Multiplexing Readout of TES Microcalorimeters Based on Analog Baseband Feedback
- Institute of Space and Astronautical Science, 3-1-1 Yoshinodai, Sagamihara, 229-8510 Japan (Japan)
- SRON Netherlands Institute for Space Research, Sorbonnelaan 2, 3584CA Utrecht (Netherlands)
A TES microcalorimeter array is a promising spectrometer with excellent energy resolution and a moderate imaging capability. To realize a large format array in space, multiplexing the TES signals at the low tempersture stage is mandatory. We are developing frequency division multiplexing (FDM) based on baseband feedback technique. In FDM, each TES is AC-biased with a different carrier frequency. Signals from several pixels are summed and then read out by one SQUID. The maximum number of multiplexed pixels are limited by the frequency band in which the SQUID can be operated in a flux-locked loop, which is {approx}1 MHz with standard flux-locked loop circuit. In the baseband feedback, the signal ({approx}10 kHz band) from the TES is once demodulated. Then a reconstructed copy of the modulated signal with an appropriate phase is fed back to the SQUID input coil to maintain an approximately constant magnetic flux. This can be implemented even for large cable delays and automatically suppresses the carrier. We developed a prototype electronics for the baseband feedback based on an analog phase sensitive detector (PSD) and a multiplier. Combined with Seiko 80-SSA SQUID amp, open-loop gain of 8 has been obtained for 10 kHz baseband signal at 5 MHz carrier frequency, with a moderate noise contribution of 27pA/{radical}(Hz) at input.
- OSTI ID:
- 21325814
- Journal Information:
- AIP Conference Proceedings, Vol. 1185, Issue 1; Conference: LTD13: 13. international workshop on low temperature detectors, Stanford, CA (United States), 20-24 Jul 2009; Other Information: DOI: 10.1063/1.3292397; (c) 2009 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance of an X-ray single pixel TES microcalorimeter under DC and AC biasing
Direct current superconducting quantum interference device spectrometer for pulsed nuclear magnetic resonance and nuclear quadrupole resonance at frequencies up to 5 MHz