Title: Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery

Abstract

Purpose: The authors propose a novel optimization model for volumetric modulated arc therapy (VMAT) planning that directly optimizes deliverable leaf trajectories in the treatment plan optimization problem, and eliminates the need for a separate arc-sequencing step. Methods: In this model, a 360° arc is divided into a given number of arc segments in which the leaves move unidirectionally. This facilitates an algorithm that determines the optimal piecewise linear leaf trajectories for each arc segment, which are deliverable in a given treatment time. Multileaf collimator constraints, including maximum leaf speed and interdigitation, are accounted for explicitly. The algorithm is customized to allow for VMAT delivery using constant gantry speed and dose rate, however, the algorithm generalizes to variable gantry speed if beneficial. Results: The authors demonstrate the method for three different tumor sites: a head-and-neck case, a prostate case, and a paraspinal case. The authors first obtain a reference plan for intensity modulated radiotherapy (IMRT) using fluence map optimization and 20 intensity-modulated fields in equally spaced beam directions, which is beyond the standard of care. Modeling the typical clinical setup for the treatment sites considered, IMRT plans using seven or nine beams are also computed. Subsequently, VMAT plans are optimized bymore » dividing the 360° arc into 20 corresponding arc segments. Assuming typical machine parameters (a dose rate of 600 MU/min, and a maximum leaf speed of 3 cm/s), it is demonstrated that the optimized VMAT plans with 2–3 min delivery time are of noticeably better quality than the 7–9 beam IMRT plans. The VMAT plan quality approaches the quality of the 20-beam IMRT benchmark plan for delivery times between 3 and 4 min. Conclusions: The results indicate that high quality treatments can be delivered in a single arc with 20 arc segments if sufficient time is allowed for modulation in each segment.« less

Authors:
;  [1]
  1. Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, 30 Fruit Street, Boston, Massachusetts 02114 (United States)
Publication Date:
OSTI Identifier:
22251298
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 41; Journal Issue: 1; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; ALGORITHMS; BENCHMARKS; COLLIMATORS; DOSE RATES; HEAD; LIMITING VALUES; NECK; NEOPLASMS; OPTIMIZATION; PLANNING; PROSTATE; RADIOTHERAPY; SHIELDING

