Title: Implementation of test for quality assurance in nuclear medicine gamma camera

Abstract

In nuclear medicine (NM) over 90% of procedures are performed for diagnostic purposes. To ensure adequate diagnostic quality of images and the optimization of the doses received by patients originated from the radioactive material is essential for regular monitoring and equipment performance through a quality assurance program (QAP). The QAP consists of 15 proposed performance tomographic and not tomographic gamma camera (GC) tests, and is based on recommendations of international organizations. We describe some results of the performance parameters of QAP applied to a GC model e.cam Siemens, of the Department of NM of the National Cancer Institute of Mexico (INCan). The results were: (1) The average intrinsic spatial resolution (R{sub in}) was 4.67 {+-} 0.25 mm at the limit of acceptance criterion of 4.4 mm. (2) The sensitivity extrinsic (S{sub ext}), with maximum variations of 1.8% (less than 2% which is the criterion of acceptance). (3) Rotational Uniformity (U{sub rot}), with values of integral uniformity (IU) in the useful field of view detector (UFOV), with maximum percentage change of 0.97% and monthly variations equal angles, ranging from 0.13 to 0.99% less than 1%. (4) The displacement of the center of rotation (DCOR), indicated a maximum deviation of 0.155 {+-}more » 0.039 mm less than 4.795 mm, an absolute deviation of less than 0.5 where pixel 0.085 pixel is suggested, the criteria are assigned to low-energy collimator high resolution. (5) In tomographic uniformity (U{sub tomo}), UI values (%) and percentage noise level (rms%) were 7.54 {+-} 1.53 and 4.18 {+-} 1.69 which are consistent with the limits of acceptance of 7.0-12.0% and 3.0-6.0% respectively. The smallest cold sphere has a diameter of 11.4 mm. The implementation of a QAP allows for high quality diagnostic images, optimization of the doses given to patients, a reduction of exposure to occupationally exposed workers (POE, by its Spanish acronym), and generally improves the productivity of the service. This proposal can be used to develop a similar QAP in other facilities and may serve as a precedent for the proposed regulations for quality assurance (QA) teams in MN.« less

Authors:
; ;  [1]
  1. Department of Nuclear Medicine, National Cancer Institute San Fernando Avenue No.22, Col. Section XVI (Mexico)
Publication Date:
OSTI Identifier:
22075695
Resource Type:
Journal Article
Journal Name:
AIP Conference Proceedings
Additional Journal Information:
Journal Volume: 1494; Journal Issue: 1; Conference: 12. Mexican symposium on medical physics, Oaxaca (Mexico), 16-19 Mar 2012; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-243X
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; 62 RADIOLOGY AND NUCLEAR MEDICINE; DOSIMETRY; GAMMA CAMERAS; IMAGES; MONITORING; MONTHLY VARIATIONS; NEOPLASMS; NUCLEAR INDUSTRY; NUCLEAR MEDICINE; PATIENTS; PERSONNEL; QUALITY ASSURANCE; RADIATION DOSES; RADIOACTIVE MATERIALS; RADIOISOTOPES; SENSITIVITY; SPATIAL RESOLUTION

Citation Formats

Montoya Moreno, A., Rodriguez Laguna, A., and Trujillo Zamudio, Flavio E. Implementation of test for quality assurance in nuclear medicine gamma camera. United States: N. p., 2012. Web. doi:10.1063/1.4764607.
Montoya Moreno, A., Rodriguez Laguna, A., & Trujillo Zamudio, Flavio E. Implementation of test for quality assurance in nuclear medicine gamma camera. United States. doi:10.1063/1.4764607.
Montoya Moreno, A., Rodriguez Laguna, A., and Trujillo Zamudio, Flavio E. Tue . "Implementation of test for quality assurance in nuclear medicine gamma camera". United States. doi:10.1063/1.4764607.
@article{osti_22075695,
title = {Implementation of test for quality assurance in nuclear medicine gamma camera},
author = {Montoya Moreno, A. and Rodriguez Laguna, A. and Trujillo Zamudio, Flavio E},
abstractNote = {In nuclear medicine (NM) over 90% of procedures are performed for diagnostic purposes. To ensure adequate diagnostic quality of images and the optimization of the doses received by patients originated from the radioactive material is essential for regular monitoring and equipment performance through a quality assurance program (QAP). The QAP consists of 15 proposed performance tomographic and not tomographic gamma camera (GC) tests, and is based on recommendations of international organizations. We describe some results of the performance parameters of QAP applied to a GC model e.cam Siemens, of the Department of NM of the National Cancer Institute of Mexico (INCan). The results were: (1) The average intrinsic spatial resolution (R{sub in}) was 4.67 {+-} 0.25 mm at the limit of acceptance criterion of 4.4 mm. (2) The sensitivity extrinsic (S{sub ext}), with maximum variations of 1.8% (less than 2% which is the criterion of acceptance). (3) Rotational Uniformity (U{sub rot}), with values of integral uniformity (IU) in the useful field of view detector (UFOV), with maximum percentage change of 0.97% and monthly variations equal angles, ranging from 0.13 to 0.99% less than 1%. (4) The displacement of the center of rotation (DCOR), indicated a maximum deviation of 0.155 {+-} 0.039 mm less than 4.795 mm, an absolute deviation of less than 0.5 where pixel 0.085 pixel is suggested, the criteria are assigned to low-energy collimator high resolution. (5) In tomographic uniformity (U{sub tomo}), UI values (%) and percentage noise level (rms%) were 7.54 {+-} 1.53 and 4.18 {+-} 1.69 which are consistent with the limits of acceptance of 7.0-12.0% and 3.0-6.0% respectively. The smallest cold sphere has a diameter of 11.4 mm. The implementation of a QAP allows for high quality diagnostic images, optimization of the doses given to patients, a reduction of exposure to occupationally exposed workers (POE, by its Spanish acronym), and generally improves the productivity of the service. This proposal can be used to develop a similar QAP in other facilities and may serve as a precedent for the proposed regulations for quality assurance (QA) teams in MN.},
doi = {10.1063/1.4764607},
journal = {AIP Conference Proceedings},
issn = {0094-243X},
number = 1,
volume = 1494,
place = {United States},
year = {2012},
month = {10}
}