skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Characterization of noise sources for two generations of computed radiography systems using powder and crystalline photostimulable phosphors

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2750973· OSTI ID:20953510
;  [1]
  1. KCARE, King's College Hospital, Denmark Hill, London, SE5 9RS (United Kingdom)

The performances of two generations of computed radiography (CR) were tested and compared in terms of resolution and noise characteristics. The main aim was to characterize and quantify the noise sources in the images. The systems tested were (1) Agfa CR 25.0, a flying spot reader with powder phosphor image plates (MD 40.0); and (2) the Agfa DX-S, a line-scanning CR reader with needle crystal phosphor image plates (HD 5.0). For both systems, the standard metrics of presampled modulation transfer function (MTF), normalized noise power spectra (NNPS) and detective quantum efficiency (DQE) were measured using standard radiation quality RQA5 as defined by the International Electrotechnical Commission. The various noise sources contributing to the NNPS were separated by using knowledge of their relationship with air kerma, MTF, absorption efficiency and antialiasing filters. The DX-S MTF was superior compared with the CR 25.0. The maximum difference in MTF between the DX-S scan and CR 25.0 subscan directions was 0.13 at 1.3 mm{sup -1}. For a nominal detector air kerma of 4 {mu}Gy, the peak DQE of the DX-S was 43({+-}3)%, which was over double that of the CR 25.0 of 18({+-}2)%. The additive electronic noise was negligible on the CR 25.0 but calculated to be constant 3.4x10{sup -7} ({+-}0.4x10{sup -7}) mm{sup 2} at 3.9 {mu}Gy on the DX-S. The DX-S has improved image quality compared with a traditional flying spot reader. The separation of the noise sources indicates that the improvements in DQE of the DX-S are due not only to the higher quantum, efficiency and MTF, but also the lower structure, secondary quantum, and excess noise.

OSTI ID:
20953510
Journal Information:
Medical Physics, Vol. 34, Issue 8; Other Information: DOI: 10.1118/1.2750973; (c) 2007 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English