Title: SU-E-T-508: End to End Testing of a Prototype Eclipse Module for Planning Modulated Arc Therapy On the Siemens Platform

Abstract

Purpose: The latest clinical implementation of the Siemens Artiste linac allows for delivery of modulated arcs (mARC) using full-field flattening filter free (FFF) photon beams. The maximum doserate of 2000 MU/min is well suited for high dose treatments such as SBRT. We tested and report on the performance of a prototype Eclipse TPS module supporting mARC capability on the Artiste platform. Method: our spine SBRT patients originally treated with 12/13 field static-gantry IMRT (SGIMRT) were chosen for this study. These plans were designed to satisfy RTOG0631 guidelines with a prescription of 16Gy in a single fraction. The cases were re-planned as mARC plans in the prototype Eclipse module using the 7MV FFF beam and required to satisfy RTOG0631 requirements. All plans were transferred from Eclipse, delivered on a Siemens Artiste linac and dose-validated using the Delta4 system. Results: All treatment plans were straightforwardly developed, in timely fashion, without challenge or inefficiency using the prototype module. Due to the limited number of segments in a single arc, mARC plans required 2-3 full arcs to yield plan quality comparable to SGIMRT plans containing over 250 total segments. The average (3%/3mm) gamma pass-rate for all arcs was 98.5±1.1%, thus demonstrating both excellent dosemore » prediction by the AAA dose algorithm and excellent delivery fidelity. Mean delivery times for the mARC plans(10.5±1.7min) were 50-70% lower than the SGIMRT plans(26±2min), with both delivered at 2000 MU/min. Conclusion: A prototype Eclipse module capable of planning for Burst Mode modulated arc delivery on the Artiste platform has been tested and found to perform efficiently and accurately for treatment plan development and delivered-dose prediction. Further investigation of more treatment sites is being carried out and data will be presented.« less

Authors:
 [1];  [2];  [3]; ; ;  [4];  [5];  [6]
  1. Huntsman Cancer Hospital, Salt Lake City, UT (United States)
  2. University of Utah Hospitals, Salt Lake City, UT (United States)
  3. Varian Medical Systems France, Buc Cedex (France)
  4. University Utah, Salt Lake City, UT (United States)
  5. University of Utah, Salt Lake City, UT (United States)
  6. Huntsman Cancer Hospital, The University of Utah, Salt Lake City, UT (United States)
Publication Date:
OSTI Identifier:
22369747
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; ALGORITHMS; FORECASTING; IMPLEMENTATION; LINEAR ACCELERATORS; PATIENTS; PERFORMANCE; RADIATION DOSES; RADIOTHERAPY; RECOMMENDATIONS; VERTEBRAE

