Resonant excitation of injected atomic ions in an ion trap mass spectrometer
Abstract
Resonant excitation of ion motion in an ion trap mass spectrometer(ITMS) is not a new idea. Most previous reports have characterized the resonant absorption profile (loss of observed ion signal versus excitation frequency) with respect to space charge, and the amplitude (V{sub ac}) and duration (T) of the excitation pulse. These reports have often focused on (single frequency) resonant excitation to effect collisionally induced dissociation for MS/MS. The authors report studies of single frequency and broadband resonant excitation applied to selective ion accumulation or ejection in elemental mass spectrometry.
- Authors:
-
- Pacific Northwest Lab., Richland, WA (United States); and others
- Publication Date:
- OSTI Identifier:
- 210694
- Report Number(s):
- CONF-9505261-
TRN: 96:001157-0133
- Resource Type:
- Conference
- Resource Relation:
- Conference: 43. American Society of Mass Spectrometry (ASMS) conference on mass spectrometry and allied topics, Atlanta, GA (United States), 21-26 May 1995; Other Information: PBD: 1995; Related Information: Is Part Of Proceedings of the 43rd ASMS conference on mass spectrometry and allied topics; PB: 1411 p.
- Country of Publication:
- United States
- Language:
- English
- Subject:
- 40 CHEMISTRY; SILVER; ICP MASS SPECTROSCOPY; CADMIUM; STRONTIUM; YTTRIUM; ZIRCONIUM
Citation Formats
Eiden, G C, Barinaga, C J, and Koppenaal, D W. Resonant excitation of injected atomic ions in an ion trap mass spectrometer. United States: N. p., 1995.
Web.
Eiden, G C, Barinaga, C J, & Koppenaal, D W. Resonant excitation of injected atomic ions in an ion trap mass spectrometer. United States.
Eiden, G C, Barinaga, C J, and Koppenaal, D W. Sun .
"Resonant excitation of injected atomic ions in an ion trap mass spectrometer". United States.
@article{osti_210694,
title = {Resonant excitation of injected atomic ions in an ion trap mass spectrometer},
author = {Eiden, G C and Barinaga, C J and Koppenaal, D W},
abstractNote = {Resonant excitation of ion motion in an ion trap mass spectrometer(ITMS) is not a new idea. Most previous reports have characterized the resonant absorption profile (loss of observed ion signal versus excitation frequency) with respect to space charge, and the amplitude (V{sub ac}) and duration (T) of the excitation pulse. These reports have often focused on (single frequency) resonant excitation to effect collisionally induced dissociation for MS/MS. The authors report studies of single frequency and broadband resonant excitation applied to selective ion accumulation or ejection in elemental mass spectrometry.},
doi = {},
url = {https://www.osti.gov/biblio/210694},