Purpose: To develop a strategy for defining meaningful tolerance limits and studying the sensitivity of IMRT QA gamma criteria by inducing known errors in QA plans. Methods: IMRT QA measurements (ArcCHECK, Sun Nuclear) were compared to QA plan calculations with induced errors. Many (>24) gamma comparisons between data and calculations were performed for each of several kinds of cases and classes of induced error types with varying magnitudes (e.g. MU errors ranging from -10% to +10%), resulting in over 3,000 comparisons. Gamma passing rates for each error class and case were graphed against error magnitude to create error curves in order to represent the range of missed errors in routine IMRT QA using various gamma criteria. Results: This study demonstrates that random, case-specific, and systematic errors can be detected by the error curve analysis. Depending on location of the peak of the error curve (e.g., not centered about zero), 3%/3mm threshold=10% criteria may miss MU errors of up to 10% and random MLC errors of up to 5 mm. Additionally, using larger dose thresholds for specific devices may increase error sensitivity (for the same X%/Ymm criteria) by up to a factor of two. This analysis will allow clinics to select more meaningful gamma criteria based on QA device, treatment techniques, and acceptable error tolerances. Conclusion: We propose a strategy for selecting gamma parameters based on the sensitivity of gamma criteria and individual QA devices to induced calculation errors in QA plans. Our data suggest large errors may be missed using conventional gamma criteria and that using stricter criteria with an increased dose threshold may reduce the range of missed errors. This approach allows quantification of gamma criteria sensitivity and is straightforward to apply to other combinations of devices and treatment techniques.

OSTI ID:
22409669
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

IMRT QA: Selecting gamma criteria based on error detection sensitivity
Journal Article · Fri Apr 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22409669

TU-C-BRE-09: Performance Comparisons of Patient Specific IMRT QA Methodologies Using ROC Analysis
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22409669

SU-F-BRE-13: Replacing Pre-Treatment Phantom QA with 3D In-Vivo Portal Dosimetry for IMRT Breast Cancer
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22409669