Title: SU-E-J-35: Using CBCT as the Alternative Method of Assessing ITV Volume

Abstract

Purpose To study the accuracy of Internal Target Volumes (ITVs) created on cone beam CT (CBCT) by comparing the visible target volume on CBCT to volumes (GTV, ITV, and PTV) outlined on free breathing (FB) CT and 4DCT. Methods A Quasar Cylindrical Motion Phantom with a 3cm diameter ball (14.14 cc) embedded within a cork insert was set up to simulate respiratory motion with a period of 4 seconds and amplitude of 2cm superioinferiorly and 1cm anterioposteriorly. FBCT and 4DCT images were acquired. A PTV-4D was created on the 4DCT by applying a uniform margin of 5mm to the ITV-CT. PTV-FB was created by applying a margin of the motion range plus 5mm, i.e. total of 1.5cm laterally and 2.5cm superioinferiorly to the GTV outlined on the FBCT. A dynamic conformal arc was planned to treat the PTV-FB with 1mm margin. A CBCT was acquired before the treatment, on which the target was delineated. During the treatment, the position of the target was monitored using the EPID in cine mode. Results ITV-CBCT and ITV-CT were measured to be 56.6 and 62.7cc, respectively, with a Dice Coefficient (DC) of 0.94 and disagreement in center of mass (COM) of 0.59 mm. Onmore » the other hand, GTV-FB was 11.47cc, 19% less than the known volume of the ball. PTV-FB and PTV-4D were 149 and 116 cc, with a DC of 0.71. Part of the ITV-CT was not enclosed by the PTV-FB despite the large margin. The cine EPID images have confirmed geometrical misses of the target. Similar under-coverage was observed in one clinical case and captured by the CBCT, where the implanted fiducials moved outside PTV-FB. Conclusion ITV-CBCT is in good agreement with ITV-CT. When 4DCT was not available, CBCT can be an effective alternative in determining and verifying the PTV margin.« less

Authors:
; ; ; ;  [1]
  1. Rush University Medical Center, Chicago, IL (United States)
Publication Date:
OSTI Identifier:
22494061
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 42; Journal Issue: 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; ACCURACY; AMPLITUDES; CENTER-OF-MASS SYSTEM; COMPUTERIZED TOMOGRAPHY; IMAGES; PHANTOMS; QUASARS; RESPIRATION

