skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Compatibility of an Ultraselective Microcatheter and Epirubicin Loaded 300–500-μm DC Bead in Ex Vivo Study

Journal Article · · Cardiovascular and Interventional Radiology
; ; ; ;  [1]
  1. Nara Medical University, Department of Radiology and IVR Center (Japan)

PurposeThe purpose of this study is to examine whether epirubicin loaded DC Bead 300–500 μm in size can pass through a 1.8-Fr ultraselective microcatheter in ex vivo study.MethodsEpirubicin (25 mg/1 mL) loaded 100–300 and 300–500 μm DC Bead were tested. Both sizes were diluted 5, 10, and 30 times using contrast material. Ultraselective microcatheter with the outer diameter of 1.8 Fr and the inner diameter of .017 inch (431.8 μm) was used. The diluted DC Bead was injected at a speed of 1 mL/min, and the pressure was continuously measured. The microspheres’ shapes after ejection were observed by a stereomicroscope.ResultsThe maximum pressure of contrast material alone was 8.40 ± 0.21 psi. The maximum pressure in 5, 10, and 30 times dilution groups of 100–300 μm were 9.67 ± 1.18, 9.25 ± 0.25, and 9.71 ± 0.28 psi, respectively, whereas 21.10 ± 10.2, 10.48 ± 2.14, 10.09 ± 0.37 psi, respectively in 300–500 μm groups. The maximum pressure in 5 times dilution group of 300–500 μm was significantly higher than the other groups (P < 0.05). In 300–500 μm, 4 of 10 measurements showed high pressure over 24 psi (the maximum value was 43.5 psi) in 5 times dilution group, whereas in 10 times and 30 times dilution groups, all measurements showed less than 12 psi. No damages of microspheres were found.ConclusionsEpirubicin loaded DC Bead 300–500 μm in size can pass through a 1.8-Fr ultraselective microcatheter. To avoid high resistance due to microspheres’ aggregation, dilution more than 10 times is needed.

OSTI ID:
22469770
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 38, Issue 5; Other Information: Copyright (c) 2015 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); http://www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English