Magnetostatic focal spot correction for x-ray tubes operating in strong magnetic fields using iterative optimization
- Department of Radiology, Stanford University, Stanford, California 94305 (United States) and Department of Bioengineering, Stanford University, Stanford, California 94305 (United States)
Purpose: Combining x-ray fluoroscopy and MR imaging systems for guidance of interventional procedures has become more commonplace. By designing an x-ray tube that is immune to the magnetic fields outside of the MR bore, the two systems can be placed in close proximity to each other. A major obstacle to robust x-ray tube design is correcting for the effects of the magnetic fields on the x-ray tube focal spot. A potential solution is to design active shielding that locally cancels the magnetic fields near the focal spot. Methods: An iterative optimization algorithm is implemented to design resistive active shielding coils that will be placed outside the x-ray tube insert. The optimization procedure attempts to minimize the power consumption of the shielding coils while satisfying magnetic field homogeneity constraints. The algorithm is composed of a linear programming step and a nonlinear programming step that are interleaved with each other. The coil results are verified using a finite element space charge simulation of the electron beam inside the x-ray tube. To alleviate heating concerns an optimized coil solution is derived that includes a neodymium permanent magnet. Any demagnetization of the permanent magnet is calculated prior to solving for the optimized coils. The temperature dynamics of the coil solutions are calculated using a lumped parameter model, which is used to estimate operation times of the coils before temperature failure. Results: For a magnetic field strength of 88 mT, the algorithm solves for coils that consume 588 A/cm{sup 2}. This specific coil geometry can operate for 15 min continuously before reaching temperature failure. By including a neodymium magnet in the design the current density drops to 337 A/cm{sup 2}, which increases the operation time to 59 min. Space charge simulations verify that the coil designs are effective, but for oblique x-ray tube geometries there is still distortion of the focal spot shape along with deflections of approximately 3 mm in the radial and circumferential directions on the anode. Conclusions: Active shielding is an attractive solution for correcting the effects of magnetic fields on the x-ray focal spot. If extremely long fluoroscopic exposure times are required, longer operation times can be achieved by including a permanent magnet with the active shielding design.
- OSTI ID:
- 22099009
- Journal Information:
- Medical Physics, Vol. 39, Issue 9; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Effect of longitudinal magnetic fields on a simulated in-line 6 MV linac
Method for measuring the focal spot size of an x-ray tube using a coded aperture mask and a digital detector
Related Subjects
71 CLASSICAL AND QUANTUM MECHANICS
GENERAL PHYSICS
ALGORITHMS
CURRENT DENSITY
DEMAGNETIZATION
ELECTRON BEAMS
FINITE ELEMENT METHOD
FLUOROSCOPY
HEATING
ITERATIVE METHODS
MAGNETIC FIELDS
NEODYMIUM
NMR IMAGING
OPTIMIZATION
PERMANENT MAGNETS
SHIELDING
SIMULATION
X RADIATION
X-RAY TUBES