Citation Formats

Papp, Dávid, E-mail: Papp.David@mgh.harvard.edu, and Unkelbach, Jan. Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery. United States: N. p., 2014. Web. doi:10.1118/1.4835435.
Papp, Dávid, E-mail: Papp.David@mgh.harvard.edu, & Unkelbach, Jan. Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery. United States. doi:10.1118/1.4835435.
Papp, Dávid, E-mail: Papp.David@mgh.harvard.edu, and Unkelbach, Jan. Wed . "Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery". United States. doi:10.1118/1.4835435.
@article{osti_22251298,
title = {Direct leaf trajectory optimization for volumetric modulated arc therapy planning with sliding window delivery},
author = {Papp, Dávid, E-mail: Papp.David@mgh.harvard.edu and Unkelbach, Jan},
abstractNote = {Purpose: The authors propose a novel optimization model for volumetric modulated arc therapy (VMAT) planning that directly optimizes deliverable leaf trajectories in the treatment plan optimization problem, and eliminates the need for a separate arc-sequencing step. Methods: In this model, a 360° arc is divided into a given number of arc segments in which the leaves move unidirectionally. This facilitates an algorithm that determines the optimal piecewise linear leaf trajectories for each arc segment, which are deliverable in a given treatment time. Multileaf collimator constraints, including maximum leaf speed and interdigitation, are accounted for explicitly. The algorithm is customized to allow for VMAT delivery using constant gantry speed and dose rate, however, the algorithm generalizes to variable gantry speed if beneficial. Results: The authors demonstrate the method for three different tumor sites: a head-and-neck case, a prostate case, and a paraspinal case. The authors first obtain a reference plan for intensity modulated radiotherapy (IMRT) using fluence map optimization and 20 intensity-modulated fields in equally spaced beam directions, which is beyond the standard of care. Modeling the typical clinical setup for the treatment sites considered, IMRT plans using seven or nine beams are also computed. Subsequently, VMAT plans are optimized by dividing the 360° arc into 20 corresponding arc segments. Assuming typical machine parameters (a dose rate of 600 MU/min, and a maximum leaf speed of 3 cm/s), it is demonstrated that the optimized VMAT plans with 2–3 min delivery time are of noticeably better quality than the 7–9 beam IMRT plans. The VMAT plan quality approaches the quality of the 20-beam IMRT benchmark plan for delivery times between 3 and 4 min. Conclusions: The results indicate that high quality treatments can be delivered in a single arc with 20 arc segments if sufficient time is allowed for modulation in each segment.},
doi = {10.1118/1.4835435},
journal = {Medical Physics},
number = 1,
volume = 41,
place = {United States},
year = {Wed Jan 15 00:00:00 EST 2014},
month = {Wed Jan 15 00:00:00 EST 2014}
}
  • Purpose: The authors propose a novel optimization model for volumetric modulated arc therapy (VMAT) planning that directly optimizes deliverable leaf trajectories in the treatment plan optimization problem, and eliminates the need for a separate arc-sequencing step. Methods: In this model, a 360° arc is divided into a given number of arc segments in which the leaves move unidirectionally. This facilitates an algorithm that determines the optimal piecewise linear leaf trajectories for each arc segment, which are deliverable in a given treatment time. Multileaf collimator constraints, including maximum leaf speed and interdigitation, are accounted for explicitly. The algorithm is customized tomore » allow for VMAT delivery using constant gantry speed and dose rate, however, the algorithm generalizes to variable gantry speed if beneficial. Results: The authors demonstrate the method for three different tumor sites: a head-and-neck case, a prostate case, and a paraspinal case. The authors first obtain a reference plan for intensity modulated radiotherapy (IMRT) using fluence map optimization and 20 intensity-modulated fields in equally spaced beam directions, which is beyond the standard of care. Modeling the typical clinical setup for the treatment sites considered, IMRT plans using seven or nine beams are also computed. Subsequently, VMAT plans are optimized by dividing the 360° arc into 20 corresponding arc segments. Assuming typical machine parameters (a dose rate of 600 MU/min, and a maximum leaf speed of 3 cm/s), it is demonstrated that the optimized VMAT plans with 2–3 min delivery time are of noticeably better quality than the 7–9 beam IMRT plans. The VMAT plan quality approaches the quality of the 20-beam IMRT benchmark plan for delivery times between 3 and 4 min. Conclusions: The results indicate that high quality treatments can be delivered in a single arc with 20 arc segments if sufficient time is allowed for modulation in each segment.« less
  • Purpose: To assess the potential benefit of trajectory modulated arc therapy (TMAT) for treatments of small benign intracranial tumor, pituitary adenoma. Methods: A TMAT planning platform that incorporates complex source motion trajectory involving synchronized gantry rotation with translational and rotational couch movement was used for the study. The platform couples an interactive trajectory generation tool with a VMAT algorithm that performs multi-resolution, progressive sampling MLC optimization on a user-designed trajectory. A continuous couch rotation of 160° angular span with ±20° mini gantry arcs was used to emulate a non-coplanar horizontal arc-like trajectory. Compared to conventional non-coplanar gantry arcs (60°-100° gantrymore » rotation with couch kicks), TMAT limited the unnecessary low to medium dose spread in the anterior and posterior directions, where primary OARs (e.g., brainstem, optic chiasm, optic nerves, and lens) are in close proximity to the targeted pituitary tumor volume. For 5 standard fractionation pituitary adenoma cases (50.4Gy/28fractions), TMAT and non-coplanar VMAT plans were generated