Title: Poster — Thur Eve — 21: Off-axis dose perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam for the PTW microLion and Exradin A1SL ionization chambers

Abstract

A PTW microLion liquid ionization chamber and an Exradin A1SL air-filled ionization chamber have been modeled using the egs-chamber user code of the EGSnrc system to determine their perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam. A model of the Varian CL21EX linear accelerator was constructed using the BEAMnrc Monte Carlo code, and was validated by comparing measured PDDs and profiles from the microLion and A1SL chambers to calculated results that included chamber models. Measured PDDs for a 5 × 5 cm{sup 2} field for the microLion chamber agreed with calculations to within 1.5% beyond a depth of 0.5 cm, and the A1SL PDDs agreed within 1.0% beyond 1.0 cm. Measured and calculated profiles at 10 cm depth agreed within 1.0% for both chambers inside the field, and within 4.0% near the field edge. Local percent differences increased up to 15% at 4 cm outside the field. The ratio of dose to water in the absence of the chamber relative to dose in the chamber's active volume as a function of off-axis distance was calculated using the egs-chamber correlated sampling technique. The dose ratio was nearly constant inside the field and consistent withmore » the stopping power ratios of water to detector material, but varied up to 3.3% near the field edge and 5.2% at 4 cm outside the field. Once these perturbation effects are fully characterized for more field sizes and detectors, they could be applied to clinical water tank measurements for improved dosimetric accuracy.« less

Authors:
; ; ;  [1]
  1. Medical Physics Unit, McGill University, Montréal, Québec (Canada)
Publication Date:
OSTI Identifier:
22407895
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 8; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
07 ISOTOPES AND RADIATION SOURCES; 60 APPLIED LIFE SCIENCES; ACCURACY; IONIZATION CHAMBERS; LINEAR ACCELERATORS; MONTE CARLO METHOD; PHOTON BEAMS; STOPPING POWER; WATER

