Test-bench calibration of the Double-Slit Ion Energy Analyzer for Heavy Ion Beam Probing Diagnostics
Journal Article
·
· AIP Conference Proceedings
- Institute of Plasma Physics, NSC KIPT, Kharkov (Ukraine)
- INF RRC 'Kurchatov Institute', Moscow (Russian Federation)
This work presents the results of the design study, investigations of the features and a calibration of double-slit ion energy analyzer prototype at the test-bench in IPP NSC KIPT, Kharkov, Ukraine, with beam energy up to 10 keV. Hitherto obtained accuracy for both analyzer slits is {delta}G-2x10{sup -3} and {delta}F-3x10{sup -3}, which is limited by the existing step motor accuracy 0.1 mm for entrance slits moving. The presented study work is considered as the first step to the multi-slit analyzer elaboration.
- OSTI ID:
- 21136916
- Journal Information:
- AIP Conference Proceedings, Vol. 993, Issue 1; Conference: PLASMA 2007: International conference on research and applications of plasmas; 4. German-Polish conference on plasma diagnostics for fusion and applications; 6. French-Polish seminar on thermal plasma in space and laboratory, Greifswald (Germany), 16-19 Oct 2007; Other Information: DOI: 10.1063/1.2909119; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Neutron Source Facility of the National Science Center “Kharkiv Institute of Physics and Technology” at Kharkiv, Ukraine
Plasma wakefield diagnostics using probe electron beam and microchannel plates
Measuring and monitoring KIPT Neutron Source Facility Reactivity
Technical Report
·
Mon Oct 03 00:00:00 EDT 2022
·
OSTI ID:21136916
+6 more
Plasma wakefield diagnostics using probe electron beam and microchannel plates
Conference
·
Tue Dec 31 00:00:00 EST 1996
·
OSTI ID:21136916
Measuring and monitoring KIPT Neutron Source Facility Reactivity
Technical Report
·
Sat Aug 01 00:00:00 EDT 2015
·
OSTI ID:21136916