Title: Sci—Fri PM: Topics — 08: The Role and Benefits of Electromagnetic Needle-Tracking Technologies in Brachytherapy

Abstract

In modern brachytherapy, application of large doses of ionizing radiation in a limited number of fractions is frequent. Furthermore, as with any surgical procedures, brachytherapy is subject to learning curve effects. In this context, there could be advantages of integrating real-time tracking of needles/catheters to existing protocols given the recent prominent advances in tracking technologies. In this work, we review the use of an electromagnetic tracking system (EMTS) based on the second generation Aurora® Planar Field Generator (Northern Digital Inc) and custom design needles (Philips Healthcare) for brachytherapy applications. The position and orientation information is obtained from 5 degrees of freedom sensors. Basic system performance characterization is performed in well-controlled conditions to establish accuracy and reproducibility as well as potential interference from standard brachytherapy equipment. The results show that sensor locations can be tracked to within 0.04mm (la) when located within 26cm of the generator. Orientation accuracy of the needle remained within ±1° in the same region, but rose quickly at larger distances. The errors on position and orientation strongly dependent the sensor position in the characterization volume (500×500×500mm{sup 3}). The presence of an ultrasound probe was shown to have negligible effects on tracking accuracy. The use of EMTS formore » automatic catheter/applicator reconstruction was also explored. Reconstruction time was less than 10 sec/channel and tips identification was within 0.69±0.29mm of the reference values. Finally, we demonstrate that hollow needle designs with special EM adaptation also allow for real-time seed drop position estimation. In phantom experiments showed that drop positions were on average within 1.6±0.9mm of the reference position measured from μCT. Altogether, EMTS offer promising benefits in a wide range of brachytherapy applications.« less

Authors:
; ; ;  [1]; ;  [2];  [3]
  1. Département de Radio-Oncologie et Centre de Recherche du CHU de Québec, CHU de Québec, Québec (Québec), and Département de Physique, de Génie Physique et d'Optique et Centre de recherche en sur le Cancer, Université Laval, Québec (Québec) (Canada)
  2. Biomedical Systems, Philips Group Innovation, High Tech Campus 34 (HTC 34), Eindhoven (Netherlands)
  3. Integrated Clinical Solutions and Marketing, Philips Healthcare, Veenpluis 4-6, Best (Netherlands)
Publication Date:
OSTI Identifier:
22407711
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 8; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; 60 APPLIED LIFE SCIENCES; ACCURACY; BRACHYTHERAPY; ELECTROMAGNETIC FIELDS; MONITORING; PHANTOMS; REVIEWS; SURGERY

