Title: Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: A technical feasibility study

Abstract

Purpose: Target sites affected by organ motion require a time resolved (4D) dose calculation. Typical 4D dose calculations use 4D-CT as a basis. Unfortunately, 4D-CT images have the disadvantage of being a 'snap-shot' of the motion during acquisition and of assuming regularity of breathing. In addition, 4D-CT acquisitions involve a substantial additional dose burden to the patient making many, repeated 4D-CT acquisitions undesirable. Here the authors test the feasibility of an alternative approach to generate patient specific 4D-CT data sets. Methods: In this approach motion information is extracted from 4D-MRI. Simulated 4D-CT data sets [which the authors call 4D-CT(MRI)] are created by warping extracted deformation fields to a static 3D-CT data set. The employment of 4D-MRI sequences for this has the advantage that no assumptions on breathing regularity are made, irregularities in breathing can be studied and, if necessary, many repeat imaging studies (and consequently simulated 4D-CT data sets) can be performed on patients and/or volunteers. The accuracy of 4D-CT(MRI)s has been validated by 4D proton dose calculations. Our 4D dose algorithm takes into account displacements as well as deformations on the originating 4D-CT/4D-CT(MRI) by calculating the dose of each pencil beam based on an individual time stamp of whenmore » that pencil beam is applied. According to corresponding displacement and density-variation-maps the position and the water equivalent range of the dose grid points is adjusted at each time instance. Results: 4D dose distributions, using 4D-CT(MRI) data sets as input were compared to results based on a reference conventional 4D-CT data set capturing similar motion characteristics. Almost identical 4D dose distributions could be achieved, even though scanned proton beams are very sensitive to small differences in the patient geometry. In addition, 4D dose calculations have been performed on the same patient, but using 4D-CT(MRI) data sets based on variable breathing patterns to show the effect of possible irregular breathing on active scanned proton therapy. Using a 4D-CT(MRI), including motion irregularities, resulted in significantly different proton dose distributions. Conclusions: The authors have demonstrated that motion information from 4D-MRI can be used to generate realistic 4D-CT data sets on the basis of a single static 3D-CT data set. 4D-CT(MRI) presents a novel approach to test the robustness of treatment plans in the circumstance of patient motion.« less

Authors:
 [1];  [2];  [1];  [2];  [1]
  1. Center for Proton Therapy, Paul Scherrer Institut, 5232 Villigen-PSI (Switzerland)
  2. (Switzerland)
Publication Date:
OSTI Identifier:
22121018
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 40; Journal Issue: 6; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; ALGORITHMS; COMPUTERIZED TOMOGRAPHY; FEASIBILITY STUDIES; NMR IMAGING; PATIENTS; PROTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; RESPIRATION; TIME RESOLUTION

Citation Formats

Boye, Dirk, Computer Vision Laboratory, ETH Zuerich, 8092 Zuerich, Lomax, Tony, ETH Zuerich, 8092 Zuerich, and Knopf, Antje. Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: A technical feasibility study. United States: N. p., 2013. Web. doi:10.1118/1.4801914.
Boye, Dirk, Computer Vision Laboratory, ETH Zuerich, 8092 Zuerich, Lomax, Tony, ETH Zuerich, 8092 Zuerich, & Knopf, Antje. Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: A technical feasibility study. United States. doi:10.1118/1.4801914.
Boye, Dirk, Computer Vision Laboratory, ETH Zuerich, 8092 Zuerich, Lomax, Tony, ETH Zuerich, 8092 Zuerich, and Knopf, Antje. 2013. "Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: A technical feasibility study". United States. doi:10.1118/1.4801914.
