Abstract

Purpose. We evaluated the Gunther Tulip (GT) retrievable inferior vena cava (IVC) filter with regard to indications, filtration efficacy, complications, retrieval window, and use of anticoagulation. Method. A retrospective study was performed of 147 patients (64 men, 83 women; mean age 58.8 years) who underwent retrievable GT filter insertion between 2001 and 2005. The indications for placement included a diagnosis of pulmonary embolism or deep venous thrombosis with a contraindication to anticoagulation (n = 68), pulmonary embolism or deep venous thrombosis while on anticoagulation (n = 49), prophylactic filter placement for high-risk surgical patients with a past history of pulmonary embolism or deep venous thrombosis (n = 20), and a high risk of pulmonary embolism or deep venous thrombosis (n = 10). Forty-nine of the 147 patients did not receive anticoagulation (33.7%) while 96 of 147 patients did, 82 of these receiving warfarin (56.5%), 11 receiving low-molecular weight heparins (7.58%), and 3 receiving antiplatelet agents alone (2.06%). Results. Filter placement was successful in 147 patients (100%). Two patients had two filters inserted. Of the 147 patients, filter deployment was on a permanent basis in 102 and with an intention to retrieve in 45 patients. There were 36 (80%) successful retrievalsmore » and 9 (20%) failed retrievals. The mean time to retrieval was 33.6 days. The reasons for failed retrieval included filter struts tightly adherent to the IVC wall (5/9), extreme filter tilt (2/9), and extensive filter thrombus (2/9). Complications included pneumothorax (n = 4), failure of filter expansion (n = 1), and breakthrough pulmonary embolism (n = 1). No IVC thrombotic episodes were recorded. Discussion. The Gunther Tulip retrievable filter can be used as a permanent or a retrievable filter. It is safe and efficacious. GT filters can be safely retrieved at a mean time interval of 33.6 days. The newly developed Celect filter may extend the retrieval interval.« less

Authors:
; ; ; ;  [1]
  1. Beaumont Hospital, Department of Radiology (Ireland), E-mail: mlee@rcsi.ie
Publication Date:
OSTI Identifier:
21091047
Resource Type:
Journal Article
Resource Relation:
Journal Name: Cardiovascular and Interventional Radiology; Journal Volume: 30; Journal Issue: 1; Other Information: DOI: 10.1007/s00270-006-0093-9; Copyright (c) 2007 Springer Science+Business Media, Inc.; www.springer-ny.com; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; BLOOD VESSELS; DIAGNOSIS; FILTERS; FILTRATION; PATIENTS; SURGERY; THROMBOSIS; VASCULAR DISEASES

Citation Formats

Looby, S., Given, M.F., Geoghegan, T., McErlean, A., and Lee, M.J. Gunther Tulip Retrievable Inferior Vena Caval Filters: Indications, Efficacy, Retrieval, and Complications. United States: N. p., 2007. Web. doi:10.1007/S00270-006-0093-9.
Looby, S., Given, M.F., Geoghegan, T., McErlean, A., & Lee, M.J. Gunther Tulip Retrievable Inferior Vena Caval Filters: Indications, Efficacy, Retrieval, and Complications. United States. doi:10.1007/S00270-006-0093-9.
Looby, S., Given, M.F., Geoghegan, T., McErlean, A., and Lee, M.J. Thu . "Gunther Tulip Retrievable Inferior Vena Caval Filters: Indications, Efficacy, Retrieval, and Complications". United States. doi:10.1007/S00270-006-0093-9.
@article{osti_21091047,
title = {Gunther Tulip Retrievable Inferior Vena Caval Filters: Indications, Efficacy, Retrieval, and Complications},
author = {Looby, S. and Given, M.F. and Geoghegan, T. and McErlean, A. and Lee, M.J.},
abstractNote = {Purpose. We evaluated the Gunther Tulip (GT) retrievable inferior vena cava (IVC) filter with regard to indications, filtration efficacy, complications, retrieval window, and use of anticoagulation. Method. A retrospective study was performed of 147 patients (64 men, 83 women; mean age 58.8 years) who underwent retrievable GT filter insertion between 2001 and 2005. The indications for placement included a diagnosis of pulmonary embolism or deep venous thrombosis with a contraindication to anticoagulation (n = 68), pulmonary embolism or deep venous thrombosis while on anticoagulation (n = 49), prophylactic filter placement for high-risk surgical patients with a past history of pulmonary embolism or deep venous thrombosis (n = 20), and a high risk of pulmonary embolism or deep venous thrombosis (n = 10). Forty-nine of the 147 patients did not receive anticoagulation (33.7%) while 96 of 147 patients did, 82 of these receiving warfarin (56.5%), 11 receiving low-molecular weight heparins (7.58%), and 3 receiving antiplatelet agents alone (2.06%). Results. Filter placement was successful in 147 patients (100%). Two patients had two filters inserted. Of the 147 patients, filter deployment was on a permanent basis in 102 and with an intention to retrieve in 45 patients. There were 36 (80%) successful retrievals and 9 (20%) failed retrievals. The mean time to retrieval was 33.6 days. The reasons for failed retrieval included filter struts tightly adherent to the IVC wall (5/9), extreme filter tilt (2/9), and extensive filter thrombus (2/9). Complications included pneumothorax (n = 4), failure of filter expansion (n = 1), and breakthrough pulmonary embolism (n = 1). No IVC thrombotic episodes were recorded. Discussion. The Gunther Tulip retrievable filter can be used as a permanent or a retrievable filter. It is safe and efficacious. GT filters can be safely retrieved at a mean time interval of 33.6 days. The newly developed Celect filter may extend the retrieval interval.},
doi = {10.1007/S00270-006-0093-9},
journal = {Cardiovascular and Interventional Radiology},
number = 1,
volume = 30,
place = {United States},
year = {Thu Feb 15 00:00:00 EST 2007},
month = {Thu Feb 15 00:00:00 EST 2007}
}