Title: SU-E-T-228: Liquid Ionisation Chamber Array and MicroDiamond Measurements with the CyberKnife System

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888558· OSTI ID:22351063
; ; ;  [1];  [2];  [3]
  1. University of Oldenburg, Oldenburg, Lower Saxony (Germany)
  2. CyberKnife Zentrum Norddeutschland, Guestrow (Germany)
  3. Georg August University, Goettingen, Niedersachsen (Germany)

Purpose: The aim of this study was to measure the dose profile and output factors with a CyberKnife accelerator using a TM60019 microDiamond detector and a 1000SRS liquid chamber array (both PTW Freiburg, Germany). Methods: An MP3 water phantom (PTW, Freiburg) was positioned along the robotic world coordinate system. The TM60019 detector was adjusted to the center of the according fields and the semiconductor axis was aligned with the beam direction. Profiles at 5cm water depth and SSD = 80 cm were measured along the robotic x axis and y axis for the cylindrical collimators of the CyberKnife (diameter 60, 50, 40, 30, 20, 15, 12.5, 10, 7.5 and 5mm). To determine the output factors the dose profile was measured at 0.1 mm steps around the field center to find the maximum dose value. The liquid chamber array (1000SRS) measurement was performed with the same setup, but with RW3 buildup. Results: The 1000SRS measurements closely conform with the TM60019 profile measurement in all profile regions and for all collimator sizes. The profile measurement is influenced by the almost equal spatial resolution of the TM60019 detector (radius of the sensitive area 1.1mm) and of the 1000SRS liquid chamber array (single chamber width 2.3mm). The measured dose profiles have not been corrected for this limited spatial resolution. Rather we purpose to consider that spatial dose averaging over 2 mm wide regions might be justified in view of patient positioning inaccuracies and of the spaces in tissue participating in the biological radiation responses. Conclusion: The 1000SRS data points conform with the TM60019 profile measurements at all profile regions showing the applicability of liquid ion chamber arrays with the CyberKnife system.

OSTI ID:
22351063
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English