skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Errors in particle tracking velocimetry with high-speed cameras

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3589267· OSTI ID:22062326
; ;  [1]
  1. Department of Physics and Astronomy, University of Iowa, Iowa City, Iowa 52242 (United States)

Velocity errors in particle tracking velocimetry (PTV) are studied. When using high-speed video cameras, the velocity error may increase at a high camera frame rate. This increase in velocity error is due to particle-position uncertainty, which is one of the two sources of velocity errors studied here. The other source of error is particle acceleration, which has the opposite trend of diminishing at higher frame rates. Both kinds of errors can propagate into quantities calculated from velocity, such as the kinetic temperature of particles or correlation functions. As demonstrated in a dusty plasma experiment, the kinetic temperature of particles has no unique value when measured using PTV, but depends on the sampling time interval or frame rate. It is also shown that an artifact appears in an autocorrelation function computed from particle positions and velocities, and it becomes more severe when a small sampling-time interval is used. Schemes to reduce these errors are demonstrated.

OSTI ID:
22062326
Journal Information:
Review of Scientific Instruments, Vol. 82, Issue 5; Other Information: (c) 2011 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English