We present a nanocalorimeter designed for the measurement of the dynamic heat capacity of thin films. The microfabricated sensor, the thermal conditioning of the sensor, as well as the highly stable and low noise electronic chain allow measurements of the real and imaginary parts of the complex specific heat with a resolution Δ C/C of about 10{sup −5}. The performances of this quasi-differential nanocalorimeter were tested on a model of polymeric glass-former, the polyvinyl acetate (PVAc). The high stability and low noise of the device are essential for accurate studies on non-equilibrium slow relaxing systems such as glasses.

OSTI ID:
22482620
Journal Information:
Review of Scientific Instruments, Vol. 86, Issue 11; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748