Title: High-efficiency cross-beam magnetic electron-impact source for improved miniature Mattauch-Herzog mass spectrometer performance

Abstract

We describe a newly designed cross-beam magnetic electron-impact ion source (CBM-EI). We demonstrate its superiority in comparison with a conventional source (CB-EI) when used with a commercial miniature sector-field-type, non-scanning mass spectrometer featuring Mattauch-Herzog geometry (MH-MS) and a permanent sector-field magnet. This paper clearly shows the value of the CBM-EI for enhancing MH-MS sensitivity. Unlike secondary electron-multiplier type detectors, the pixelated detector (IonCCD Trade-Mark-Sign ) used in the commercial MH-MS has no gain. The MH-MS/IonCCD system is therefore challenged to compete with time-of-flight and quadrupole MS systems due to their higher ion transmissions and detector gains. Using the new CBM-EI, we demonstrate an instrument sensitivity increase of 20-fold to 100-fold relative to the CB-EI-equipped instrument. This remarkable signal increase by the simple addition of the magnet assembly arises from the magnet-induced gyromotion of the thermionic electrons, which vastly increases the effective path length of the electrons through the ionization region, and the collimated nature of the electron flux, which optimizes the ion transmission through the 100-{mu}m object slit of the MH-MS. Some or all of the realized sensitivity increase may be exchanged for an increase in resolution and/or mass range through the use of a narrower object slit, or formore » a reduction in ion-source pressure to limit quenching. The CBM-EI should facilitate development of a differentially pumped ion source to extend the lifetime of the filament, especially in otherwise intractable applications associated with oxidizing and corrosive samples.« less

Authors:
;  [1]
  1. OI Analytical/CMS Field Products, 2148 Pelham Parkway, Bldg. 400, Pelham, Alabama 35124 (United States)
Publication Date:
OSTI Identifier:
22093635
Resource Type:
Journal Article
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 83; Journal Issue: 6; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0034-6748
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 43 PARTICLE ACCELERATORS; BEAMS; EFFICIENCY; ELECTRON MULTIPLIERS; ELECTRON SOURCES; ELECTRONS; GAIN; ION SOURCES; IONIZATION; LIFETIME; MAGNETS; MASS; MASS SPECTROMETERS; RESOLUTION; SENSITIVITY; TIME-OF-FLIGHT METHOD; TRANSMISSION

Citation Formats

Hadjar, O., and Fowler, W. K. High-efficiency cross-beam magnetic electron-impact source for improved miniature Mattauch-Herzog mass spectrometer performance. United States: N. p., 2012. Web. doi:10.1063/1.4729115.
Hadjar, O., & Fowler, W. K. High-efficiency cross-beam magnetic electron-impact source for improved miniature Mattauch-Herzog mass spectrometer performance. United States. doi:10.1063/1.4729115.
Hadjar, O., and Fowler, W. K. Fri . "High-efficiency cross-beam magnetic electron-impact source for improved miniature Mattauch-Herzog mass spectrometer performance". United States. doi:10.1063/1.4729115.
@article{osti_22093635,
title = {High-efficiency cross-beam magnetic electron-impact source for improved miniature Mattauch-Herzog mass spectrometer performance},
author = {Hadjar, O. and Fowler, W. K.},
abstractNote = {We describe a newly designed cross-beam magnetic electron-impact ion source (CBM-EI). We demonstrate its superiority in comparison with a conventional source (CB-EI) when used with a commercial miniature sector-field-type, non-scanning mass spectrometer featuring Mattauch-Herzog geometry (MH-MS) and a permanent sector-field magnet. This paper clearly shows the value of the CBM-EI for enhancing MH-MS sensitivity. Unlike secondary electron-multiplier type detectors, the pixelated detector (IonCCD Trade-Mark-Sign ) used in the commercial MH-MS has no gain. The MH-MS/IonCCD system is therefore challenged to compete with time-of-flight and quadrupole MS systems due to their higher ion transmissions and detector gains. Using the new CBM-EI, we demonstrate an instrument sensitivity increase of 20-fold to 100-fold relative to the CB-EI-equipped instrument. This remarkable signal increase by the simple addition of the magnet assembly arises from the magnet-induced gyromotion of the thermionic electrons, which vastly increases the effective path length of the electrons through the ionization region, and the collimated nature of the electron flux, which optimizes the ion transmission through the 100-{mu}m object slit of the MH-MS. Some or all of the realized sensitivity increase may be exchanged for an increase in resolution and/or mass range through the use of a narrower object slit, or for a reduction in ion-source pressure to limit quenching. The CBM-EI should facilitate development of a differentially pumped ion source to extend the lifetime of the filament, especially in otherwise intractable applications associated with oxidizing and corrosive samples.},
doi = {10.1063/1.4729115},
journal = {Review of Scientific Instruments},
issn = {0034-6748},
number = 6,
volume = 83,
place = {United States},
year = {2012},
month = {6}
}