Citation Formats

Beaulieu, L., Racine, E., Boutaleb, S., Filion, O., Poulin, E., Hautvast, G., and Binnekamp, D.. Sci—Fri PM: Topics — 08: The Role and Benefits of Electromagnetic Needle-Tracking Technologies in Brachytherapy. United States: N. p., 2014. Web. doi:10.1118/1.4894955.
Beaulieu, L., Racine, E., Boutaleb, S., Filion, O., Poulin, E., Hautvast, G., & Binnekamp, D.. Sci—Fri PM: Topics — 08: The Role and Benefits of Electromagnetic Needle-Tracking Technologies in Brachytherapy. United States. doi:10.1118/1.4894955.
Beaulieu, L., Racine, E., Boutaleb, S., Filion, O., Poulin, E., Hautvast, G., and Binnekamp, D.. Fri . "Sci—Fri PM: Topics — 08: The Role and Benefits of Electromagnetic Needle-Tracking Technologies in Brachytherapy". United States. doi:10.1118/1.4894955.
@article{osti_22407711,
title = {Sci—Fri PM: Topics — 08: The Role and Benefits of Electromagnetic Needle-Tracking Technologies in Brachytherapy},
author = {Beaulieu, L. and Racine, E. and Boutaleb, S. and Filion, O. and Poulin, E. and Hautvast, G. and Binnekamp, D.},
abstractNote = {In modern brachytherapy, application of large doses of ionizing radiation in a limited number of fractions is frequent. Furthermore, as with any surgical procedures, brachytherapy is subject to learning curve effects. In this context, there could be advantages of integrating real-time tracking of needles/catheters to existing protocols given the recent prominent advances in tracking technologies. In this work, we review the use of an electromagnetic tracking system (EMTS) based on the second generation Aurora® Planar Field Generator (Northern Digital Inc) and custom design needles (Philips Healthcare) for brachytherapy applications. The position and orientation information is obtained from 5 degrees of freedom sensors. Basic system performance characterization is performed in well-controlled conditions to establish accuracy and reproducibility as well as potential interference from standard brachytherapy equipment. The results show that sensor locations can be tracked to within 0.04mm (la) when located within 26cm of the generator. Orientation accuracy of the needle remained within ±1° in the same region, but rose quickly at larger distances. The errors on position and orientation strongly dependent the sensor position in the characterization volume (500×500×500mm{sup 3}). The presence of an ultrasound probe was shown to have negligible effects on tracking accuracy. The use of EMTS for automatic catheter/applicator reconstruction was also explored. Reconstruction time was less than 10 sec/channel and tips identification was within 0.69±0.29mm of the reference values. Finally, we demonstrate that hollow needle designs with special EM adaptation also allow for real-time seed drop position estimation. In phantom experiments showed that drop positions were on average within 1.6±0.9mm of the reference position measured from μCT. Altogether, EMTS offer promising benefits in a wide range of brachytherapy applications.},
doi = {10.1118/1.4894955},
journal = {Medical Physics},
number = 8,
volume = 41,
place = {United States},
year = {Fri Aug 15 00:00:00 EDT 2014},
month = {Fri Aug 15 00:00:00 EDT 2014}
}
  • Purpose: We have developed a semi-automated dose accumulation workflow for Head and Neck Cancer (HNC) patients to evaluate volumetric and dosimetric changes that take place during radiotherapy. This work will be used to assess how dosimetric changes affect both toxicity and disease control, hence inform the feasibility and design of a prospective HNC adaptive trial. Methods: RayStation 4.5.2 features deformable image registration (DIR), where structures already defined on the planning CT image set can be deformably mapped onto cone-beam computed tomography (CBCT) images, accounting for daily treatment set-up shifts and changes in patient anatomy. The daily delivered dose can bemore » calculated on each CBCT and mapped back to the planning CT to allow dose accumulation. The process is partially automated using Python scripts developed in collaboration with RaySearch. Results: To date we have performed dose accumulation on 18 HNC patients treated at our institution during 2013–2015 under REB approval. Our semi-automated process establishes clinical feasibility. Generally, dose accumulation for the entire treatment course of one case takes 60–120 minutes: importing all CBCTs requires 20–30 minutes as each patient has 30 to 40 treated fractions; image registration and dose accumulation require 60–90 minutes. This is in contrast to the process without automated scripts where dose accumulation alone would take 3–5 hours. Conclusions: We have developed a reliable workflow for retrospective dose tracking in HNC using RayStation. The process has been validated for HNC patients treated on both Elekta and Varian linacs with CBCTs acquired on XVI and OBI platforms respectively.« less