Citation Formats

O'Grady, K, Davis, S D, Papaconstadopoulos, P, and Seuntjens, J. Poster — Thur Eve — 21: Off-axis dose perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam for the PTW microLion and Exradin A1SL ionization chambers. United States: N. p., 2014. Web. doi:10.1118/1.4894877.
O'Grady, K, Davis, S D, Papaconstadopoulos, P, & Seuntjens, J. Poster — Thur Eve — 21: Off-axis dose perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam for the PTW microLion and Exradin A1SL ionization chambers. United States. doi:10.1118/1.4894877.
O'Grady, K, Davis, S D, Papaconstadopoulos, P, and Seuntjens, J. Fri . "Poster — Thur Eve — 21: Off-axis dose perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam for the PTW microLion and Exradin A1SL ionization chambers". United States. doi:10.1118/1.4894877.
@article{osti_22407895,
title = {Poster — Thur Eve — 21: Off-axis dose perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam for the PTW microLion and Exradin A1SL ionization chambers},
author = {O'Grady, K and Davis, S D and Papaconstadopoulos, P and Seuntjens, J},
abstractNote = {A PTW microLion liquid ionization chamber and an Exradin A1SL air-filled ionization chamber have been modeled using the egs-chamber user code of the EGSnrc system to determine their perturbation effects in water in a 5 × 5 cm{sup 2} 18 MV photon beam. A model of the Varian CL21EX linear accelerator was constructed using the BEAMnrc Monte Carlo code, and was validated by comparing measured PDDs and profiles from the microLion and A1SL chambers to calculated results that included chamber models. Measured PDDs for a 5 × 5 cm{sup 2} field for the microLion chamber agreed with calculations to within 1.5% beyond a depth of 0.5 cm, and the A1SL PDDs agreed within 1.0% beyond 1.0 cm. Measured and calculated profiles at 10 cm depth agreed within 1.0% for both chambers inside the field, and within 4.0% near the field edge. Local percent differences increased up to 15% at 4 cm outside the field. The ratio of dose to water in the absence of the chamber relative to dose in the chamber's active volume as a function of off-axis distance was calculated using the egs-chamber correlated sampling technique. The dose ratio was nearly constant inside the field and consistent with the stopping power ratios of water to detector material, but varied up to 3.3% near the field edge and 5.2% at 4 cm outside the field. Once these perturbation effects are fully characterized for more field sizes and detectors, they could be applied to clinical water tank measurements for improved dosimetric accuracy.},
doi = {10.1118/1.4894877},
journal = {Medical Physics},
number = 8,
volume = 41,
place = {United States},
year = {Fri Aug 15 00:00:00 EDT 2014},
month = {Fri Aug 15 00:00:00 EDT 2014}
}
  • In cases where surgery is not possible for lung cancer treatment, stereotactic body radiation therapy (SBRT) may be an option. One problem when treating this type of cancer is the motion of the lungs caused by the patient's respiration. It is possible to reduce the impact of this movement with the use of respiratory gating. By combining respiratory gating with a flattening filter free (FFF) photon beam linac, the increased treatment time caused by a reduced beam-on time of respiratory gating methods can be compensated by the inherent increased dose rate of FFF beams. This project's aim is to createmore » hardware and software interfaces allowing free respiration gating on an Elekta Synergy-S linac specially modified to deliver 6 MV FFF photon beams. First, a printed circuit board was created for reading the signal from a Bellows Belt from Philips (a respiration monitor belt) and transmitting an On/Off signal to the accelerator. A software was also developed to visualize patient respiration. Secondly, a FFF model was created with the Pinnacle treatment planning system from Philips. Gamma (Γ) analysis (2%, 2 mm) was used to evaluate model. For fields going from 5.6 × 5.6 to 12 × 12 cm{sup 2}, central axis depth dose model fitting shows an average gamma value of 0.2 and 100% of gamma values remain under the Γ = 1 limit. For smaller fields (0.8 × 0.8 and 1.6 × 1.6 cm{sup 2}), Pinnacle has more trouble trying to fit the measurements, overestimating dose in penumbra and buildup regions.« less
  • Purpose: To assess radiation dose from the cone beam CT (CBCT) component of SPECT/CT studies and to compare with other CT examinations performed in our institution. Methods: We used an anthropomorphic chest phantom and the 6 cc ion chamber to measure entrance breast dose for several CBCT and diagnostic CT acquisition protocols. The CBCT effective dose was calculated with ImPACT software; the CT effective dose was evaluated from the DLP value and conversion factor, dependent on the anatomic region. The RADAR medical procedure radiation dose calculator was used to assess the nuclear medicine component of exam dose. Results: The entrancemore » dose to the breast measured with the anthropomorphic phantom was 0.48 mGy and 9.41 mGy for cardiac and chest CBCT scans; and 4.59 mGy for diagnostic thoracic CT. The effective doses were 0.2 mSv, 3.2 mSv and 2.8 mSv respectively. For a small patient represented by the anthropomorphic phantom, the dose from the diagnostic CT was lower than from the CBCT scan, as a result of the exposure reduction options available on modern CT scanners. The CBCT protocols used the same fixed scanning techniques. The diagnostic CT dose based on the patient data was 35% higher than the phantom dose. For most SPECT/CT studies the dose from the CBCT component was comparable with the dose from the radiopharmaceutical. Conclusions: The patient radiation dose from the cone beam CT scan can be higher than that from a diagnostic CT and should be taken into consideration in evaluating total SPECT/CT patient dose.« less
