skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: 500-MHz x-ray counting with a Si-APD and a fast-pulse processing system

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.3463339· OSTI ID:21410344
; ;  [1]
  1. High Energy Accelerator Research Organization, 1-1 Oho, Tsukuba-shi, Ibaraki-ken 305-0801 (Japan)

We introduce a counting system of up to 500 MHz for synchrotron x-ray high-rate measurements. A silicon avalanche photodiode detector was used in the counting system. The fast-pulse circuit of the amplifier was designed with hybrid ICs to prepare an ASIC system for a large-scale pixel array detector in near future. The fast amplifier consists of two cascading emitter-followers using 10-GHz band transistors. A count-rate of 3.25x10{sup 8} s{sup -1} was then achieved using the system for 8-keV x-rays. However, a baseline shift by adopting AC-coupling in the amplifier disturbed us to observe the maximum count of 4.49x10{sup 8} s{sup -1}, determined by electron-bunch filling into a ring accelerator. We also report that an amplifier with a baseline restorer was tested in order to keep the baseline level to be 0 V even at high input rates.

OSTI ID:
21410344
Journal Information:
AIP Conference Proceedings, Vol. 1234, Issue 1; Conference: SRI 2009: 10. international conference on radiation instrumentation, Melbourne (Australia), 27 Sep - 2 Oct 2009; Other Information: DOI: 10.1063/1.3463339; (c) 2010 American Institute of Physics; ISSN 0094-243X
Country of Publication:
United States
Language:
English