Title: Fibre optics wavemeters calibration using a self-referenced optical frequency comb

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4904973· OSTI ID:22392304
; ;  [1]
  1. Instituto de Óptica IO-CSIC, C/Serrano 121, 28006 Madrid (Spain)

Self-referenced optical frequency combs enable the measurement of optical frequencies with a very high accuracy, achieving uncertainties close to the atomic clock used as reference (<10{sup −13} s). In this paper, we present the technique for the measurement of laser frequencies for optical communications followed at IO-CSIC and its application to the calibration of two wavemeters in the 1.5 μm optical communication window. Calibration uncertainties down to 12 MHz and 59 MHz were obtained, respectively, for each of the devices. Furthermore, the long-term behaviour of the higher resolution wavemeter was studied during a 750 h period of sustained operation, exhibiting a dispersion in the measurements of 7.72 MHz. Temperature dependence of the device was analysed, enabling to further reduce dispersion down to a 2.15 MHz range, with no significant temporal deviations.

OSTI ID:
22392304
Journal Information:
Review of Scientific Instruments, Vol. 86, Issue 1; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English