Citation Formats

Huang, L, Sarkar, V, Spiessens, S, Rassiah-Szegedi, P, Huang, Y, Salter, B, Zhao, H, and Szegedi, M. SU-E-T-508: End to End Testing of a Prototype Eclipse Module for Planning Modulated Arc Therapy On the Siemens Platform. United States: N. p., 2014. Web. doi:10.1118/1.4888841.
Huang, L, Sarkar, V, Spiessens, S, Rassiah-Szegedi, P, Huang, Y, Salter, B, Zhao, H, & Szegedi, M. SU-E-T-508: End to End Testing of a Prototype Eclipse Module for Planning Modulated Arc Therapy On the Siemens Platform. United States. doi:10.1118/1.4888841.
Huang, L, Sarkar, V, Spiessens, S, Rassiah-Szegedi, P, Huang, Y, Salter, B, Zhao, H, and Szegedi, M. Sun . "SU-E-T-508: End to End Testing of a Prototype Eclipse Module for Planning Modulated Arc Therapy On the Siemens Platform". United States. doi:10.1118/1.4888841.
@article{osti_22369747,
title = {SU-E-T-508: End to End Testing of a Prototype Eclipse Module for Planning Modulated Arc Therapy On the Siemens Platform},
author = {Huang, L and Sarkar, V and Spiessens, S and Rassiah-Szegedi, P and Huang, Y and Salter, B and Zhao, H and Szegedi, M},
abstractNote = {Purpose: The latest clinical implementation of the Siemens Artiste linac allows for delivery of modulated arcs (mARC) using full-field flattening filter free (FFF) photon beams. The maximum doserate of 2000 MU/min is well suited for high dose treatments such as SBRT. We tested and report on the performance of a prototype Eclipse TPS module supporting mARC capability on the Artiste platform. Method: our spine SBRT patients originally treated with 12/13 field static-gantry IMRT (SGIMRT) were chosen for this study. These plans were designed to satisfy RTOG0631 guidelines with a prescription of 16Gy in a single fraction. The cases were re-planned as mARC plans in the prototype Eclipse module using the 7MV FFF beam and required to satisfy RTOG0631 requirements. All plans were transferred from Eclipse, delivered on a Siemens Artiste linac and dose-validated using the Delta4 system. Results: All treatment plans were straightforwardly developed, in timely fashion, without challenge or inefficiency using the prototype module. Due to the limited number of segments in a single arc, mARC plans required 2-3 full arcs to yield plan quality comparable to SGIMRT plans containing over 250 total segments. The average (3%/3mm) gamma pass-rate for all arcs was 98.5±1.1%, thus demonstrating both excellent dose prediction by the AAA dose algorithm and excellent delivery fidelity. Mean delivery times for the mARC plans(10.5±1.7min) were 50-70% lower than the SGIMRT plans(26±2min), with both delivered at 2000 MU/min. Conclusion: A prototype Eclipse module capable of planning for Burst Mode modulated arc delivery on the Artiste platform has been tested and found to perform efficiently and accurately for treatment plan development and delivered-dose prediction. Further investigation of more treatment sites is being carried out and data will be presented.},
doi = {10.1118/1.4888841},
journal = {Medical Physics},
number = 6,
volume = 41,
place = {United States},
year = {Sun Jun 01 00:00:00 EDT 2014},
month = {Sun Jun 01 00:00:00 EDT 2014}
}
  • Purpose: To compare and analyze the characteristics of intensity-modulated arc therapy(IMAT) versus fixed-gantry intensity-modulated radiotherapy(IMRT) in treatment of non-small-cell lung cancer. Methods: Twelve patients treated in our radiotherapy center were selected for this study. The patient subsequently underwent 4D-CT simulation.Margins of 5mm and 10mm were added to the ITV to generate the CTV and PTV respectively. Three treatment plans (IMRT,one single arc (RA1),double arcs (RA2))were generated with Eclipse ver.8.6 planning systems. Using a dose level of 75Gy in 15fractions to the ITV,60Gy in 15fractions to the CTV and 45Gy in 15fractions to the PTV respectively. The target and normol tissuemore » volumes were compared,as were the dosimetry parameters. Results: There were no significant differences in CI of ITV,PTV,HI of ITV,CTV and PTV, V5,V10,V15,V20,V25,V30,V45,V50 of total-lung and mean lung dose (all p>0.05). However, the differences were significant in terms of CI of CTV,V5 of B-P (all p<0.05). On the MU, IMRT=1540MU,RA1=1006 MU and RA2=1096 MU. (F=12.00,P=0.000).On the treatment time, IMRT= 13.5min,RA1= 1.5min,and RA2=2.5 min (F= 30.11,P=0.000 ). Conclusion: IMAT is equal to IMRT in dosimetril evaluation. Due to much less Mu and delivery time,IMAT is an ideal technique in treating patients by reduceing the uncomfortable influnce which could effect the treatment.« less
  • Purpose: To compare the plan quality and performance of Simultaneous Integrated Boost (SIB) Treatment plan between Seven field (7F) and Nine field(9F) Intensity Modulated Radiotherapies and Single Arc (SA) and Dual Arc (DA) Volumetric Modulated Arc Therapy( VMAT). Methods: Retrospective planning study of 16 patients treated in Elekta Synergy Platform (mlci2) by 9F-IMRT were replanned with 7F-IMRT, Single Arc VMAT and Dual Arc VMAT using CMS, Monaco Treatment Planning System (TPS) with Monte Carlo simulation. Target delineation done as per Radiation Therapy Oncology Protocols (RTOG 0225&0615). Dose Prescribed as 70Gy to Planning Target Volumes (PTV70) and 61Gy to PTV61 inmore » 33 fraction as a SIB technique. Conformity Index(CI), Homogeneity Index(HI) were used as analysis parameter for Target Volumes as well as Mean dose and Max dose for Organ at Risk(OAR,s).Treatment Delivery Time(min), Monitor unit per fraction (MU/fraction), Patient specific quality assurance were also analysed. Results: A Poor dose coverage and Conformity index (CI) was observed in PTV70 by 7F-IMRT among other techniques. SA-VMAT achieved poor dose coverage in PTV61. No statistical significance difference observed in OAR,s except Spinal cord (P= 0.03) and Right optic nerve (P=0.03). DA-VMAT achieved superior target coverage, higher CI (P =0.02) and Better HI (P=0.03) for PTV70 other techniques (7F-IMRT/9F-IMRT/SA-VMAT). A better dose spare for Parotid glands and spinal cord were seen in DA-VMAT. The average treatment delivery time were 5.82mins, 6.72mins, 3.24mins, 4.3mins for 7F-IMRT, 9F-IMRT, SA-VMAT and DA-VMAT respectively. Significance difference Observed in MU/fr (P <0.001) and Patient quality assurance pass rate were >95% (Gamma analysis (Γ3mm, 3%). Conclusion: DA-VAMT showed better target dose coverage and achieved better or equal performance in sparing OARs among other techniques. SA-VMAT offered least Treatment Time than other techniques but achieved poor target coverage. DA-VMAT offered shorter delivery time than 7F-IMRT and 9F-IMRT without compromising the plan quality.« less
