Title: Absolute calibration of an electron spectrometer using high energy electrons produced by the laser-plasma interaction

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2969655· OSTI ID:21124109
; ; ;  [1]
  1. National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 2, 1-1-1 Umezono, Tsukuba, Ibaraki 305-8568 (Japan)

An in situ observation system has been developed to observe the absolute electron energy spectra of electron beams generated by laser-plasma interaction. A phosphor screen (DRZ) coupled with a charge coupled device camera is used to detect the electrons. A new method is proposed to calibrate the absolute sensitivity of the detection system for a wide energy range with a single shot by using an electron beam generated by laser-plasma interaction. The sensitivity of the system is found to be high, which is comparable to that of an imaging plate.

OSTI ID:
21124109
Journal Information:
Review of Scientific Instruments, Vol. 79, Issue 8; Other Information: DOI: 10.1063/1.2969655; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English