Title: SU-E-J-53: A Phantom Design to Assist Patient Position Verification System in Daily Image-Guided RT and Comprehensive QA Measurements

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924140· OSTI ID:22494075
;  [1]
  1. Willis-Knighton Medical Center, Shreveport, LA (United States)

Purpose This study is to implement a homemade novel device with surface locking couch index to check daily radiograph (DR) function of adaPTInsight™, stereoscopic image guided system (SIGS), for proton therapy. The comprehensive daily QA checks of proton pencil beam output, field size, flatness and symmetry of spots and energy layers will be followed by using MatriXX dosimetry device. Methods The iBa MatriXX device was used to perform daily dosimetry which is also used to perform SIGS checks. A set of markers were attached to surface of MatriXX device in alignment of DRR of reconstructed CT images and daily DR. The novel device allows MatriXX to be fit into the cradle which was locked by couch index bars on couch surface. This will keep the MatriXX at same XY plane daily with exact coordinates. Couch height Z will be adjusted according to imaging to check isocenter-laser coincidence accuracy. Results adaPTInsight™ provides robotic couch to move in 6-degree coordinate system to align the dosimetry device to be within 1.0 mm / 1.0°. The daily constancy was tightened to be ± 0.5 mm / 0.3° compared to 1.0 mm / 1.0° before. For gantry at 0° and couch all 0° angles (@ Rt ARM 0 setting), offsets measured of the couch systems were ≤ 0.5° in roll, yaw and pitch dimensions. Conclusion Simplicity of novel device made daily image guided QA consistent with accuracy. The offset of the MatriXX isocenter-laser coincident was reproducible. Such easy task not only speeds up the setup, but it increases confidence level in detailed daily comprehensive measurements. The total SIGS alignment time has been shortened with less setup error. This device will enhance our experiences for the future QA when cone beam CT imaging modality becomes available at proton therapy center.

OSTI ID:
22494075
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

SU-F-J-188: Clinical Implementation of in Room Mobile CT for Image Guided Proton Therapy
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22494075

SU-E-T-662: Quick and Efficient Daily QA for Compact PBS Proton Therapy Machine
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22494075

SU-E-T-268: Proton Radiosurgery End-To-End Testing Using Lucy 3D QA Phantom
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22494075