skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Active suppression of air refractive index fluctuation using a Fabry-Perot cavity and a piezoelectric volume actuator

Journal Article · · Applied Optics
DOI:https://doi.org/10.1364/AO.50.000053· OSTI ID:22036564

Air refractive index fluctuation ({Delta}n{sub air}) is one of the largest uncertainty sources in precision interferometry systems that require a resolution of nanometer order or less. We introduce a method for the active suppression of {Delta}n{sub air} inside a normal air-environment chamber using a Fabry-Perot cavity and a piezoelectric volume actuator. The temporal air refractive index (n{sub air}) at a local point is maintained constant with an expanded uncertainty of {approx}4.2x10{sup -9} (k=2), a sufficiently low uncertainty for precise measurements unaffected by {Delta}n{sub air} to be made inside a chamber.

OSTI ID:
22036564
Journal Information:
Applied Optics, Vol. 50, Issue 1; Other Information: (c) 2010 Optical Society of America; Country of input: International Atomic Energy Agency (IAEA); ISSN 0003-6935
Country of Publication:
United States
Language:
English