@article{osti_22121018,
title = {Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: A technical feasibility study},
author = {Boye, Dirk and Computer Vision Laboratory, ETH Zuerich, 8092 Zuerich and Lomax, Tony and ETH Zuerich, 8092 Zuerich and Knopf, Antje},
abstractNote = {Purpose: Target sites affected by organ motion require a time resolved (4D) dose calculation. Typical 4D dose calculations use 4D-CT as a basis. Unfortunately, 4D-CT images have the disadvantage of being a 'snap-shot' of the motion during acquisition and of assuming regularity of breathing. In addition, 4D-CT acquisitions involve a substantial additional dose burden to the patient making many, repeated 4D-CT acquisitions undesirable. Here the authors test the feasibility of an alternative approach to generate patient specific 4D-CT data sets. Methods: In this approach motion information is extracted from 4D-MRI. Simulated 4D-CT data sets [which the authors call 4D-CT(MRI)] are created by warping extracted deformation fields to a static 3D-CT data set. The employment of 4D-MRI sequences for this has the advantage that no assumptions on breathing regularity are made, irregularities in breathing can be studied and, if necessary, many repeat imaging studies (and consequently simulated 4D-CT data sets) can be performed on patients and/or volunteers. The accuracy of 4D-CT(MRI)s has been validated by 4D proton dose calculations. Our 4D dose algorithm takes into account displacements as well as deformations on the originating 4D-CT/4D-CT(MRI) by calculating the dose of each pencil beam based on an individual time stamp of when that pencil beam is applied. According to corresponding displacement and density-variation-maps the position and the water equivalent range of the dose grid points is adjusted at each time instance. Results: 4D dose distributions, using 4D-CT(MRI) data sets as input were compared to results based on a reference conventional 4D-CT data set capturing similar motion characteristics. Almost identical 4D dose distributions could be achieved, even though scanned proton beams are very sensitive to small differences in the patient geometry. In addition, 4D dose calculations have been performed on the same patient, but using 4D-CT(MRI) data sets based on variable breathing patterns to show the effect of possible irregular breathing on active scanned proton therapy. Using a 4D-CT(MRI), including motion irregularities, resulted in significantly different proton dose distributions. Conclusions: The authors have demonstrated that motion information from 4D-MRI can be used to generate realistic 4D-CT data sets on the basis of a single static 3D-CT data set. 4D-CT(MRI) presents a novel approach to test the robustness of treatment plans in the circumstance of patient motion.},
doi = {10.1118/1.4801914},
journal = {Medical Physics},
number = 6,
volume = 40,
place = {United States},
year = 2013,
month = 6
}
  • Purpose: The dose difference between three-dimensional dose (3D dose) and 4D dose which considers motion due to respiratory can be varied according to geometrical relationship between planning target volume (PTV) and organ at risk (OAR). The purpose of the study is to investigate the dose difference between 3D and 4D dose using overlap volume histogram (OVH) which is an indicator that quantify geometrical relationship between a PTV and an OAR. Methods: Five liver cancer patients who previously treated stereotactic body radiotherapy (SBRT) were investigated. Four-dimensional computed tomography (4DCT) images were acquired for all patients. ITV-based treatment planning was performed. 3Dmore » dose was calculated on the end-exhale phase image as a reference phase image. 4D dose accumulation was implemented from all phase images using dose warping technique used deformable image registration (DIR) algorithm (Horn and Schunck optical flow) in DIRART. In this study OVH was used to quantify geometrical relationship between a PTV and an OAR. OVH between a PTV and a selected OAR was generated for each patient case and compared for all cases. The dose difference between 3D and 4D dose for normal organ was calculated and compared for all cases according to OVH. Results: The 3D and 4D dose difference for OAR was analyzed using dose-volume histogram (DVH). On the basis of a specific point which corresponds to 10% of OAR volume overlapped with expanded PTV, mean dose difference was 34.56% in minimum OVH distance case and 13.36% in maximum OVH distance case. As the OVH distance increased, mean dose difference between 4D and 3D dose was decreased. Conclusion: The tendency of dose difference variation was verified according to OVH. OVH is seems to be indicator that has a potential to predict the dose difference between 4D and 3D dose. This work was supported by the Radiation Technology R&D program (No. 2013M2A2A7043498) and the Mid-career Researcher Program (2014R1A2A1A10050270) through the National Research Foundation of Korea funded by the Ministry of Science, ICT&Future Planning.« less
