Fast-MICP for frameless image-guided surgery
- Department of Electrical Engineering, Chang Gung University, Tao-Yuan 333, Taiwan (China)
Purpose: In image-guided surgery (IGS) systems, image-to-physical registration is critical for reliable anatomical information mapping and spatial guidance. Conventional stereotactic frame-based or fiducial-based approaches provide accurate registration but are not patient-friendly. This study proposes a frameless cranial IGS system that uses computer vision techniques to replace the frame or fiducials with the natural features of the patient. Methods: To perform a cranial surgery with the proposed system, the facial surface of the patient is first reconstructed by stereo vision. Accuracy is ensured by capturing parallel-line patterns projected from a calibrated LCD projector. Meanwhile, another facial surface is reconstructed from preoperative computed tomography (CT) images of the patient. The proposed iterative closest point (ICP)-based algorithm [fast marker-added ICP (Fast-MICP)] is then used to register the two facial data sets, which transfers the anatomical information from the CT images to the physical space. Results: Experimental results reveal that the Fast-MICP algorithm reduces the computational cost of marker-added ICP (J.-D. Lee et al., ''A coarse-to-fine surface registration algorithm for frameless brain surgery,'' in Proceedings of International Conference of the IEEE Engineering in Medicine and Biology Society, 2007, pp. 836-839) to 10% and achieves comparable registration accuracy, which is under 3 mm target registration error (TRE). Moreover, two types of optical-based spatial digitizing devices can be integrated for further surgical navigation. Anatomical information or image-guided surgical landmarks can be projected onto the patient to obtain an immersive augmented reality environment. Conclusion: The proposed frameless IGS system with stereo vision obtains TRE of less than 3 mm. The proposed Fast-MICP registration algorithm reduces registration time by 90% without compromising accuracy.
- OSTI ID:
- 22098522
- Journal Information:
- Medical Physics, Vol. 37, Issue 9; Other Information: (c) 2010 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Calibration of time-of-flight cameras for accurate intraoperative surface reconstruction
Robust frameless stereotactic localization in extra-cranial radiotherapy