  • The INTRABEAM system (Carl Zeiss, Oberkochen, Germany) is a miniature x-ray generator for use in intraoperative radiotherapy and brachytherapy. The device accelerates electrons to up to 50 keV, which are then steered down an evacuated needle probe to strike a thin gold target. For accurate dosimetry of the INTRABEAM system, it is important that the photon spectrum be well understood. Measurements based on air-kerma are heavily impacted by photon spectra, particularly for low photon energies due to the large photoelectric contribution in air mass energy absorption coefficient. While low energy photons have little clinical significance at treatment depths, they maymore » have a large effect on air-kerma measurements. In this work, we have developed an EGSnrc-based monte carlo (MC) model of the Zeiss INTRABEAM system to study the source photon spectra and half-value layers (HVLs) of the bare probe and with various spherical applicators. HVLs were calculated using the analytical attenuation of air-kerma spectra. The calculated bare probe spectrum was compared with simulated and measured results taken from literature. Differences in the L-line energies of gold were found between the spectra predicted by EGSnrc and Geant4. This is due to M and N shell averaging during atomic transitions in EGSnrc. The calculated HVLs of the bare probe and spherical applicators are consistent with literature reported measured values.« less
  • The planning for PBSI is done with the patient's ipsilateral arm raised, however, anatomical changes and variations are unavoidable as the patient resumes her daily activities, potentially resulting in significant deviations in implant geometry from the treatment plan. This study aims to quantify the impact of the ipsilateral arm position on the geometry and dosimetry of the implant at eight weeks, evaluated on post-plans using the MIM Symphony™ software (MIM Software, Cleveland, OH). The average dose metrics for the three patients treated at the TBCC thus far using rigid fusion and contour transfer for the arms up position were 76%more » for the CTV V100, 61% for the PTV V100, and 37% for the PTV V200; and for the arms down position 81% for the CTV V100, 64% for the PTV V100, and 42% for the PTV V200. Qualitative analysis of the post-implant CT for one of the three patients showed poor agreement between the seroma contour transferred from the pre-implant CT and the seroma visible on the post-implant CT. To obtain a clinically accurate plan for that patient, contour modifications were used, yielding improved dose metric averages for the arms-up position for all three patients of 87% for the CTV V100, 68% for the PTV V100, and 39% for the PTV V200. Overall, the data available shows that dosimetric parameters increase with the patient's arm down, both in terms of coverage and in terms of the hot spot, and accrual of more patients may confirm this in a larger population.« less
  • The Union for International Cancer Control (UICC) is a membership-based, non-governmental organization with a mandate to “…to unite the cancer community to reduce the global cancer burden, to promote greater equity, and to integrate cancer control into the world health and development agenda.” COMP is an associate member of the UICC. It is well recognized by the UICC that there are major gaps between high, and low and middle income countries, in terms of access to cancer services including access to radiation therapy. In this context, the UICC has developed a Global Task Force on Radiotherapy for Cancer Control withmore » a charge to answer a single question: “What does it cost to close the gap between what exists today and reasonable access to radiotherapy globally?” The Task Force consists of leaders internationally recognized for their radiation treatment related expertise (radiation oncologists, medical physicists, radiation therapists) as well as those with global health and economics specialization. The Task Force has developed three working groups: (1) to look at the global burden of cancer; (2) to look at the infrastructure requirements (facilities, equipment, personnel); and (3) to consider outcomes in terms of numbers of lives saved and palliated patients. A report is due at the World Cancer Congress in December 2014. This presentation reviews the infrastructure considerations under analysis by the second work group. The infrastructure parameters being addressed include capital costs of buildings and equipment and operating costs, which include human resources, equipment servicing and quality control, and general overhead.« less
  • In vitro evidence has suggested that radiation induced bystander effects may enhance non-local cell killing which may influence radiotherapy treatment planning paradigms. This work applies a bystander effect model, which has been derived from published in vitro data, to calculate equivalent uniform dose (EUD) and tumour control probability (TCP) and compare them with predictions from standard linear quadratic (LQ) models that assume a response due only to local absorbed dose. Comparisons between the models were made under increasing dose heterogeneity scenarios. Dose throughout the CTV was modeled with normal distributions, where the degree of heterogeneity was then dictated by changingmore » the standard deviation (SD). The broad assumptions applied in the bystander effect model are intended to place an upper limit on the extent of the results in a clinical context. The bystander model suggests a moderate degree of dose heterogeneity yields as good or better outcome compared to a uniform dose in terms of EUD and TCP. Intermediate risk prostate prescriptions of 78 Gy over 39 fractions had maximum EUD and TCP values at SD of around 5Gy. The plots only dropped below the uniform dose values for SD ∼ 10 Gy, almost 13% of the prescribed dose. The bystander model demonstrates the potential to deviate from the common local LQ model predictions as dose heterogeneity through a prostate CTV is varies. The results suggest the potential for allowing some degree of dose heterogeneity within a CTV, although further investigations of the assumptions of the bystander model are warranted.« less