Title: TU-F-BRE-03: Application of a Novel Mass-Density Compensation Optimization Method to Improve the Response of a Liquid-Filled Ionization Chamber in Nonstandard Fields

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889311· OSTI ID:22407806
 [1];  [2]
  1. CHUM - Notre Dame, Montreal, QC (Canada)
  2. National Physical Laboratory, Teddington, Middlesex (United Kingdom)

Purpose: To improve the response of the microLion detector (PTW 31018) in small field conditions by optimizing the density of the detector's non-sensitive components using a novel method based on the detector's dose response function. Methods: The central values h(0,0) of the perturbation functions for the microLion detector and a volume of water equivalent to its sensitive volume were calculated using the Monte Carlo user code egs-chamber by scoring the dose absorbed by a 6 MV photon pencil beam incident on their centroids. Values of h(0,0) were plotted as a function of the density of the microLion's graphite electrode with the detector placed in the axial orientation. The optimized density was found by finding the minimal value of h(0,0). Results: A density of 1.37 g/cm3 was found to minimize the perturbation function of the microLion detector. The modified microLion's response was then evaluated in small square fields with sides in the range of 5 – 40 mm and found to be consistent with highly watere-quivalent detectors such as a scintillating detector (Exradin W1) and a generic alanine detector in both axial and radial orientations. Conclusion: This work illustrates a novel method which can used to optimize the design of radiation detectors in small fields. This method should also work with other optimization parameters (e.g. thickness of electrode). Density-compensated detectors have the potential to eliminate the need to evaluate nonstandard field correction factors as described by the IAEA-AAPM formalism (Alfonso et al.) and simplify future dosimetry protocols for SRS/SBRT modalities. Finally, we also expect an improvement in the response of density-compensated detectors for composite IMRT fields.

OSTI ID:
22407806
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English