and compared under equivalent objectives/constraints. TMAT delivery was implemented and demonstrated on Varian TrueBeam via XML scripts. Results: Both techniques showed good target coverage while OARs were able to meet the constraints on QUANTEC guidelines. Notably, TMAT decreased the dose deposition in the anterior-to-posterior direction surrounding PTV. TMAT significantly reduced the mean doses on brainstem, optic nerves, eyes and lens by 47.29%±13.17%, 28.51%±8.68%, 80.82%±8.71% and 65.38%±19.99% compared with VMAT, all p≤0.01. Percentage reductions of maximum point dose in eyes and lens were 75.68%±10.30% and 70.72%±18.62% respectively for TMAT versus VMAT, all p≤0.01. A representative isocentric TMAT pituitary plan was delivered via an XML script with 200 control points and 282 MUs. Conclusion: Deliverable TMAT plans were achieved in developer mode in TrueBeam. TMAT was shown to be superior for pituitary adenoma irradiation in terms of OARs sparing.« less
  • This article is aimed to compare the dosimetric differences between volumetric modulated arc therapy (VMAT) and intensity-modulated radiotherapy (IMRT) for Stage I-II nasal natural killer/T-cell lymphoma (NNKTL). Ten patients with Stage I-II NNKTL treated with IMRT were replanned with VMAT (2 arcs). The prescribed dose of the planning target volume (PTV) was 50 Gy in 25 fractions. The VMAT plans with the Anisotropic Analytical Algorithm (Version 8.6.15) were based on an Eclipse treatment planning system; the monitor units (MUs) and treatment time (T) were scored to measure the expected treatment efficiency. All the 10 patients under the study were subjectmore » to comparisons regarding the quality of target coverage, the efficiency of delivery, and the exposure of normal adjacent organs at risk (OARs). The study shows that VMAT was associated with a better conformal index (CI) and homogeneity index (HI) (both p < 0.05) but slightly higher dose to OARs than IMRT. The MUs with VMAT (650.80 ± 24.59) were fewer than with IMRT (1300.10 ± 57.12) (relative reduction of 49.94%, p = 0.00) when using 2-Gy dose fractions. The treatment time with VMAT (3.20 ± 0.02 minutes) was shorter than with IMRT (7.38 ± 0.18 minutes) (relative reduction of 56.64%, p = 0.00). We found that VMAT and IMRT both provide satisfactory target dosimetric coverage and OARs sparing clinically. Likely to deliver a bit higher dose to OARs, VMAT in comparison with IMRT, is still a better choice for treatment of patients with Stage I-II NNKTL, thanks to better dose distribution, fewer MUs, and shorter delivery time.« less
  • This study aims to evaluate treatment plans generated by Step-and-Shoot (SS), Sliding Window (SW) and Volumetric Modulated Arc Therapy (VMAT) in order to assess the differences in dose volume histograms of planning target volume (PTV) and organs at risk (OAR), conformity indices, radiobiological evaluations, and plan quality for prostate cancer cases. Six prostate cancer patients treated in our center were selected for this retrospective study. Treatment plans were generated with Eclipse version 8.9 using 10 MV photon beams. For VMAT, Varian Rapid Arc with 1 or 2 arcs, and for SS and SW IMRT, 7-9 fields were used. Each planmore » had three PTVs with prescription doses of 81, 59.4, and 45 Gy to prostate, to prostate and lymph nodes, and to pelvis, respectively. Doses to PTV and OAR and the conformal indices (COIN) were compared among three techniques. The equivalent uniform dose (EUD), tumor control probability (TCP) and normal tissue complication probability (NTCP) were calculated and compared. The mean doses to the PTV prostate on average were 83 Gy and the percent differences of mean dose among all techniques were below 0.28. For bladder and rectum, the percent differences of mean dose among all techniques were below 2.2. The COIN did not favour any particular delivery method over the other. The TCP was higher with SS and SW for four patients and higher with VMAT for two patients. The NTCP for the rectum was the lowest with VMAT in five out of the six patients. The results show similar target coverage in general.« less
  • Purpose: To perform a comparison of two pelvic lymph node volume delineation strategies used in intensity-modulated radiotherapy (IMRT) for high risk prostate cancer and to determine the role of volumetric modulated arc therapy (VMAT). Methods and Materials: Eighteen consecutive patients accrued to an ongoing clinical trial were identified according to either the nodal contouring strategy as described based on lymphotropic nanoparticle-enhanced magnetic resonance imaging technology (9 patients) or the current Radiation Therapy Oncology Group (RTOG) consensus guidelines (9 patients). Radiation consisted of 45 Gy to prostate, seminal vesicles, and lymph nodes, with a simultaneous integrated boost to the prostate alone,more » to a total dose of 67.5 Gy delivered in 25 fractions. Prospective acute genitourinary and gastrointestinal toxicities were compared at baseline, during radiotherapy, and 3 months after radiotherapy. Each patient was retrospectively replanned using the opposite method of nodal contouring, and plans were normalized for dosimetric comparison. VMAT plans were also generated according to the RTOG method for comparison. Results: RTOG plans resulted in a significantly lower rate of genitourinary frequency 3 months after treatment. The dosimetric comparison showed that the RTOG plans resulted in both favorable planning target volume (PTV) coverage and lower organs at risk (OARs) and integral (ID) doses. VMAT required two to three arcs to achieve adequate treatment plans, we did not observe consistent dosimetric benefits to either the PTV or the OARs, and a higher ID was observed. However, treatment times were significantly shorter with VMAT. Conclusion: The RTOG guidelines for pelvic nodal volume delineation results in favorable dosimetry and acceptable acute toxicities for both the target and OARs. We are unable to conclude that VMAT provides a benefit compared with IMRT.« less