Abstract

Langmuir probes are standard instruments for plasma density measurements on many sounding rockets. These probes can be operated in swept-bias as well as in fixed-bias modes. In swept-bias Langmuir probes, contamination effects are frequently visible as a hysteresis between consecutive up and down voltage ramps. This hysteresis, if not corrected, leads to poorly determined plasma densities and temperatures. With a properly chosen sweep function, the contamination parameters can be determined from the measurements and correct plasma parameters can then be determined. In this paper, we study the contamination effects on fixed-bias Langmuir probes, where no hysteresis type effect is seen in the data. Even though the contamination is not evident from the measurements, it does affect the plasma density fluctuation spectrum as measured by the fixed-bias Langmuir probe. We model the contamination as a simple resistor-capacitor circuit between the probe surface and the plasma. We find that measurements of small scale plasma fluctuations (meter to sub-meter scale) along a rocket trajectory are not affected, but the measured amplitude of large scale plasma density variation (tens of meters or larger) is attenuated. From the model calculations, we determine amplitude and cross-over frequency of the contamination effect on fixed-bias probes for differentmore » contamination parameters. The model results also show that a fixed bias probe operating in the ion-saturation region is affected less by contamination as compared to a fixed bias probe operating in the electron saturation region.« less

Authors:
 [1];  [2]
  1. Institut fuer Experimentelle und Angewandte Physik, Christian-Albrechts-Universitaet zu Kiel, 24098 Kiel (Germany)
  2. Department of Physical Sciences, Embry-Riddle Aeronautical University, Daytona Beach, Florida 32114 (United States)
Publication Date:
OSTI Identifier:
22093979
Resource Type:
Journal Article
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 83; Journal Issue: 11; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0034-6748
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; AMPLITUDES; CONTAMINATION; ELECTRON TEMPERATURE; FLUCTUATIONS; HYSTERESIS; ION TEMPERATURE; IONS; LANGMUIR PROBE; PLASMA; PLASMA DENSITY; RESISTORS; SATURATION; SPECTRA; SURFACES

Citation Formats

Steigies, C. T., and Barjatya, A. Contamination effects on fixed-bias Langmuir probes. United States: N. p., 2012. Web. doi:10.1063/1.4764582.
Steigies, C. T., & Barjatya, A. Contamination effects on fixed-bias Langmuir probes. United States. doi:10.1063/1.4764582.
Steigies, C. T., and Barjatya, A. Thu . "Contamination effects on fixed-bias Langmuir probes". United States. doi:10.1063/1.4764582.
@article{osti_22093979,
title = {Contamination effects on fixed-bias Langmuir probes},
author = {Steigies, C. T. and Barjatya, A.},
abstractNote = {Langmuir probes are standard instruments for plasma density measurements on many sounding rockets. These probes can be operated in swept-bias as well as in fixed-bias modes. In swept-bias Langmuir probes, contamination effects are frequently visible as a hysteresis between consecutive up and down voltage ramps. This hysteresis, if not corrected, leads to poorly determined plasma densities and temperatures. With a properly chosen sweep function, the contamination parameters can be determined from the measurements and correct plasma parameters can then be determined. In this paper, we study the contamination effects on fixed-bias Langmuir probes, where no hysteresis type effect is seen in the data. Even though the contamination is not evident from the measurements, it does affect the plasma density fluctuation spectrum as measured by the fixed-bias Langmuir probe. We model the contamination as a simple resistor-capacitor circuit between the probe surface and the plasma. We find that measurements of small scale plasma fluctuations (meter to sub-meter scale) along a rocket trajectory are not affected, but the measured amplitude of large scale plasma density variation (tens of meters or larger) is attenuated. From the model calculations, we determine amplitude and cross-over frequency of the contamination effect on fixed-bias probes for different contamination parameters. The model results also show that a fixed bias probe operating in the ion-saturation region is affected less by contamination as compared to a fixed bias probe operating in the electron saturation region.},
doi = {10.1063/1.4764582},
journal = {Review of Scientific Instruments},
issn = {0034-6748},
number = 11,
volume = 83,
place = {United States},
year = {2012},
month = {11}
}