Absolute calibration of an electron spectrometer using high energy electrons produced by the laser-plasma interaction
Journal Article
·
· Review of Scientific Instruments
- National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 2, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8568 (Japan)
An in situ observation system has been developed to observe the absolute electron energy spectra of electron beams generated by laser-plasma interaction. A phosphor screen (DRZ) coupled with a charge coupled device camera is used to detect the electrons. A new method is proposed to calibrate the absolute sensitivity of the detection system for a wide energy range with a single shot by using an electron beam generated by laser-plasma interaction. The sensitivity of the system is found to be high, which is comparable to that of an imaging plate.
- OSTI ID:
- 21124109
- Journal Information:
- Review of Scientific Instruments, Vol. 79, Issue 8; Other Information: DOI: 10.1063/1.2969655; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Note: Absolute calibration of two DRZ phosphor screens using ultrashort electron bunch
Absolute calibration for a broad range single shot electron spectrometer
High-resolution imaging spectrometer for recording absolutely calibrated far ultraviolet spectra from laser-produced plasmas
Journal Article
·
Wed Feb 15 00:00:00 EST 2012
· Review of Scientific Instruments
·
OSTI ID:21124109
+2 more
Absolute calibration for a broad range single shot electron spectrometer
Journal Article
·
Sun Oct 15 00:00:00 EDT 2006
· Review of Scientific Instruments
·
OSTI ID:21124109
+6 more
High-resolution imaging spectrometer for recording absolutely calibrated far ultraviolet spectra from laser-produced plasmas
Journal Article
·
Wed Oct 15 00:00:00 EDT 2008
· Review of Scientific Instruments
·
OSTI ID:21124109
+5 more