  • Purpose: The aim of this study was to evaluate the appropriateness of using computed tomography (CT) to cone-beam CT (CBCT) deformable image registration (DIR) for the application of calculating the “dose of the day” received by a head and neck patient. Methods: NiftyReg is an open-source registration package implemented in our institution. The affine registration uses a Block Matching-based approach, while the deformable registration is a GPU implementation of the popular B-spline Free Form Deformation algorithm. Two independent tests were performed to assess the suitability of our registrations methodology for “dose of the day” calculations in a deformed CT. Amore » geometric evaluation was performed to assess the ability of the DIR method to map identical structures between the CT and CBCT datasets. Features delineated in the planning CT were warped and compared with features manually drawn on the CBCT. The authors computed the dice similarity coefficient (DSC), distance transformation, and centre of mass distance between features. A dosimetric evaluation was performed to evaluate the clinical significance of the registrations errors in the application proposed and to identify the limitations of the approximations used. Dose calculations for the same intensity-modulated radiation therapy plan on the deformed CT and replan CT were compared. Dose distributions were compared in terms of dose differences (DD), gamma analysis, target coverage, and dose volume histograms (DVHs). Doses calculated in a rigidly aligned CT and directly in an extended CBCT were also evaluated. Results: A mean value of 0.850 in DSC was achieved in overlap between manually delineated and warped features, with the distance between surfaces being less than 2 mm on over 90% of the pixels. Deformable registration was clearly superior to rigid registration in mapping identical structures between the two datasets. The dose recalculated in the deformed CT is a good match to the dose calculated on a replan CT. The DD is smaller than 2% of the prescribed dose on 90% of the body's voxels and it passes a 2% and 2 mm gamma-test on over 95% of the voxels. Target coverage similarity was assessed in terms of the 95%-isodose volumes. A mean value of 0.962 was obtained for the DSC, while the distance between surfaces is less than 2 mm in 95.4% of the pixels. The method proposed provided adequate dose estimation, closer to the gold standard than the other two approaches. Differences in DVH curves were mainly due to differences in the OARs definition (manual vs warped) and not due to differences in dose estimation (dose calculated in replan CT vs dose calculated in deformed CT). Conclusions: Deforming a planning CT to match a daily CBCT provides the tools needed for the calculation of the “dose of the day” without the need to acquire a new CT. The initial clinical application of our method will be weekly offline calculations of the “dose of the day,” and use this information to inform adaptive radiotherapy (ART). The work here presented is a first step into a full implementation of a “dose-driven” online ART.« less
  • Purpose: The introduction of radioembolization with microspheres represents a significant step forward in the treatment of patients with metastatic disease to the liver. This technique uses semiempirical formulae based on body surface area or liver and target volumes to calculate the required total activity for a given patient. However, this treatment modality lacks extremely important information, which is the three-dimensional (3D) dose delivered by microspheres to different organs after their administration. The absence of this information dramatically limits the clinical efficacy of this modality, specifically the predictive power of the treatment. Therefore, the aim of this study is to developmore » a 3D dose calculation technique that is based on the PET imaging of the infused microspheres.Methods: The Fluka Monte Carlo code was used to calculate the voxel dose kernel for {sup 90}Y source with voxel size equal to that of the PET scan. The measured PET activity distribution was converted to total activity distribution for the subsequent convolution with the voxel dose kernel to obtain the 3D dose distribution. In addition, dose-volume histograms were generated to analyze the dose to the tumor and critical structures.Results: The 3D inpatient