Citation Formats

Liao, Y, Turian, J, Templeton, A, Redler, G, and Chu, J. SU-E-J-35: Using CBCT as the Alternative Method of Assessing ITV Volume. United States: N. p., 2015. Web. doi:10.1118/1.4924122.
Liao, Y, Turian, J, Templeton, A, Redler, G, & Chu, J. SU-E-J-35: Using CBCT as the Alternative Method of Assessing ITV Volume. United States. doi:10.1118/1.4924122.
Liao, Y, Turian, J, Templeton, A, Redler, G, and Chu, J. Mon . "SU-E-J-35: Using CBCT as the Alternative Method of Assessing ITV Volume". United States. doi:10.1118/1.4924122.
@article{osti_22494061,
title = {SU-E-J-35: Using CBCT as the Alternative Method of Assessing ITV Volume},
author = {Liao, Y and Turian, J and Templeton, A and Redler, G and Chu, J},
abstractNote = {Purpose To study the accuracy of Internal Target Volumes (ITVs) created on cone beam CT (CBCT) by comparing the visible target volume on CBCT to volumes (GTV, ITV, and PTV) outlined on free breathing (FB) CT and 4DCT. Methods A Quasar Cylindrical Motion Phantom with a 3cm diameter ball (14.14 cc) embedded within a cork insert was set up to simulate respiratory motion with a period of 4 seconds and amplitude of 2cm superioinferiorly and 1cm anterioposteriorly. FBCT and 4DCT images were acquired. A PTV-4D was created on the 4DCT by applying a uniform margin of 5mm to the ITV-CT. PTV-FB was created by applying a margin of the motion range plus 5mm, i.e. total of 1.5cm laterally and 2.5cm superioinferiorly to the GTV outlined on the FBCT. A dynamic conformal arc was planned to treat the PTV-FB with 1mm margin. A CBCT was acquired before the treatment, on which the target was delineated. During the treatment, the position of the target was monitored using the EPID in cine mode. Results ITV-CBCT and ITV-CT were measured to be 56.6 and 62.7cc, respectively, with a Dice Coefficient (DC) of 0.94 and disagreement in center of mass (COM) of 0.59 mm. On the other hand, GTV-FB was 11.47cc, 19% less than the known volume of the ball. PTV-FB and PTV-4D were 149 and 116 cc, with a DC of 0.71. Part of the ITV-CT was not enclosed by the PTV-FB despite the large margin. The cine EPID images have confirmed geometrical misses of the target. Similar under-coverage was observed in one clinical case and captured by the CBCT, where the implanted fiducials moved outside PTV-FB. Conclusion ITV-CBCT is in good agreement with ITV-CT. When 4DCT was not available, CBCT can be an effective alternative in determining and verifying the PTV margin.},
doi = {10.1118/1.4924122},
journal = {Medical Physics},
number = 6,
volume = 42,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2015},
month = {Mon Jun 15 00:00:00 EDT 2015}
}
  • Purpose: To develop a CBCT HU correction method using a patient specific HU to mass density conversion curve based on a novel image registration and organ mapping method for head-and-neck radiation therapy. Methods: There are three steps to generate a patient specific CBCT HU to mass density conversion curve. First, we developed a novel robust image registration method based on sparseness analysis to register the planning CT (PCT) and the CBCT. Second, a novel organ mapping method was developed to transfer the organs at risk (OAR) contours from the PCT to the CBCT and corresponding mean HU values of eachmore » OAR were measured in both the PCT and CBCT volumes. Third, a set of PCT and CBCT HU to mass density conversion curves were created based on the mean HU values of OARs and the corresponding mass density of the OAR in the PCT. Then, we compared our proposed conversion curve with the traditional Catphan phantom based CBCT HU to mass density calibration curve. Both curves were input into the treatment planning system (TPS) for dose calculation. Last, the PTV and OAR doses, DVH and dose distributions of CBCT plans are compared to the original treatment plan. Results: One head-and-neck cases which contained a pair of PCT and CBCT was used. The dose differences between the PCT and CBCT plans using the proposed method are −1.33% for the mean PTV, 0.06% for PTV D95%, and −0.56% for the left neck. The dose differences between plans of PCT and CBCT corrected using the CATPhan based method are −4.39% for mean PTV, 4.07% for PTV D95%, and −2.01% for the left neck. Conclusion: The proposed CBCT HU correction method achieves better agreement with the original treatment plan compared to the traditional CATPhan based calibration method.« less
  • Purpose: To investigate the feasibility of using nanoparticle markers to validate liver tumor motion together with a deformation field map-based four dimensional (4D) cone-beam computed tomography (CBCT) reconstruction method. Methods: A technique for lung 4D-CBCT reconstruction has been previously developed using a deformation field map (DFM)-based strategy. In this method, each phase of the 4D-CBCT is considered as a deformation of a prior CT volume. The DFM is solved by a motion modeling and free-form deformation (MM-FD) technique, using a data fidelity constraint and the deformation energy minimization. For liver imaging, there is low contrast of a liver tumor inmore » on-board projections. A validation of liver tumor motion using implanted gold nanoparticles, along with the MM-FD deformation technique is implemented to reconstruct onboard 4D CBCT liver radiotherapy images. These nanoparticles were placed around the liver tumor to reflect the tumor positions in both CT simulation and on-board image acquisition. When reconstructing each phase of the 4D-CBCT, the migrations of the gold nanoparticles act as a constraint to regularize the deformation field, along with the data fidelity and the energy minimization constraints. In this study, multiple tumor diameters and positions were simulated within the liver for on-board 4D-CBCT imaging. The on-board 4D-CBCT reconstructed by the proposed method was compared with the “ground truth” image. Results: The preliminary data, which uses reconstruction for lung radiotherapy suggests that the advanced reconstruction algorithm including the gold nanoparticle constraint will Resultin volume percentage differences (VPD) between lesions in reconstructed images by MM-FD and “ground truth” on-board images of 11.5% (± 9.4%) and a center of mass shift of 1.3 mm (± 1.3 mm) for liver radiotherapy. Conclusion: The advanced MM-FD technique enforcing the additional constraints from gold nanoparticles, results in improved accuracy for reconstructing on-board 4D-CBCT of liver tumor. Varian medical systems research grant.« less