  • Lung tumours move due to respiratory motion. This is managed during planning by acquiring a 4DCT and capturing the excursion of the GTV (gross tumour volume) throughout the breathing cycle within an IGTV (Internal Gross Tumour Volume) contour. Patients undergo a verification cone-beam CT (CBCT) scan immediately prior to treatment. 3D reconstructed images do not consider tumour motion, resulting in image artefacts, such as blurring. This may lead to difficulty in identifying the tumour on reconstructed images. It would be valuable to create a 4DCBCT reconstruction of the tumour motion to confirm that does indeed remain within the planned IGTV.more » CBCT projections of a Quasar Respiratory Motion Phantom are acquired in Treatment mode (half-fan scan) on a Varian TrueBeam accelerator. This phantom contains a mobile, low-density lung insert with an embedded 3cm diameter tumour object. It is programmed to create a 15s periodic, 2cm (sup/inf) displacement. A Varian Real-time Position Management (RPM) tracking-box is placed on the phantom breathing platform. Breathing phase information is automatically integrated into the projection image files. Using in-house Matlab programs and RTK (Reconstruction Tool Kit) open-source toolboxes, the projections are re-binned into 10 phases and a 4DCBCT scan reconstructed. The planning IGTV is registered to the 4DCBCT and the tumour excursion is verified to remain within the planned contour. This technique successfully reconstructs 4DCBCT images using clinical modes for a breathing phantom. UBC-BCCA ethics approval has been obtained to perform 4DCBCT reconstructions on lung patients (REB#H12-00192). Clinical images will be accrued starting April 2014.« less
  • We investigate Cherenkov emission (CE) by radiotherapy beams via radiation dose-versus-CE correlation analyses, CE detection optimization by means of a spectral shift towards the near-infrared (NIR) window of biological tissue, and comparison of CE to on-board MV imaging. Dose-CE correlation was investigated via simulation and experiment. A Monte Carlo (MC) CE simulator was designed using Geant4. Experimental phantoms include: water; tissue-simulating phantom composed of water, Intralipid®, and beef blood; plastic phantom with solid water insert. The detector system comprises an optical fiber and diffraction-grating spectrometer incorporating a front/back-illuminated CCD. The NIR shift was carried out with CdSe/ZnS quantum dots (QDs),more » emitting at (650±10) nm. CE and MV images were acquired with a CMOS camera and electronic portal imaging device. MC and experimental studies indicate a strong linear dose-CE correlation (Pearson coefficient > 0.99). CE by an 18-MeV beam was effectively NIR-shifted in water and a tissue-simulating phantom, exhibiting a significant increase at 650 nm for QD depths up to 10 mm. CE images exhibited relative contrast superior to MV images by a factor of 30. Our work supports the potential for application of CE in radiotherapy online imaging for patient setup and treatment verification, since CE is intrinsic to the beam and non-ionizing and QDs can be used to improve CE detectability, potentially yielding image quality superior to MV imaging for the case of low-density-variability, low-optical-attenuation materials (ex: breast/oropharynx). Ongoing work involves microenvironment functionalization of QDs and application of multi-channel spectrometry for simultaneous acquisition of dosimetric and tumor oxygenation signals.« less
  • Effective doses were calculated from the delivery of 6 MV, 15 MV, and 18 MV conventional and intensity-modulated radiation therapy (IMRT) prostate treatment plans. ICRP-60 tissue weighting factors were used for the calculations. Photon doses were measured in phantom for all beam energies. Neutron spectra were measured for 15 MV and 18 MV and ICRP-74 quality conversion factors used to calculate ambient dose equivalents. The ambient dose equivalents were corrected for each tissue using neutron depth dose data from the literature. The depth corrected neutron doses were then used as a measure of the neutron component of the ICRP protectionmore » quantity, organ equivalent dose. IMRT resulted in an increased photon dose to many organs. However, the IMRT treatments resulted in an overall decrease in effective dose compared to conventional radiotherapy. This decrease correlates to the ability of an intensity-modulated field to minimize dose to critical normal structures in close proximity to the treatment volume. In a comparison of the three beam energies used for the IMRT treatments, 6 MV resulted in the lowest effective dose, while 18 MV resulted in the highest effective dose. This is attributed to the large neutron contribution for 18 MV compared to no neutron contribution for 6 MV.« less