Abstract

Wavelet transform based de-noising like wavelet shrinkage, gives the good results in CT. This procedure affects very little the spatial resolution. Some applications are reconstruction methods, while others are a posteriori de-noising methods. De-noising after reconstruction is very difficult because the noise is non-stationary and has unknown distribution. Therefore, methods which work on the sinogram-space don’t have this problem, because they always work over a known noise distribution at this point. On the other hand, the posterior fossa in a head CT is a very complex region for physicians, because it is commonly affected by artifacts and noise which are not eliminated during the reconstruction procedure. This can leads to some false positive evaluations. The purpose of our present work is to compare different wavelet shrinkage de-noising filters to reduce noise, particularly in images of the posterior fossa within CT scans in the sinogram-space. This work describes an experimental search for the best wavelets, to reduce Poisson noise in Computed Tomography (CT) scans. Results showed that de-noising with wavelet filters improved the quality of posterior fossa region in terms of an increased CNR, without noticeable structural distortions.

Authors:
 [1]; ; ;  [2]
  1. Centro de Ingeniería Clínica. Guacalote y Circunvalación, Santa Clara 50200 (Cuba)
  2. Centro de Estudios de Electrónica y Tecnologías de la Información (CEETI), Universidad Central Marta Abreu de las Villas, Carretera a Camajuaní, km. 5 1/2 Santa Clara 54830 (Cuba)
Publication Date:
OSTI Identifier:
22390514
Resource Type:
Journal Article
Resource Relation:
Journal Name: AIP Conference Proceedings; Journal Volume: 1626; Journal Issue: 1; Conference: 13. Mexican Symposium on Medical Physics, Leon, Guanajuato (Mexico), 15-16 Mar 2014; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; CAT SCANNING; COMPARATIVE EVALUATIONS; FILTERS; IMAGE PROCESSING; IMAGES; NOISE; SHRINKAGE; SPACE; SPATIAL RESOLUTION

Citation Formats

Pita-Machado, Reinado, Perez-Diaz, Marlen, E-mail: mperez@uclv.edu.cu, Lorenzo-Ginori, Juan V., E-mail: mperez@uclv.edu.cu, and Bravo-Pino, Rolando, E-mail: mperez@uclv.edu.cu. The use of wavelet filters for reducing noise in posterior fossa Computed Tomography images. United States: N. p., 2014. Web. doi:10.1063/1.4901386.
Pita-Machado, Reinado, Perez-Diaz, Marlen, E-mail: mperez@uclv.edu.cu, Lorenzo-Ginori, Juan V., E-mail: mperez@uclv.edu.cu, & Bravo-Pino, Rolando, E-mail: mperez@uclv.edu.cu. The use of wavelet filters for reducing noise in posterior fossa Computed Tomography images. United States. doi:10.1063/1.4901386.
Pita-Machado, Reinado, Perez-Diaz, Marlen, E-mail: mperez@uclv.edu.cu, Lorenzo-Ginori, Juan V., E-mail: mperez@uclv.edu.cu, and Bravo-Pino, Rolando, E-mail: mperez@uclv.edu.cu. 2014. "The use of wavelet filters for reducing noise in posterior fossa Computed Tomography images". United States. doi:10.1063/1.4901386.
@article{osti_22390514,
title = {The use of wavelet filters for reducing noise in posterior fossa Computed Tomography images},
author = {Pita-Machado, Reinado and Perez-Diaz, Marlen, E-mail: mperez@uclv.edu.cu and Lorenzo-Ginori, Juan V., E-mail: mperez@uclv.edu.cu and Bravo-Pino, Rolando, E-mail: mperez@uclv.edu.cu},
abstractNote = {Wavelet transform based de-noising like wavelet shrinkage, gives the good results in CT. This procedure affects very little the spatial resolution. Some applications are reconstruction methods, while others are a posteriori de-noising methods. De-noising after reconstruction is very difficult because the noise is non-stationary and has unknown distribution. Therefore, methods which work on the sinogram-space don’t have this problem, because they always work over a known noise distribution at this point. On the other hand, the posterior fossa in a head CT is a very complex region for physicians, because it is commonly affected by artifacts and noise which are not eliminated during the reconstruction procedure. This can leads to some false positive evaluations. The purpose of our present work is to compare different wavelet shrinkage de-noising filters to reduce noise, particularly in images of the posterior fossa within CT scans in the sinogram-space. This work describes an experimental search for the best wavelets, to reduce Poisson noise in Computed Tomography (CT) scans. Results showed that de-noising with wavelet filters improved the quality of posterior fossa region in terms of an increased CNR, without noticeable structural distortions.},
doi = {10.1063/1.4901386},
journal = {AIP Conference Proceedings},
number = 1,
volume = 1626,
place = {United States},
year = 2014,
month =
}
  • Purpose: To investigate the possibility of exposure dose reduction of the cone-beam computed tomography (CBCT) in an image guided patient positioning system by using 6 noise suppression filters. Methods: First, a reference dose (RD) and low-dose (LD)-CBCT (X-ray volume imaging system, Elekta Co.) images were acquired with a reference dose of 86.2 mGy (weighted CT dose index: CTDIw) and various low doses of 1.4 to 43.1 mGy, respectively. Second, an automated rigid registration for three axes was performed for estimating setup errors between a planning CT image and the LD-CBCT images, which were processed by 6 noise suppression filters, i.e.,more » averaging filter (AF), median filter (MF), Gaussian filter (GF), bilateral filter (BF), edge preserving smoothing filter (EPF) and adaptive partial median filter (AMF). Third, residual errors representing the patient positioning accuracy were calculated as an Euclidean distance between the setup error vectors estimated using the LD-CBCT image and RD-CBCT image. Finally, the relationships between the residual error and CTDIw were obtained for 6 noise suppression filters, and then the CTDIw for LD-CBCT images processed by the noise suppression filters were measured at the same residual error, which was obtained with the RD-CBCT. This approach was applied to an anthropomorphic pelvic phantom and two cancer patients. Results: For the phantom, the exposure dose could be reduced from 61% (GF) to 78% (AMF) by applying the noise suppression filters to the CBCT images. The exposure dose in a prostate cancer case could be reduced from 8% (AF) to 61% (AMF), and the exposure dose in a lung cancer case could be reduced from 9% (AF) to 37% (AMF). Conclusion: Using noise suppression filters, particularly an adaptive partial median filter, could be feasible to decrease the additional exposure dose to patients in image guided patient positioning systems.« less
  • The errors associated with derivation of cerebral blood flow values by the xenon-enhanced computed tomography (CT) method have been evaluated as a function of tissue heterogeneity and CT noise. The results of this study indicate that CT noise introduces large errors in the derived flow value when data for a single, unprocessed voxel are used for this purpose. CT noise increases the derived flow values in a systematic way. Tissue heterogeneity results in a systematic error which lowers the derived flow values. Errors due to both parameters are computed for typical and extreme conditions.