Title: Poster — Thur Eve — 57: Evaluation of laryngeal mucosal dose with conventional linac and TomoTherapy

Abstract

The purpose of this study was to examine whether or not underdosing occurs in the mucosal layer during treatment of glottis cancer. A larynx phantom was produced and regions at risk of recurrence due to suspected underdosing were identified and wells drilled into the phantom for flush placement of TLDs. Seven interest points were chosen. CT simulation was completed prior to the wells being drilled, and again afterwards with the TLD locations indicated using BBs. Treatment plans created for this investigation included: 3DCRT using Elekta-XiO (n=9) and VMAT created using Elekta-Monaco (n=9), both delivered on an Elekta linac; standard TomoTherapy plan (n=11) and a directionally blocked TomoTherapy plan to approximate a 3D-conformal approach (n=5). Imaging dose during TomoTherapy deliveries was accounted for. The average TLD result at each interest point was compared to the planned value using a paired t-test. There was no significant difference between the planned and measured 3DCRT dose (268.9 vs. 267.0 cGy, respectively; p>0.05). Similarly, the planned and measured TomoTherapy treatment did not show any significant differences (271.7 vs 269.7 cGy; p>0.05). In the blocked TomoTherapy plan, significant overdosing was seen (274.5 vs 294.9 cGy; p<0.05) and underdosing was not seen in the VMAT treatment (303.5more » vs 321.8 cGy; p>0.05). Further investigation is ongoing to ensure appropriate normalization of results and to investigate the overdosing noted with the blocked TomoTherapy plan. Results from this study suggest that significant underdosing does not occur in the conventional treatment of early glottic cancer using 6MV photons.« less

Authors:
;  [1];  [2]
  1. Department of Medical Physics, The Ottawa Hospital Cancer Centre, Ottawa, Ontario (Canada)
  2. Department of Radiation Oncology, The Ottawa Hospital Cancer Centre, Ottawa, Ontario (Canada)
Publication Date:
OSTI Identifier:
22407896
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 8; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; LARYNX; LINEAR ACCELERATORS; NEOPLASMS; PHANTOMS

