Title: Dual energy CT: How well can pseudo-monochromatic imaging reduce metal artifacts?

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4905106· OSTI ID:22413410
; ; ;  [1];  [2]
  1. German Cancer Research Center (DKFZ), Heidelberg 69120 (Germany)
  2. Friedrich-Alexander-University (FAU), Erlangen 91054 (Germany)

Purpose: Dual Energy CT (DECT) provides so-called monoenergetic images based on a linear combination of the original polychromatic images. At certain patient-specific energy levels, corresponding to certain patient- and slice-dependent linear combination weights, e.g., E = 160 keV corresponds to α = 1.57, a significant reduction of metal artifacts may be observed. The authors aimed at analyzing the method for its artifact reduction capabilities to identify its limitations. The results are compared with raw data-based processing. Methods: Clinical DECT uses a simplified version of monochromatic imaging by linearly combining the low and the high kV images and by assigning an energy to that linear combination. Those pseudo-monochromatic images can be used by radiologists to obtain images with reduced metal artifacts. The authors analyzed the underlying physics and carried out a series expansion of the polychromatic attenuation equations. The resulting nonlinear terms are responsible for the artifacts, but they are not linearly related between the low and the high kV scan: A linear combination of both images cannot eliminate the nonlinearities, it can only reduce their impact. Scattered radiation yields additional noncanceling nonlinearities. This method is compared to raw data-based artifact correction methods. To quantify the artifact reduction potential of pseudo-monochromatic images, they simulated the FORBILD abdomen phantom with metal implants, and they assessed patient data sets of a clinical dual source CT system (100, 140 kV Sn) containing artifacts induced by a highly concentrated contrast agent bolus and by metal. In each case, they manually selected an optimal α and compared it to a raw data-based material decomposition in case of simulation, to raw data-based material decomposition of inconsistent rays in case of the patient data set containing contrast agent, and to the frequency split normalized metal artifact reduction in case of the metal implant. For each case, the contrast-to-noise ratio (CNR) was assessed. Results: In the simulation, the pseudo-monochromatic images yielded acceptable artifact reduction results. However, the CNR in the artifact-reduced images was more than 60% lower than in the original polychromatic images. In contrast, the raw data-based material decomposition did not significantly reduce the CNR in the virtual monochromatic images. Regarding the patient data with beam hardening artifacts and with metal artifacts from small implants the pseudo-monochromatic method was able to reduce the artifacts, again with the downside of a significant CNR reduction. More intense metal artifacts, e.g., as those caused by an artificial hip joint, could not be suppressed. Conclusions: Pseudo-monochromatic imaging is able to reduce beam hardening, scatter, and metal artifacts in some cases but it cannot remove them. In all cases, the CNR is significantly reduced, thereby rendering the method questionable, unless special post processing algorithms are implemented to restore the high CNR from the original images (e.g., by using a frequency split technique). Raw data-based dual energy decomposition methods should be preferred, in particular, because the CNR penalty is almost negligible.

OSTI ID:
22413410
Journal Information:
Medical Physics, Vol. 42, Issue 2; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

WE-A-BRF-01: Dual-Energy CT Imaging in Diagnostic Imaging and Radiation Therapy
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22413410

Segmentation-free empirical beam hardening correction for CT
Journal Article · Sun Feb 15 00:00:00 EST 2015 · Medical Physics · OSTI ID:22413410

Dual energy exposure control (DEEC) for computed tomography: Algorithm and simulation study
Journal Article · Sat Nov 15 00:00:00 EST 2008 · Medical Physics · OSTI ID:22413410