dose distribution can be reconstructed from the PET data of a patient scanned after the infusion of microspheres. A total of seven patients have been analyzed so far using the proposed reconstruction method. Four patients underwent treatment with SIR-Spheres for liver metastases from colorectal cancer and three patients were treated with Therasphere for hepatocellular cancer. A total of 14 target tumors were contoured on post-treatment PET-CT scans for dosimetric evaluation. Mean prescription activity was 1.7 GBq (range: 0.58–3.8 GBq). The resulting mean maximum measured dose to targets was 167 Gy (range: 71–311 Gy). Mean minimum dose to 70% of target (D70) was 68 Gy (range: 25–155 Gy). Mean minimum dose to 90% of target (D90) was 53 Gy (range: 13–125 Gy).Conclusions: A three-dimensional inpatient dose reconstruction method has been developed that is based on the PET/CT data of a patient treated with {sup 90}Y microspheres. It allows for a complete description of the absorbed dose by the tumor and critical structures. It represents the first step in building predictive models for treatment outcomes for patients receiving this therapeutic modality as well as it allows for better analysis of patients' dose response and will ultimately improve future treatment administration.« less
  • Purpose: To assess dose calculated by the 3DVH software (Sun Nuclear Systems, Melbourne, FL) against TLD measurements and treatment planning system calculations in anthropomorphic phantoms. Methods: The IROC Houston (RPC) head and neck (HN) and lung phantoms were scanned and plans were generated using Eclipse (Varian Medical Systems, Milpitas, CA) following IROC Houston procedures. For the H and N phantom, 6 MV VMAT and 9-field dynamic MLC (DMLC) plans were created. For the lung phantom 6 MV VMAT and 15 MV 9-field dynamic MLC (DMLC) plans were created. The plans were delivered to the phantoms and to an ArcCHECK (Sunmore » Nuclear Systems, Melbourne, FL). The head and neck phantom contained 8 TLDs located at PTV1 (4), PTV2 (2), and OAR Cord (2). The lung phantom contained 4 TLDs, 2 in the PTV, 1 in the cord, and 1 in the heart. Daily outputs were recorded before each measurement for correction. 3DVH dose reconstruction software was used to project the calculated dose to patient anatomy. Results: For the HN phantom, the maximum difference between 3DVH and TLDs was -3.4% and between 3DVH and Eclipse was 1.2%. For the lung plan the maximum difference between 3DVH and TLDs was 4.3%, except for the spinal cord for which 3DVH overestimated the TLD dose by 12%. The maximum difference between 3DVH and Eclipse was 0.3%. 3DVH agreed well with Eclipse because the dose reconstruction algorithm uses the diode measurements to perturb the dose calculated by the treatment planning system; therefore, if there is a problem in the modeling or heterogeneity correction, it will be carried through to 3DVH. Conclusion: 3DVH agreed well with Eclipse and TLD measurements. Comparison of 3DVH with film measurements is ongoing. Work supported by PHS grant CA10953 and CA81647 (NCI, DHHS)« less
  • While each of the three different screening comparisons made in this report (i.e., within-medium evaluation, between-media evaluation and relative importance grouping) individually provides information potentially of value in focusing future studies, each one is subject to a variety of limitations, the most important being associated with the absence or variable quality of environmental data for a number of the contaminants and media. These screening exercises are intended to provide an initial framework for approaching the study of an extremely complex site. Other approaches could very well yield somewhat different priorities, and the identification or reinterpretation of data in subsequent detailedmore » studies are likely to invalidate some of the results of these screening exercises. However, these evaluations provide a logical approach to defining initial off- site health impact study priorities for the ORR. Therefore, while care must be taken in attempting to make any broad generalizations or greatly simplifying assumptions with regard to the potential health hazards posed by the complex releases from the Reservation, Table 6-1 represents an attempt to summarize a set of recommendations that are derived from the screening exercises presented in this report. Table 6-1 identifies the facilities, processes and contaminants believed to have the highest potential for resulting in off-site health impacts. Table 6-2 identifies contaminants for which no ranking could be performed as part of this feasibility study, because of the absence of any appropriate data for any environmental medium.« less