  • Purpose: Four-dimensional-computed-tomography(4D-CT) imaging for target-volume delineation and cone-beam-tomography(CBCT) for treatment localization are widely utilized in lung-SABR.Aim of this study was to perform a quantitative-assessment and inter-comparison of Internal-targetvolumes( ITV) drawn on various phases of breathing-cycle 4D-CT-scans, Maximum-intensity-projection(MIP), average-intensity-projection(AIP)and static CT-scans of lung-motion-phantom to simulate lung-SABR patient geometry. We also analyzed and compared the ITVs drawn on freebreathing- CBCT. Materials and Methods: 4D-CT-scans were acquired on Philips big-bore 16slice CT and Bellows-respiratory monitoring-system using retrospective phase-binning method. Each respiratory cycle divided into 10-phases. Quasar-Phantom with lung-inserts and 3cm-diameter nylonball to simulate tumor and was placed on respiratory-motion-platform for 4D-CT and CBCT-acquisition.more » Amplitudes of motions: 0.5,1.0,2.0,3.0cm in superior-inferior direction with breathing-cycle time of 6,5,4,6sec, respectively used.4D-CTs with 10-phases(0%to90%)for each excursion-set and 3D-CT for static-phantom exported to iPlan treatment-planningsystem( TPS).Tumor-volumes delineated in all phases of 4D-CT, MIP,AIP,CBCT scans using fixed-HU-threshold(−500to1000)values automatically.For each 4D-dataset ITV obtained by unifying the tumorcontours on all phases.CBCT-ITV-volumes were drawn in Eclipse-TPS. Results: Mean volume of tumor contours for all phases compared with static 3D-CT were 0.62±0.08%, 1.67±0.26%, 4.77±0.54% and 9.27±1.23% for 0.5cm,1cm,2cm,3cm excursions respectively. Differences of mean Union-ITV with MIP-ITV were close(≤2.4%).Mean Union-ITV from expected-theoretical values differed from −4.9% to 3.8%.Union-ITV and MIP-ITV were closer within 2.3%. AIP-ITVs were underestimated from 14 to 32% compared to union-ITV for all motion datasets. Differences of −5.9% to −44% and −5% to 6.7% for CBCT-ITV from MIP-ITV and AIP-ITV respectively.Motion excursions and centroid positions were within 2mm for 4D-CT and CBCT-ITVs to that of expected values. Conclusion: 4D-CT MIP-ITV and Union-ITV showed very good agreement that validates that ITV can be fast contoured on MIP.Contouring ITV in AIP must be avoided as it significantly underestimates the volume with all excursions. Free breathing CBCT-ITV showed good agreement with AIP-ITV but underestimated the MIP-ITV. Estimation of excursions and centroid values for 4D-CT and CBCT were in good agreement with expected values.« less
  • Purpose: To acquire upright CBCT images using the onboard imager of a Varian TrueBeam. An easy to implement upright imaging protocol could allow for widespread upright radiation therapy which would greatly benefit certain patients. These include thoracic cancer patients (because lung volume increases in a seated position) and patients who experience substantial discomfort during supine treatment. Methods: To acquire upright CBCT images, the gantry head remained stationary at 0 degrees with the KV imager arms extended to their lateral positions. Phantoms were placed upright at the end of the treatment couch. During a scan, the couch rotated from 270 tomore » 90 degrees while continuous fluoroscopic projections were taken by the onboard imager. To extend the field-of-view, this sequence was performed twice: once with the KV detector longitudinally offset +14.5cm and once with it longitudinally offset −14.5cm. The resulting two image sets were stitched together before reconstruction. The imaging beam parameters were chosen to deliver a dose similar to that given during a simulation CT. Image quality was evaluated for spatial linearity, high and low contrast resolution, and HU linearity using CatPhan and anthropomorphic phantoms. A deformable registration technique was used to evaluate HU mapping from a simulation CT. Results: Spatial linearity and high contrast resolution were maintained in upright CBCT when compared to simulation CT. However, low contrast resolution and HU linearity degraded. Streak artifacts were caused by the limited 180 degree arc of the couch, and the stitching process created a sharp artifact at the center of the reconstruction. The deformable registration was robust in the HU mapping even with these artifacts and the loss of HU linearity. Conclusions: The image quality obtained from upright CBCT was sufficient for treatment planning. The success of this novel technique is an important step towards a future clinical protocol. This project was funded in part by Varian Medical Systems, Inc.« less
  • Purpose: Ensuring reproducibility of target localization is critical to accurate stereotactic body radiation treatment (SBRT) for lung and soft tissue metastatic lesions. To characterize interfraction variability in set-up and evaluate PTV margins utilized for SBRT, daily CBCTs were used to calculate delivered target and OAR doses compared to those expected from planning. Methods: CBCT images obtained prior to each fraction of SBRT for a lung and thyroid metastatic lesion were evaluated. The target CTV/ITV and OARs on each of 8 CBCT data sets were contoured. Using MIM fusion software and Pinnacle{sup 3} RTP system, delivered dose distribution was reconstructed onmore » each CBCT, utilizing translational shifts performed prior to treatment. Actual delivered vs. expected doses received by target CTV/ITV and adjacent critical structures were compared to characterize accuracy of pre-treatment translational shifts and PTV margins. Results: The planned CTV/ITV D95% and V100% were 4595cGy and 91.47% for the lung lesion, and 3010cGy and 96.34% for the thyroid lesion. Based on CBCT analysis, actual mean D95% and V100% for lung ITV were 4542±344.4cGy and 91.54±3.45%; actual mean D95% and V100% for thyroid metastasis CTV were 3005±25.98cGy and 95.20±2.522%. For the lung lesion, ipsilateral lung V20, heart V32 (cc) and spinal cord (.03 cc) max were 110.15cc, 3.33cc, and 1680cGy vs. 110.27±14.79cc, 6.74±3.76cc, and 1711±46.56cGy for planned vs. delivered doses, respectively. For the thyroid metastatic lesion, esophagus V18, trachea (.03 cc) max, and spinal cord (.03 cc) max were 0.35cc, 2555cGy, and 850cGy vs. 0.16±0.13cc, 2147±367cGy, and 838±45cGy for planned vs. delivered treatments, respectively. Conclusion: Minimal variability in SBRT target lesion dose delivered based on pre-treatment CBCT-based translational shifts suggests tighter PTV margins may be considered to further decrease dose to surrounding critical structures. Guidelines for optimal target alignment during CBCT-guidance for lung and soft tissue metastatic lesions treated with SBRT are being established.« less