Citation Formats

Nusrat, H, Lekx, K, and Eapen, L. Poster — Thur Eve — 57: Evaluation of laryngeal mucosal dose with conventional linac and TomoTherapy. United States: N. p., 2014. Web. doi:10.1118/1.4894917.
Nusrat, H, Lekx, K, & Eapen, L. Poster — Thur Eve — 57: Evaluation of laryngeal mucosal dose with conventional linac and TomoTherapy. United States. doi:10.1118/1.4894917.
Nusrat, H, Lekx, K, and Eapen, L. Fri . "Poster — Thur Eve — 57: Evaluation of laryngeal mucosal dose with conventional linac and TomoTherapy". United States. doi:10.1118/1.4894917.
@article{osti_22407896,
title = {Poster — Thur Eve — 57: Evaluation of laryngeal mucosal dose with conventional linac and TomoTherapy},
author = {Nusrat, H and Lekx, K and Eapen, L},
abstractNote = {The purpose of this study was to examine whether or not underdosing occurs in the mucosal layer during treatment of glottis cancer. A larynx phantom was produced and regions at risk of recurrence due to suspected underdosing were identified and wells drilled into the phantom for flush placement of TLDs. Seven interest points were chosen. CT simulation was completed prior to the wells being drilled, and again afterwards with the TLD locations indicated using BBs. Treatment plans created for this investigation included: 3DCRT using Elekta-XiO (n=9) and VMAT created using Elekta-Monaco (n=9), both delivered on an Elekta linac; standard TomoTherapy plan (n=11) and a directionally blocked TomoTherapy plan to approximate a 3D-conformal approach (n=5). Imaging dose during TomoTherapy deliveries was accounted for. The average TLD result at each interest point was compared to the planned value using a paired t-test. There was no significant difference between the planned and measured 3DCRT dose (268.9 vs. 267.0 cGy, respectively; p>0.05). Similarly, the planned and measured TomoTherapy treatment did not show any significant differences (271.7 vs 269.7 cGy; p>0.05). In the blocked TomoTherapy plan, significant overdosing was seen (274.5 vs 294.9 cGy; p<0.05) and underdosing was not seen in the VMAT treatment (303.5 vs 321.8 cGy; p>0.05). Further investigation is ongoing to ensure appropriate normalization of results and to investigate the overdosing noted with the blocked TomoTherapy plan. Results from this study suggest that significant underdosing does not occur in the conventional treatment of early glottic cancer using 6MV photons.},
doi = {10.1118/1.4894917},
journal = {Medical Physics},
number = 8,
volume = 41,
place = {United States},
year = {Fri Aug 15 00:00:00 EDT 2014},
month = {Fri Aug 15 00:00:00 EDT 2014}
}
  • Lung tumours move due to respiratory motion. This is managed during planning by acquiring a 4DCT and capturing the excursion of the GTV (gross tumour volume) throughout the breathing cycle within an IGTV (Internal Gross Tumour Volume) contour. Patients undergo a verification cone-beam CT (CBCT) scan immediately prior to treatment. 3D reconstructed images do not consider tumour motion, resulting in image artefacts, such as blurring. This may lead to difficulty in identifying the tumour on reconstructed images. It would be valuable to create a 4DCBCT reconstruction of the tumour motion to confirm that does indeed remain within the planned IGTV.more » CBCT projections of a Quasar Respiratory Motion Phantom are acquired in Treatment mode (half-fan scan) on a Varian TrueBeam accelerator. This phantom contains a mobile, low-density lung insert with an embedded 3cm diameter tumour object. It is programmed to create a 15s periodic, 2cm (sup/inf) displacement. A Varian Real-time Position Management (RPM) tracking-box is placed on the phantom breathing platform. Breathing phase information is automatically integrated into the projection image files. Using in-house Matlab programs and RTK (Reconstruction Tool Kit) open-source toolboxes, the projections are re-binned into 10 phases and a 4DCBCT scan reconstructed. The planning IGTV is registered to the 4DCBCT and the tumour excursion is verified to remain within the planned contour. This technique successfully reconstructs 4DCBCT images using clinical modes for a breathing phantom. UBC-BCCA ethics approval has been obtained to perform 4DCBCT reconstructions on lung patients (REB#H12-00192). Clinical images will be accrued starting April 2014.« less
  • The purpose of this work is twofold: 1) to measure dose profiles under lead shielding at the level of the lens for a range of clinical electron energies via film dosimetry; and, 2) to assess the validity of the Pinnacle treatment planning system (TPS) in calculating the penumbral doses under lead shielding with the heterogeneous electron algorithm. First, a film calibration curve that spanned the electron energies of interest, 6–18MeV, was created. Next, EBT3 film and lead shielding were incorporated into a solid water phantom with the film positioned 7mm below the lead and a variable thickness of bolus onmore » top. This geometry was reproduced in the Pinnacle TPS and used to calculate dose profiles using the heterogeneous electron algorithm. The measured vs. calculated dose profiles were normalized to d{sub max} in a homogeneous phantom with no lead shielding and compared. Pinnacle consistently overestimated the dose distal to the lead shielding with significant discrepancies occurring near the edge of the lead shield reaching 25% at the edge and 35% in the open field region. The film measurements showed that a minimum lead margin of 5mm extending beyond the diameter of the lens is required to adequately shield the lens to ≤10% of the dose at d{sub max}. These measurements allow for a reasonable estimate of the dose to the lens from anterior electron beams. They also allow for clinicians to assess the extent of the lead margin required to reduce the lens dose to an acceptable amount prior to radiotherapy treatment.« less
  • Use of helical TomoTherapy based MVCT imaging for adaptive planning requires the image values (HU) to remain stable over the course of treatment. In the past, the image value stability was suboptimal, which required frequent change to the image value to density calibration curve to avoid dose errors on the order of 2–4%. The stability of the image values at our center was recently improved by stabilizing the dose rate of the machine (dose control servo) and performing daily MVCT calibration corrections. In this work, we quantify the stability of the image values over treatment time by comparing patient treatmentmore » image density derived using MVCT and KVCT. The analysis includes 1) MVCT - KVCT density difference histogram, 2) MVCT vs KVCT density spectrum, 3) multiple average profile density comparison and 4) density difference in homogeneous locations. Over two months, the imaging beam stability was compromised several times due to a combination of target wobbling, spectral calibration, target change and magnetron issues. The stability of the image values were analyzed over the same period. Results show that the impact on the patient dose calculation is 0.7% +− 0.6%.« less
  • Introduction: RADPOS 4D dosimetry system consists of a microMOSFET dosimeter combined with an electromagnetic positioning sensor, which allows for performing real-time dose and position measurements simultaneously. In this report the use of RADPOS as an independent quality assurance (QA) tool during CyberKnife 4D radiotherapy treatment is described. In addition to RADPOS, GAFCHROMIC® films were used for simultaneous dose measurement. Methods: RADPOS and films were calibrated in a Solid Water® phantom at 1.5 cm depth, SAD= 80 cm, using 60 mm cone. CT based treatment plan was created for a Solid Water® breast phantom containing metal fiducials and RADPOS probe. Dosemore » calculations were performed using iPlan pencil beam algorithm. Before the treatment delivery, GAFCHROMIC® film was inserted inside the breast phantom, next to the RADPOS probe. Then the phantom was positioned on the chest platform of the QUASAR, to which Synchrony LED optical markers were also attached. Position logging began for RADPOS and the Synchrony tracking system, the QUASAR motion was initiated and the treatment was delivered. Results: RADPOS position measurements very closely matched the LED marker positions recorded by the Synchrony camera tracking system. The RADPOS measured dose was 2.5% higher than the average film measured dose, which is within the experimental uncertainties. Treatment plan calculated dose was 4.1 and 1.6% lower than measured by RADPOS and film, respectively. This is most likely due to the inferior nature of the dose calculation algorithm. Conclusions: Our study demonstrates that RADPOS system is a useful tool for independent QA of CyberKnife treatments.« less
  • Introduction: Radiation detector responses can be affected by dose rate. Due to higher dose per pulse and wider range of mu rates in FFF beams, detector responses should be characterized prior to implementation of QA protocols for FFF beams. During VMAT delivery, the MU rate may also vary dramatically within a treatment fraction. This study looks at the dose per pulse variation throughout a 3D volume for typical VMAT plans and the response characteristics for a variety of detectors, and makes recommendations on the design of QA protocols for FFF VMAT QA. Materials and Methods: Linac log file data andmore » a simplified dose calculation algorithm are used to calculate dose per pulse for a variety of clinical VMAT plans, on a voxel by voxel basis, as a function of time in a cylindrical phantom. Diode and ion chamber array responses are characterized over the relevant range of dose per pulse and dose rate. Results: Dose per pulse ranges from <0.1 mGy/pulse to 1.5 mGy/pulse in a typical VMAT treatment delivery using the 10XFFF beam. Diode detector arrays demonstrate increased sensitivity to dose (+./− 3%) with increasing dose per pulse over this range. Ion chamber arrays demonstrate decreased sensitivity to dose (+/− 1%) with increasing dose rate over this range. Conclusions: QA protocols should be designed taking into consideration inherent changes in detector sensitivity with dose rate. Neglecting to account for changes in detector response with dose per pulse can lead to skewed QA results.« less