skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Interfraction rotation of the prostate as evaluated by kilovoltage X-ray fiducial marker imaging in intensity-modulated radiotherapy of localized prostate cancer

Journal Article · · Medical Dosimetry
; ;  [1];  [1]
  1. Charite Universitaetsmedizin Berlin, Department of Radiation Oncology, Campus Virchow-Klinikum, Berlin (Germany)

To quantify the daily rotation of the prostate during a radiotherapy course using stereoscopic kilovoltage (kV) x-ray imaging and intraprostatic fiducials for localization and positioning correction. From 2005 to 2009, radio-opaque fiducial markers were inserted into 38 patients via perineum into the prostate. The ExacTrac/Novalis Body X-ray 6-day image acquisition system (ET/NB; BrainLab AG, Feldkirchen, Germany) was used to determine and correct the target position. During the first period in 10 patients we recorded all rotation errors but used only Y (table) for correction. For the next 28 patients we used for correction all rotational coordinates, i.e., in addition Z (superior-inferior [SI] or roll) and X (left-right [LR] or tilt/pitch) according to the fiducial marker position by use of the Robotic Tilt Module and Varian Exact Couch. Rotation correction was applied above a threshold of 1 Degree-Sign displacement. The systematic and random errors were specified. Overall, 993 software-assisted rotational corrections were performed. The interfraction rotation errors of the prostate as assessed from the radiodense surrogate markers around the three axes Y, Z, and X were on average 0.09, -0.52, and -0.01 Degree-Sign with standard deviations of 2.01, 2.30, and 3.95 Degree-Sign , respectively. The systematic uncertainty per patient for prostate rotation was estimated with 2.30, 1.56, and 4.13 Degree-Sign and the mean random components with 1.81, 2.02, and 3.09 Degree-Sign . The largest rotational errors occurred around the X-axis (pitch), but without preferring a certain orientation. Although the error around Z (roll) can be compensated on average by a transformation with 4 coordinates, a significant error around X remains and advocates the full correction with 6 coordinates. Rotational errors as assessed via daily stereoscopic online imaging are significant and dominate around X. Rotation possibly degrades the dosimetric coverage of the target volume and may require suitable strategies for correction.

OSTI ID:
22131254
Journal Information:
Medical Dosimetry, Vol. 37, Issue 4; Other Information: Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
Country of Publication:
United States
Language:
English