  • Purpose: The delivery of post-mastectomy radiotherapy (PMRT) can be challenging for patients with left-sided breast cancer due to the PTV size and proximity to critical organs. This study investigates the use of protons for PMRT in a clinically-representative cohort of patients, and quantitatively compares volumetric modulated arc therapy (VMAT) to proton therapy to have an evidence-based rationale for selecting a treatment modality for these patients. Methods: Eight left-sided PMRT patients previously treated at our clinic with VMAT were identified for the study. PTVs included the chest wall and regional lymph nodes. Passively scattered (PS) and intensity modulated proton therapy (IMPT)more » plans were constructed using the Eclipse proton planning system. The resulting plans were compared to the original VMAT plan on the basis of PTV coverage; dose homogeneity index (DHI) and conformity index (CI); dose to organs at risk (OAR); tumor control probability (TCP), normal tissue complication probability (NTCP) and secondary cancer complication probability (SCCP). Differences were tested for significance using the paired Student's t-test (p<0.01). Results: All modalities produced clinically acceptable PMRT plans. The comparison demonstrated proton treatment plans provide significantly lower NTCP values for the heart and the lung while maintaining significantly better CI and DHI. At a prescribed dose of 50.4 Gy (RBE) in the PTV, the calculated mean NTCP value for the patients decreased from 1.3% to 0.05% for the whole heart (cardiac mortality) and from 3.8% to 1.1% for the lungs (radiation pneumonitis) for both proton therapy plans from VMAT plans. Both proton modalities showed a significantly lower SCCP for the contralateral breast compared to VMAT. Conclusion: All three plans (VMAT, PS, and IMPT) provide acceptable treatment plans for PMRT. However, proton therapy shows a significant advantage over VMAT with regards to sparing OARs and may be more advantageous for patients with unfavorable anatomy.« less
  • Purpose: To perform the comparison of dose distributions and dosevolume- histograms generated by VMAT and conventional field-in-field technique for left-sided breast and chestwall cancers; to determine whether VMAT offers more dosimetric benefits than does the field-in-field technique. Methods: All VMAT and field-in-filed plans were produced in Eclipse(version 10). Five plans were generated for left-sided breast and leftsided chestwall with supraclavicular nodes, respectively. A clockwise arc (CW) and a counter-clockwise arc (CCW) were used with start and stop angles being 310o±10o and 140o±10o. Collimator angles were 30o for CW and 330o for CCW. The conformity index (CI) is the ratio ofmore » V95% over PTV. The homogeneity index (HI) is the ratio of the difference between D2% and D98% over the prescribed dose. The V5, as an indicator of low dose bath to organs-at-risk, was used for ipsilateral lung, heart, contralateral lung, and contralateral breast. The V20, as an indicator of radiation pneumonitis, was used for ipsilateral lung. Results: Breast/chestwall VMAT delivers much higher low dose bath to ipsilateral lung, contralateral lung and contralateral breast/chestwall for both intact breast and chestwall with nodes. V5 for heart is increased in VMAT plans. V20 for ipsilateral lung is lower in VMAT plans. PTV coverage is similar for both techniques. For one particular chestwall patient with supraclavicular and internal mammary nodes, VMAT offers superior dose coverage of PTVs with slightly more low-dose-wash to heart, contralateral lung and contralateral breast. Conclusion: This study indicates that there is generally no benefit using VMAT for left-sided intact breast, due to large low-dose-bath (5Gy) to normal tissues with insignificant improvement in PTV coverage. Dosimetric benefits will be seen in VMAT plans for some chestwall patients with large size, and/or internal mammary nodes, etc. Whether a chestwall patient is treated with VMAT should be carefully analyzed on an individual basis.« less
  • Purpose: A novel modulated arc therapy technique, mARC, enables superposition of step-and-shoot IMRT segments upon a subset of the optimization points (OPs) of a continuous-arc delivery. We compare two approaches to mARC planning: one with the number of OPs fixed throughout optimization, and another where the planning system determines the number of OPs in the final plan, subject to an upper limit defined at the outset. Methods: Fixed-OP mARC planning was performed for representative cases using Panther v. 5.01 (Prowess, Inc.), while variable-OP mARC planning used Monaco v. 5.00 (Elekta, Inc.). All Monaco planning used an upper limit of 91more » OPs; those OPs with minimal MU were removed during optimization. Plans were delivered, and delivery times recorded, on a Siemens Artiste accelerator using a flat 6MV beam with 300 MU/min rate. Dose distributions measured using ArcCheck (Sun Nuclear Corporation, Inc.) were compared with the plan calculation; the two were deemed consistent if they agreed to within 3.5% in absolute dose and 3.5 mm in distance-to-agreement among > 95% of the diodes within the direct beam. Results: Example cases included a prostate and a head-and-neck planned with a single arc and fraction doses of 1.8 and 2.0 Gy, respectively. Aside from slightly more uniform target dose for the variable-OP plans, the DVHs for the two techniques were similar. For the fixed-OP technique, the number of OPs was 38 and 39, and the delivery time was 228 and 259 seconds, respectively, for the prostate and head-and-neck cases. For the final variable-OP plans, there were 91 and 85 OPs, and the delivery time was 296 and 440 seconds, correspondingly longer than for fixed-OP. Conclusion: For mARC, both the fixed-OP and variable-OP approaches produced comparable-quality plans whose delivery was successfully verified. To keep delivery time per fraction short, a fixed-OP planning approach is preferred.« less