skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Accuracy of EGSnrc calculations at {sup 60}Co energies for the response of ion chambers configured with various wall materials and cavity dimensions

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3013701· OSTI ID:22095286
;  [1]
  1. Carleton Laboratory for Radiotherapy Physics, Ottawa Carleton Institute of Physics, Carleton University Campus, Ottawa, Ontario K1S 5B6 (Canada)

In this investigation, five experimental data sets are used to evaluate the ability of the EGSnrc Monte Carlo code to calculate the change in chamber response associated with changes in wall material and cavity dimension at {sup 60}Co energies. Calculations of the ratios of response per unit mass of air as a function of cavity volume for walls ranging from polystyrene to lead are generally within 1%-3% of experiments. A few exceptions, which are discussed, include 20%-30% discrepancies with experiments involving lead-walled chambers used by Attix et al. [J. Res. Natl. Bur. Stand. 60, 235-243 (1958)] and Cormack and Johns [Radiat. Res. 1, 133-157 (1954)], and 5% discrepancies for the graphite chamber of Attix et al. (relative to data for other wall materials). Simulations of the experiment by Whyte [Radiat. Res. 6, 371-379 (1957)], which varied cavity air pressure in a large cylindrical chamber, are generally within 0.5% (wall/electrode materials ranging from beryllium to copper). In all cases, the agreement between measurements and EGSnrc calculations is much better when the response as a function of cavity height or air pressure is considered for each wall material individually. High-precision measurements [Burns et al., Phys. Med. Biol. 52, 7125-7135 (2007)] of the response per unit mass as a function of cavity height for a graphite chamber are also accurately reproduced, and validate previous tests of the transport mechanics of EGSnrc. Based on the general agreement found in this work between corresponding experimental results and EGSnrc calculations it can be concluded that EGSnrc can reliably be used to calculate changes in response with changes in various wall materials and cavity dimensions at {sup 60}Co energies within a accuracy of a few percent or less.

OSTI ID:
22095286
Journal Information:
Medical Physics, Vol. 35, Issue 12; Other Information: (c) 2008 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English