean, Dmax) to PC was determined and compared with corresponding planned quantities. Results: The mean (±standard deviation) volume increase (ΔV) and thickness increase (Δt) over the course of 35 total fractions were 54±33% (11.9±7.6 cc), and 63±39% (2.9±1.9 mm), respectively. The resultant cumulative mean dose increase from planned dose to PC (ΔDmean) was 1.4±1.3% (0.9±0.8 Gy), while the maximum dose increase (ΔDmax) was 0.0±1.6% (0.0±1.1 Gy). Patients with adaptive replanning (n=6) showed a smaller mean dose increase than those without (n=7); 0.5±0.2% (0.3±0.1 Gy) vs. 2.2±1.4% (1.4±0.9 Gy). There was a statistically significant (p<0.0001) strong correlation between ΔDmean and Δt (Pearson coefficient r=0.78), and a moderate-to-strong correlation (r=0.52) between ΔDmean and ΔV. Correlation between ΔDmean and weight loss ΔW (r=0.1), as well as ΔV and ΔW (r=0.2) were negligible. Conclusion: Patients were found to undergo considerable anatomical changes to pharyngeal constrictor during H&N RT, resulting in non-negligible dose deviations from intended dose. Results are indicative that pharyngeal constrictor thickness, measured at C3 level, is a good predictor for the dose change to the organ. Daily deformable registration and dose accumulation provide a reliable means to assess important anatomical and dosimetric changes to pharyngeal constrictor occurring during treatment. This work was supported in part by a research grant from Varian Medical Systems, Palo Alto, CA.« less
  • Purpose: Accurate deformable image registration (DIR) between CT and CBCT in H&N is challenging. In this study, we propose a practical hybrid method that uses not only the pixel intensities but also organ physical properties, structure volume of interest (VOI), and interactive local registrations. Methods: Five oropharyngeal cancer patients were selected retrospectively. For each patient, the planning CT was registered to the last fraction CBCT, where the anatomy difference was largest. A three step registration strategy was tested; Step1) DIR using pixel intensity only, Step2) DIR with additional use of structure VOI and rigidity penalty, and Step3) interactive local correction.more » For Step1, a public-domain open-source DIR algorithm was used (cubic B-spline, mutual information, steepest gradient optimization, and 4-level multi-resolution). For Step2, rigidity penalty was applied on bony anatomies and brain, and a structure VOI was used to handle the body truncation such as the shoulder cut-off on CBCT. Finally, in Step3, the registrations were reviewed on our in-house developed software and the erroneous areas were corrected via a local registration using level-set motion algorithm. Results: After Step1, there were considerable amount of registration errors in soft tissues and unrealistic stretching in the posterior to the neck and near the shoulder due to body truncation. The brain was also found deformed to a measurable extent near the superior border of CBCT. Such errors could be effectively removed by using a structure VOI and rigidity penalty. The rest of the local soft tissue error could be corrected using the interactive software tool. The estimated interactive correction time was approximately 5 minutes. Conclusion: The DIR using only the image pixel intensity was vulnerable to noise and body truncation. A corrective action was inevitable to achieve good quality of registrations. We found the proposed three-step hybrid method efficient and practical for CT/CBCT registrations in H&N. My department receives grant support from Industrial partners: (a) Varian Medical Systems, Palo Alto, CA, and (b) Philips HealthCare, Best, Netherlands.« less