Title: Dose calculation for electron therapy using an improved LBR method

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4810938· OSTI ID:22121618
;  [1];  [1]
  1. Department of Physics and Astronomy, University of South Carolina, Columbia, South Carolina 29208 (United States)

Purpose: To calculate the percentage depth dose (PDD) of any irregularly shaped electron beam using a modified lateral build-up ratio (LBR) method.Methods: Percentage depth dose curves were measured using 6, 9, 12, and 15 MeV electron beam energies for applicator cone sizes of 6 Multiplication-Sign 6, 10 Multiplication-Sign 10, 14 Multiplication-Sign 14, and 20 Multiplication-Sign 20 cm{sup 2}. Circular cutouts for each cone were prepared from 2.0 cm diameter to the maximum possible size for each cone. In addition, three irregular cutouts were prepared.Results: The LBR for each circular cutout was calculated from the measured PDD curve using the open field of the 14 Multiplication-Sign 14 cm{sup 2} cone as the reference field. Using the LBR values and the radius of the circular cutouts, the corresponding lateral spread parameter [{sigma}{sub R}(z)] of the electron shower was calculated. Unlike the commonly accepted assumption that {sigma}{sub R}(z) is independent of cutout size, it is shown that its value increases linearly with circular cutout size (R). Using this characteristic of the lateral spread parameter, the PDD curves of irregularly shaped cutouts were calculated. Finally, the calculated PDD curves were compared with measured PDD curves.Conclusions: In this research, it is shown that the lateral spread parameter {sigma}{sub R}(z) increases with cutout size. For radii of circular cutout sizes up to the equilibrium range of the electron beam, the increase of {sigma}{sub R}(z) with the cutout size is linear. The percentage difference of the calculated PDD curve from the measured PDD data for irregularly shaped cutouts was under 1.0% in the region between the surface and therapeutic range of the electron beam. Similar results were obtained for four electron beam energies (6, 9, 12, and 15 MeV)

OSTI ID:
22121618
Journal Information:
Medical Physics, Vol. 40, Issue 7; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English