Title: Dosimetry of a Small Air Cavity for Clinical Electron Beams: A Monte Carlo Study

Journal Article · · Medical Dosimetry
 [1];  [2]
  1. Radiation Medicine Program, Princess Margaret Hospital and Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada)
  2. Medical Physics Department, Grand River Regional Cancer Center, Kitchener, Ontario (Canada)

This study investigated dosimetric changes in a water phantom when a small air cavity was presented at the central axis of a clinical electron beam. We used 6-, 9-, and 16-MeV electron beams with a 10 x 10 cm{sup 2} applicator and cutout produced by a Varian 21 EX linear accelerator. Percentage depth doses (PDDs) for different depths (0.5-7 cm), thicknesses (2-10 mm), and widths (1-5 cm) of air cavities were calculated using Monte Carlo simulations (EGSnrc code) validated by film measurements. By comparing PDDs of phantoms with and without the air cavity, it was found that when the depth or thickness of cavity was changed, the PDD curve below the cavity was shifted with a distance equal to the thickness of the cavity. However, when the width of the air cavity was changed, both the PDD curve and its slope within and below the cavity were changed. A larger width of the air cavity resulted in a shallower PDD curve within the cavity. The slope of the PDD curve below the cavity tended towards a value as the width of the air cavity was increased to 3-5 cm for the 6-, 9-, and 16-MeV electron beams. The dependence of the depth dose on the width of the air cavity is a result of the contribution of the electron side scattering in the water surrounding the cavity. The change in depth dose resulting from the presence of an air cavity can cause discrepancies between the calculated and actual dose during radiotherapy, unless the effects of the air cavity are properly characterized during treatment planning. From the dosimetry data in this study, neglecting an air cavity of 1-cm thickness in the build-up region of a 6-MeV electron beam resulted in a delivered dose 10-12% larger than the original prescription. Delivered doses 3% and 6% higher than the prescribed dose were observed when doses were prescribed at R{sub 80} for a 16-MeV electron beam. These results were obtained by neglecting air cavities with thicknesses equal to 2 and 4 mm, respectively, at a depth of 5 cm.

OSTI ID:
21333975
Journal Information:
Medical Dosimetry, Vol. 35, Issue 2; Other Information: DOI: 10.1016/j.meddos.2009.03.004; PII: S0958-3947(09)00012-0; Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
Country of Publication:
United States
Language:
English