Title: SU-E-J-32: Calypso(R) and Laser-Based Localization Systems Comparison for Left-Sided Breast Cancer Patients Using Deep Inspiration Breath Hold

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888084· OSTI ID:22325204
 [1]; ;  [1]; ; ;  [2];  [2]
  1. Northwest Medical Physics Center, Lynnwood, WA (United States)
  2. Seattle Cancer Care Alliance, Seattle, WA (United States)

Purpose: Our institution uses a manual laser-based system for primary localization and verification during radiation treatment of left-sided breast cancer patients using deep inspiration breath hold (DIBH). This primary system was compared with sternum-placed Calypso(R) beacons (Varian Medical Systems, CA). Only intact breast patients are considered for this analysis. Methods: During computed tomography (CT) simulation, patients have BB and Calypso(R) surface beacons positioned sternally and marked for free-breathing and DIBH CTs. During dosimetry planning, BB longitudinal displacement between free breathing and DIBH CT determines laser mark (BH mark) location. Calypso(R) beacon locations from the DIBH CT are entered at the Tracking Station. During Linac simulation and treatment, patients inhale until the cross-hair and/or lasers coincide with the BH Mark, which can be seen using our high quality cameras (Pelco, CA). Daily Calypso(R) displacement values (difference from the DIBH-CT-based plan) are recorded.The displacement mean and standard deviation was calculated for each patient (77 patients, 1845 sessions). An aggregate mean and standard deviation was calculated weighted by the number of patient fractions.Some patients were shifted based on MV ports. A second data set was calculated with Calypso(R) values corrected by these shifts. Results: Mean displacement values indicate agreement within 1±3mm, with improvement for shifted data (Table). Conclusion: Both unshifted and shifted data sets show the Calypso(R) system coincides with the laser system within 1±3mm, demonstrating either localization/verification system will Resultin similar clinical outcomes. Displacement value uncertainty unilaterally reduces when shifts are taken into account.

OSTI ID:
22325204
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English