Title: Analysis of Prostate Patient Setup and Tracking Data: Potential Intervention Strategies

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
OSTI ID:21590471
; ;  [1]
  1. Department of Radiation Oncology, Virginia Commonwealth University, Richmond, VA (United States)

Purpose: To evaluate the setup, interfraction, and intrafraction organ motion error distributions and simulate intrafraction intervention strategies for prostate radiotherapy. Methods and Materials: A total of 17 patients underwent treatment setup and were monitored using the Calypso system during radiotherapy. On average, the prostate tracking measurements were performed for 8 min/fraction for 28 fractions for each patient. For both patient couch shift data and intrafraction organ motion data, the systematic and random errors were obtained from the patient population. The planning target volume margins were calculated using the van Herk formula. Two intervention strategies were simulated using the tracking data: the deviation threshold and period. The related planning target volume margins, time costs, and prostate position 'fluctuation' were presented. Results: The required treatment margin for the left-right, superoinferior, and anteroposterior axes was 8.4, 10.8, and 14.7 mm for skin mark-only setup and 1.3, 2.3, and 2.8 mm using the on-line setup correction, respectively. Prostate motion significantly correlated among the superoinferior and anteroposterior directions. Of the 17 patients, 14 had prostate motion within 5 mm of the initial setup position for {>=}91.6% of the total tracking time. The treatment margin decreased to 1.1, 1.8, and 2.3 mm with a 3-mm threshold correction and to 0.5, 1.0, and 1.5 mm with an every-2-min correction in the left-right, superoinferior, and anteroposterior directions, respectively. The periodic corrections significantly increase the treatment time and increased the number of instances when the setup correction was made during transient excursions. Conclusions: The residual systematic and random error due to intrafraction prostate motion is small after on-line setup correction. Threshold-based and time-based intervention strategies both reduced the planning target volume margins. The time-based strategies increased the treatment time and the in-fraction position fluctuation.

OSTI ID:
21590471
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 81, Issue 3; Other Information: DOI: 10.1016/j.ijrobp.2010.07.1978; PII: S0360-3016(10)02952-4; Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

Evaluation of multiple image-based modalities for image-guided radiation therapy (IGRT) of prostate carcinoma: A prospective study
Journal Article · Mon Apr 15 00:00:00 EDT 2013 · Medical Physics · OSTI ID:21590471

Influence of intrafraction motion on margins for prostate radiotherapy
Journal Article · Thu Jun 01 00:00:00 EDT 2006 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21590471

Reduction of prostate intrafraction motion using gas-release rectal balloons
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:21590471